Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
GetHealthStatus | 0403 | GetHealthStatus | msrc_UNKNOWN | Details | |
ResetCurrentHold_method | 41F1 | Reset Current Hold | msrc_UNKNOWN | Details | |
SetCurrentHold_method | 41F2 | Set Current hold | msrc_UNKNOWN | Details | |
SW_Write_Protection_method | 40AF | Modify SW Write Protection | Select software write protection status. For configuration parameters to be editable and calibrations to be performed, both hardware and software write protection has to disabled (off). | msrc_UNKNOWN | Details |
ReadSWRev_method | 409D | Software Rev | Revisão do Software do dispositivo de campo – Essa revisão corresponde ao software ou firmware, que está incorporado no dispositivo de campo | msrc_UNKNOWN | Details |
ReadHWRev_method | 409E | Hardware Rev | Revisão do Hardware- Revisão que corresponde ao hardware eletrônico do Dispositivo de campo. | msrc_UNKNOWN | Details |
Hart_HartAccessLevel_method | 412B | Set HART Access Level | Select 'Calibratoin' for Calibrate access level. Select 'Advanced' for Advanced access level. Select 'Service' for Service access level. | msrc_UNKNOWN | Details |
reset_config_flag | 42F1 | Reset configuration change | msrc_UNKNOWN | Details | |
method_read_unit | 4361 | msrc_UNKNOWN | Details | ||
method_upload_measurement_value | 4333 | msrc_UNKNOWN | Details | ||
method_download_measurement_value | 4332 | msrc_UNKNOWN | Details | ||
ResetPHCalStatus_method | 41F3 | Reset Calibration Status | msrc_UNKNOWN | Details | |
AutoBufferCalibration_method | 414E | Auto Buffer Calibration | Auto buffer calibration. Select calibration buffer type (one of the pre-defined or user-defined) and the two calibration points before running this procedure. The procedure performs the calibration steps. The sensor temperature is used for calibration. Device automatically detects pH and temperature stability before performing calibration | msrc_UNKNOWN | Details |
OnePointPVManualCal_method | 416E | One Point Manual Calibration | msrc_UNKNOWN | Details | |
TwoPointPVManualCal_method | 416F | Two Point Manual Calibration | Two-point Manual PV cal lets the user enter the two cal point values and the temperature of the calibration buffer. User manually determines when the process value is stable. | msrc_UNKNOWN | Details |
PHTemperatureCal_method | 40C7 | Temperature Calibration | The temperature calibration is used to calibrate the temperature input from the sensor against a standard known reading. | msrc_UNKNOWN | Details |
DetectTCType_methodpH | 414C | Detect TC Type | Starts the auto-recognition process for the temperature sensor. | msrc_UNKNOWN | Details |
ResettoDefaults_methodpH | 40AB | Reset to Defaults | Warning! The device will power cycle; reset most of configurable parameters to factory default values. The parameters that are not reset are - HART device address, serial number, hardware and software version numbers. | msrc_UNKNOWN | Details |
Reset_cal_methodpH | 40C9 | Restore Defaults | msrc_UNKNOWN | Details | |
ResetTCCalStatus_method | 41F6 | Reset Calibration Status | msrc_UNKNOWN | Details | |
PVZeroCal_method | 417E | PV Zero Cal | The process value's zero calibration | msrc_UNKNOWN | Details |
PVSpanCal_method | 417F | PV Span Cal | The process value's span calibration | msrc_UNKNOWN | Details |
TCTemperatureCal_method | 40D3 | Temperature Calibration | The temperature calibration is used to calibrate the temperature input from the sensor against a standard known reading. | msrc_UNKNOWN | Details |
DetectTCType_methodTC | 417D | Detect TC Type | Starts the auto-recognition process for the temperature sensor. | msrc_UNKNOWN | Details |
ResettoDefaults_methodTC | 40AE | Reset to Defaults | Warning! The device will power cycle; reset most of configurable parameters to factory default values. The parameters that are not reset are - HART device address, serial number, hardware and software version numbers. | msrc_UNKNOWN | Details |
Reset_cal_methodTC | 40D4 | Restore Defaults | Restore default calibration co-efficients. | msrc_UNKNOWN | Details |
ResetECCalStatus_method | 41F8 | Reset Calibration Status | msrc_UNKNOWN | Details | |
ECCondOrConcCalibration_method | 4146 | Start Cond/Conc Cal | msrc_UNKNOWN | Details | |
ECTemperatureCal_method | 40CC | Temperature Cal | msrc_UNKNOWN | Details | |
DetectTCType_methodEC | 4145 | Detect TC Type | Starts the auto-recognition process for the temperature sensor. | msrc_UNKNOWN | Details |
ResettoDefaults_methodEC | 40AC | Reset to Defaults | Warning! The device will power cycle; reset most of configurable parameters to factory default values. The parameters that are not reset are - HART device address, serial number, hardware and software version numbers. | msrc_UNKNOWN | Details |
Reset_cal_methodEC | 40CD | Restore Defaults | msrc_UNKNOWN | Details | |
ResetTECalStatus_method | 41FA | Reset Calibration Status | msrc_UNKNOWN | Details | |
TECondOrConcCalibration_method | 418B | Conductivity/Concentration cal | Calibration of a conductivity/Concentration process value. | msrc_UNKNOWN | Details |
TETemperatureCal_method | 40CF | Temperature Cal | msrc_UNKNOWN | Details | |
DetectTCType_methodTE | 418A | Detect TC Type | Starts the auto-recognition process for the temperature sensor. | msrc_UNKNOWN | Details |
ResettoDefaults_methodTE | 40AD | Reset to Defaults | Warning! The device will power cycle; reset most of configurable parameters to factory default values. The parameters that are not reset are - HART device address, serial number, hardware and software version numbers. | msrc_UNKNOWN | Details |
Reset_cal_methodTE | 40D0 | Restore Defaults | msrc_UNKNOWN | Details | |
device_master_reset | 1072 | Restaurar o dispositivo | Restaurar dispositivo – comanda os eletrônicos para realizar a restauração. Esse equivalente ao ciclo de energia desligada e então retorne para o dispositivo de campo. | msrc_UNKNOWN | Details |
transmitter_loop_test | 1023 | Teste de malha | Teste de malha- Permite o usuário a manipular manualemnte a saída analógica para o valor de saída de constante selecionada | msrc_UNKNOWN | Details |
transmitter_dac_trim | 1024 | Analog Output Trim | Digital para ajuste analógico- Permite a calibração da saída analógica selecionada com uma referência externa nos endpoints de operação da saída analógica | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | Deixar o modo fixado atual | Modo atual de deixar fixado- Leva uma saída analógica para fora do modo onde a saída foi um valor constant, e retorna a saída para o modo onde a saída segue a entrada | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Abortar retorno para normal | msrc_UNKNOWN | Details | |
Name | ID | Label | Description | Type | Source |