0 |
EPROM Checksum: Replace Transmitter
or
Replace Transmitter |
1 |
RAM Error: Replace Transmitter |
2 |
Debug, IRQ: Inform Dräger
Fax: +49-451-882-2042 |
3 |
Cmd 48, Byte 0, Bit 0x10 Set |
4 |
Cmd 48, Byte 0, Bit 0x08 Set |
5 |
Debug, com1: Inform Dräger
Fax: +49-451-882-2042 |
6 |
Debug, com2: Inform Dräger
Fax: +49-451-882-2042 |
7 |
Cmd 48, Byte 0, Bit 0x01 Set |
8 |
No sensorConnect sensor
or
Check sensor for electrolyte leakage: Resistance between pins 3<->6 shall be > 10MOhms
or
Replace Sensor
or
Replace Transmitter |
9 |
Invalid sensor typeUse only sensors intended for Polytron2 |
10 |
Invalid sensor versionUse only sensors intended for Polytron2
or
Replace Transmitter |
11 |
Wrong sensor installedInstall Senser with correct Part No. or disable Sensor Lock temporarily |
12 |
Wrong Gas Database |
13 |
Invalid EEPROM versionPerform Standard Initialization
or
Perform Fundamental Initialization
or
Calibrate analog interface
or
Replace Transmitter |
14 |
Cmd 48, Byte 1, Bit 0x02 Set |
15 |
Cmd 48, Byte 1, Bit 0x01 Set |
16 |
I2C-bus error during sensor data uploadReplace Sensor
or
Replace Transmitter |
17 |
No valid sensor data read blockReplace Sensor
or
Replace Transmitter |
18 |
No valid sensor data notice blockReplace Sensor
or
Replace Transmitter |
19 |
No valid sensor data write blockReplace Sensor
or
Replace Transmitter |
20 |
Checksum of gas name table corruptedInterrupt supply voltage > 1 sec (Reset) |
21 |
Cmd 48, Byte 2, Bit 0x04 Set |
22 |
Cmd 48, Byte 2, Bit 0x02 Set |
23 |
Cmd 48, Byte 2, Bit 0x01 Set |
24 |
Manufacturer database corruptedReplace Transmitter |
25 |
EEPROM database corruptedPerform Fundamental Initialization
or
Calibrate analog interface |
26 |
I2C-bus error during sensor data downloadInterrupt supply voltage > 1 sec (Reset) |
27 |
Sensor stack 1 ErrorInform Dräger
Fax: +49-451-882-2042 |
28 |
Sensor stack 2 ErrorInform Dräger
Fax: +49-451-882-2042 |
29 |
HC11 Stack 1: Inform Dräger
Fax: +49-451-882-2042 |
30 |
HC11 Stack 2: Inform Dräger
Fax: +49-451-882-2042 |
31 |
Cmd 48, Byte 3, Bit 0x01 Set |
32 |
Cmd 48, Byte 4, Bit 0x80 Set |
33 |
Cmd 48, Byte 4, Bit 0x40 Set |
34 |
Cmd 48, Byte 4, Bit 0x20 Set |
35 |
Analog Database Checksum: Interrupt supply voltage > 1 sec (Reset) |
36 |
Install test failedCheck supply voltage |
37 |
Warning Parameters corrupted: Inform Dräger
Fax: +49-451-882-2042 |
38 |
Invalid Supply Voltage: Check supply voltage |
39 |
Invalid Loop Current: Check supply voltage; Calibrate analog interface
or
Replace Transmitter |
40 |
Cmd 48, Byte 5, Bit 0x80 Set |
41 |
Cmd 48, Byte 5, Bit 0x40 Set |
42 |
Cmd 48, Byte 5, Bit 0x20 Set |
43 |
Cmd 48, Byte 5, Bit 0x10 Set |
44 |
Cmd 48, Byte 5, Bit 0x08 Set |
45 |
Cmd 48, Byte 5, Bit 0x04 Set |
46 |
Cmd 48, Byte 5, Bit 0x02 Set |
47 |
Cmd 48, Byte 5, Bit 0x01 Set |
48 |
No valid span calibrationCalibrate Sensor (zero) and Calibrate Sensor (span) |
49 |
No valid zero calibrationCalibrate Sensor (zero) |
50 |
No write access to sensor database during calibrationInterrupt supply voltage > 1 sec (Reset) and Calibrate Sensor (zero) and Calibrate Sensor (span)
or
Replace Sensor
or
Replace Transmitter |
51 |
Cmd 48, Byte 14, Bit 0x10 Set |
52 |
Too negative concentrationClean ambient from gasses with negative cross sensitivity
or
Calibrate Sensor (zero) |
53 |
Sensor failed self testReplace Sensor |
54 |
Sensor potential incompatibleUse only sensors intended for Polytron2 |
55 |
Cmd 48, Byte 14, Bit 0x01 Set |
56 |
Debug measurement 1 ErrorInform Dräger
Fax: +49-451-882-2042 |
57 |
Debug ADC 1 ErrorInform Dräger
Fax: +49-451-882-2042 |
58 |
Debug ADC 2 ErrorInform Dräger
Fax: +49-451-882-2042 |
59 |
Incorrect sensor potentialCheck supply voltage
or
Replace Transmitter |
60 |
Pre-Amplifier Offset ErrorReplace Transmitter |
61 |
Pre-Amplifier power consumption ErrorReplace Transmitter |
62 |
Debug measurement 3 ErrorInform Dräger
Fax: +49-451-882-2042 |
63 |
ADC overflowReplace Transmitter
or
Interrupt supply voltage > 1 sec (Reset) |
64 |
ADC timing ErrorInform Dräger
Fax: +49-451-882-2042 |
65 |
Cmd 48, Byte 16, Bit 0x20 Set |
66 |
ADC common mode ErrorCheck supply voltage
or
Replace Transmitter |
67 |
ADC span ErrorCheck supply voltage
or
Replace Transmitter |
68 |
ADC zero ErrorCheck supply voltage
or
Replace Transmitter |
69 |
HC11 ADC span ErrorCheck supply voltage
or
Replace Transmitter |
70 |
HC11 ADC zero ErrorCheck supply voltage
or
Replace Transmitter |
71 |
Cmd 48, Byte 17, Bit 0x80 Set |
72 |
Cmd 48, Byte 17, Bit 0x40 Set |
73 |
Cmd 48, Byte 17, Bit 0x20 Set |
74 |
Cmd 48, Byte 17, Bit 0x10 Set |
75 |
Cmd 48, Byte 17, Bit 0x08 Set |
76 |
Cmd 48, Byte 17, Bit 0x04 Set |
77 |
Cmd 48, Byte 17, Bit 0x02 Set |
78 |
Cmd 48, Byte 17, Bit 0x01 Set |
79 |
Cmd 48, Byte 18, Bit 0x80 Set |
80 |
Cmd 48, Byte 18, Bit 0x40 Set |
81 |
Cmd 48, Byte 18, Bit 0x20 Set |
82 |
Cmd 48, Byte 18, Bit 0x10 Set |
83 |
Cmd 48, Byte 18, Bit 0x08 Set |
84 |
ADC invalid paramtersInform Dräger
Fax: +49-451-882-2042 |
85 |
ADC time outInform Dräger
Fax: +49-451-882-2042 |
86 |
ADC overrunInform Dräger
Fax: +49-451-882-2042 |
87 |
Analog span not calibrated: Calibrate analog interface |
88 |
Analog zero not calibrated: Calibrate analog interface |
89 |
Cmd 48, Byte 22, Bit 0x20 Set |
90 |
Cmd 48, Byte 22, Bit 0x10 Set |
91 |
Cmd 48, Byte 22, Bit 0x08 Set |
92 |
Date not programmedSet Clock |
93 |
Time not programmedSet Clock |
94 |
Clock back-up supply downSet Jumper X3 or replace Li-battery
or
Replace Transmitter |
95 |
Cmd 48, Byte 23, Bit 0x80 Set |
96 |
Cmd 48, Byte 23, Bit 0x40 Set |
97 |
Sensor warm-up, no measurement: Wait for sensor to warm up |
98 |
Sensor warm-up, Measurement error <4x spec.: Wait for sensor to warm up |
99 |
Sensor counter electrode voltage out of rangeStop non permissable permanent gassing
or
Replace Sensor
or
Replace Transmitter
or
Replace Transmitter |
100 |
Gas concentration is too highReduce gas concentration
or
Replace Sensor |
101 |
Sensor temperature too lowAdjust ambient temperature to permissable range
or
Replace Sensor
or
Replace Transmitter |
102 |
Sensor temperature too highAdjust ambient temperature to permissable range
or
Replace Sensor
or
Replace Transmitter |
103 |
End of sensor life due to exposureReplace Sensor |
104 |
End of sensor life due to poor sensitivityReplace Sensor |
105 |
Cmd 48, Byte 24, Bit 0x20 Set |
106 |
Cmd 48, Byte 24, Bit 0x10 Set |
107 |
Too high exposure since last (span) calibrationCalibrate Sensor (zero) and Calibrate Sensor (span) |
108 |
Calibration interval expiredCalibrate Sensor (zero) and Calibrate Sensor (span) |
109 |
Gas concentration was too highCalibrate Sensor (zero) and Calibrate Sensor (span) |
110 |
Sensor temperature was too highCalibrate Sensor (zero) and Calibrate Sensor (span) |
111 |
.1f |
112 |
.2f |
113 |
8.2f |
114 |
5.2f |
115 |
4.2f |
116 |
Upload Var List |
117 |
4d |
118 |
2d |
119 |
AlmMinMax |
120 |
RangeRef |
121 |
.3f |
122 |
02X |
123 |
7d |
124 |
8.1f |
125 |
8.3f |
126 |
d |
127 |
3d |
128 |
2x |
129 |
p2GasIndex |
130 |
p2Units1 |
131 |
p2Units2 |
132 |
p2Units3 |
133 |
p2Units4 |
134 |
p2Units5 |
135 |
p2Units6 |
136 |
p2Units7 |
137 |
p2Units8 |
138 |
p2Units9 |
139 |
p2Units10 |
140 |
p2Units11 |
141 |
p2Units12 |
142 |
p2Units13 |
143 |
p2Units14 |
144 |
p2Units15 |
145 |
p2Units16 |
146 |
p2Units17 |
147 |
p2Units18 |
148 |
p2Units19 |
149 |
p2Units20 |
150 |
p2Units21 |
151 |
p2Units22 |
152 |
p2Units23 |
153 |
p2Units24 |
154 |
p2Units25 |
155 |
p2Units26 |
156 |
p2Units27 |
157 |
p2Units28 |
158 |
p2Units29 |
159 |
p2Units30 |
160 |
p2Units31 |
161 |
p2UnitIndex |
162 |
8u |
163 |
Panic: Unknown Bit Cmd 48, Byte 19, 0x10 |
164 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x10 |
165 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x08 |
166 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x04 |
167 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x02 |
168 |
Panic: Unknown Bit Cmd 48, Byte 21, 0x01 |
169 |
SetupErrorCfg |
170 |
WrapErrorCfg |
171 |
EnterParamMode |
172 |
ExitParamMode |
173 |
EnterGasConfig |
174 |
ExitGasConfig |
175 |
Unexpected Error Command %{cmd} |
176 |
The handheld will temporarily loose communication with the Polytron. Please Select "Online" to Reconnect |
177 |
Online Info |
178 |
Mode %{1} %{2} %{3}
Online Info
%{4} %{5} %{6} %{7} %{8} %{14}
%{9} %{10} %{11} %{12} %{13} %{15} %{16} |
179 |
Master Reset |
180 |
Master Reset - Performs a master reset of the field device. Equivalent to cycling the transmitter power supply |
181 |
MaintPasswd=%{0}
Change Password? |
182 |
Enter Maintenance Password |
183 |
Entering Parameter Configuration Mode |
184 |
Saving Password |
185 |
Entering Measurement Mode |
186 |
ConfigPasswd=%{0}
Change Password? |
187 |
Enter Configuration Password |
188 |
Disable Warning Signal? |
189 |
Disabling Analog Warning Signal |
190 |
Enable Warning Signal? |
191 |
N/A! Multidrop |
192 |
Device is in Mutidrop; Analog Warning Signal is NOT Applicable |
193 |
Device is in Mutidrop; Analog Warning Signal is NOT Applicable |
194 |
Output Set to:
%{2} %{1} |
195 |
Use Special Signal |
196 |
Fault;Warning;Maintenance |
197 |
Enable Alarms |
198 |
TestGas= %{0}
MaxConc= %{3} %{1} Enter Actual Conc |
199 |
Value Stabilized?
%{2} %0 %{1} |
200 |
Value OK?
%{2} %{0} %{1} |
201 |
Yes;No |
202 |
Not doing the calibration |
203 |
Too many write attempts |
204 |
Remove Test Gas |
205 |
Is Value out of Alarm Range?
%{2} %0 %{1} |
206 |
History Buffer
%{4} %{5} %{6} %{7} %{8} %{14}
%{9} %{10} %{11} %{12} %{13} %{15} %{16}
Reset History? |
207 |
MeasGas= %{0}
Units= %{1}
Change Gas/Units? |
208 |
ExitGasSelect |
209 |
Exiting from Gas Selection (selection mode not active) |
210 |
Gas/Units NOT changed |
211 |
TestGas= %{0}
Units= %{1}
Change Test Gas or Units? |
212 |
ExitCalGasSelect |
213 |
Exiting from Cal Gas Selection (selection mode not active) |
214 |
Cal Gas and Units NOT changed |
215 |
ExitReloadDB |
216 |
Exiting from Cal Gas Selection (Reload DB) |
217 |
Reloading Measurement Gas DB |
218 |
Upload Timeout.
Continue Upload? |
219 |
Bad DB Upload.
Retry Upload? |
220 |
Offline |
221 |
p2GasInfoArray |
222 |
Warning; Some relays still active:
Alarm 1: %{2}
Alarm 2: %{1}
ErrorRelay: %{0} |
223 |
Sensor ready in
%{0} Minutes
Calibration possible in
%{1} Minutes |
224 |
4mA;20mA;Other mA;End |
225 |
Output Set to:
%{2} %{1} |
226 |
Set Output Level Again? |
227 |
Yes;No |
228 |
Fault;Warning;Maintenance;End |
229 |
About to Reset Error History |
230 |
Error History is reset |
231 |
Warning: SEND will fail while HMI Active. Please Correct |
232 |
Reloading Measurement Gas DB |
233 |
Upload Timeout.
Continue Upload? |
234 |
Entering Measurement Mode |