Loading…

方法名称: TwoPointPVManualCal_method

方法ID: 0x416F

方法标签: Two Point Manual Calibration

方法描述: Two-point Manual PV cal lets the user enter the two cal point values and the temperature of the calibration buffer. User manually determines when the process value is stable.


  • {
  • char status[3];
  • char string1[60];
  • int len1;
  • int current;
  • int selection;
  • len1 = 60;
  • _set_all_resp_code(0);
  • _set_xmtr_all_resp_code(0);
  • _set_device_status(0xFF,0);
  • _set_xmtr_device_status(0xFF,0);
  • _get_dictionary_string(19726337,string1,len1);
  • ACKNOWLEDGE(string1);
  • _add_abort_method(4118);
  • send(234,status);
  • if (status[0])
  • {
  • display_response_status(234,status[0]);
  • DELAY(4,literal_string(833));
  • process_abort();
  • }
  • if (MeasurementPh_calibrationStatus != 0 && MeasurementPh_calibrationStatus != 7 && MeasurementPh_calibrationStatus != 8 && MeasurementPh_calibrationStatus != 11 && MeasurementPh_calibrationStatus != 9 && MeasurementPh_calibrationStatus != 10 && MeasurementPh_calibrationStatus != 22 && MeasurementPh_calibrationStatus != 21)
  • {
  • _remove_abort_method(4118);
  • ACKNOWLEDGE(literal_string(834));
  • process_abort();
  • }
  • current = SetCurrentHold_method();
  • if (current == 0)
  • {
  • _remove_abort_method(4118);
  • _add_abort_method(16881);
  • _add_abort_method(4118);
  • }
  • send(225,status);
  • if (status[0])
  • {
  • display_response_status(225,status[0]);
  • DELAY(4,literal_string(833));
  • process_abort();
  • }
  • _get_dev_var_value(literal_string(858),0,16885);
  • send(226,status);
  • if (status[0])
  • {
  • display_response_status(226,status[0]);
  • DELAY(4,literal_string(833));
  • process_abort();