名称 | ID | 标签 | 描述 | 类型 | 源代码 |
---|---|---|---|---|---|
LF220_self_test | 4058 | Self test | msrc_UNKNOWN | 详细信息 | |
LF220_press_keypad_pre_warning | 4059 | 设置回路为手动 | msrc_UNKNOWN | 详细信息 | |
LF220_press_keypad_post_warning | 405A | 注意——回路可能被返回自动控制 | msrc_UNKNOWN | 详细信息 | |
static_flow_zero | 405B | EMF zero | EMF zero- Make zero-point adjustment, with the fluid within the measuring pipe of the detector completely at a halt. | msrc_UNKNOWN | 详细信息 |
LF220_sensor_trim | 405C | Calibration | msrc_UNKNOWN | 详细信息 | |
flow_cal_0 | 4099 | FLOW CAL 0% | PV% display- PV% is displayed every 1 second. Calibration start- 0% calibration start. Exit- Exit and back to menu. | msrc_UNKNOWN | 详细信息 |
flow_cal_50 | 40C1 | FLOW CAL 50% | PV% display- PV% is displayed every 1 second. Calibration start- 50% calibration start. Exit- Exit and back to menu. | msrc_UNKNOWN | 详细信息 |
flow_cal_100 | 409A | FLOW CAL 100% | PV% display- PV% is displayed every 1 second. Calibration start- 100% calibration start. Exit- Exit and back to menu. | msrc_UNKNOWN | 详细信息 |
ex_curr_dsp | 409B | EX.CURR.DSP. | msrc_UNKNOWN | 详细信息 | |
LF220_count_rate | 409D | Count rate | Count rate- For the integral counter, the amount per count (pulse or count rate) is set. | msrc_UNKNOWN | 详细信息 |
LF220_multi_range_hysteresys | 405D | Hysteresys | Hysteresys- Set the blind sector width associated with range switching. | msrc_UNKNOWN | 详细信息 |
LF220_manual_zero | 40C2 | Manual zero | Manual Zero- Manual Zero value is setting for zero offset value that can be applied to make outputs comparable to process values measured by other instruments. | msrc_UNKNOWN | 详细信息 |
LF220_fixed_pulse | 409E | Fixed pulse | Fixed pulse. | msrc_UNKNOWN | 详细信息 |
LF220_pass_word | 40C3 | Password | Password- Security code for local programing. Enter correct password by local SW, enable parameters to change. | msrc_UNKNOWN | 详细信息 |
LF220_alarm_set | 40C4 | Alarm set (ON/OFF) | Alarm setting | msrc_UNKNOWN | 详细信息 |
LF220_D_size | 405E | Meter size | Meter size- Detector Inlet Diameter. | msrc_UNKNOWN | 详细信息 |
limit_of_change | 405F | Limit of change | Limit of change- The change rate limitation provides the function for suppressing sudden variations in converter flow rate signal output generated when too much noise is contained in the flow rate signal. | msrc_UNKNOWN | 详细信息 |
main_unit | 40C5 | UNIT1 | Display UNIT1- Engineering unit on unit1 LCD of converter to be display with flow values. | msrc_UNKNOWN | 详细信息 |
sub_unit | 40C6 | UNIT2 | Display UNIT2- Engineering unit on unit2 LCD of converter to be display with flow values. | msrc_UNKNOWN | 详细信息 |
LF220_flange_types | 409C | Flange type | msrc_UNKNOWN | 详细信息 | |
pulse_width | 4060 | Pulse width | msrc_UNKNOWN | 详细信息 | |
preset_counter | 4061 | Preset count | Preset count- Preset value of preset counter is set. | msrc_UNKNOWN | 详细信息 |
alarm_upper_value | 4062 | High alarm | High alarm value- Flow-rate high limit alarm value. | msrc_UNKNOWN | 详细信息 |
alarm_lower_value | 4063 | Low alarm | Low alarm- Flow-rate low limit alarm value. | msrc_UNKNOWN | 详细信息 |
review2_method | 4064 | Review2 | msrc_UNKNOWN | 详细信息 | |
loop_test | 1002 | 回路测试 | 回路测试——允许用户将选择的模拟量输出手动调节至选择的常数输出值。 | msrc_UNKNOWN | 详细信息 |
return_to_normal | 1016 | 中止返回正常 | msrc_UNKNOWN | 详细信息 | |
leave_fixed_current_mode | 1003 | 退出固定电流模式 | 退出固定电流模式——使选择的模拟量输出退出输出为常数值的模式,使输出返回输出跟踪输入的模式。 | msrc_UNKNOWN | 详细信息 |
dac_trim | 1004 | D/A修正 | 数字到模拟修正——允许对选择的模拟量输出进行标定,在模拟量输出的操作端点有外部参考。 | msrc_UNKNOWN | 详细信息 |
名称 | ID | 标签 | 描述 | 类型 | 源代码 |