Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
KDGM_add_100 | 4064 | Add 100 | msrc_UNKNOWN | Details | |
KDGM_subtract_100 | 4065 | Subtract 100 | msrc_UNKNOWN | Details | |
KDGM_trim_4mA | 403F | Trim 4mA | D/A トリム (デジタル/アナログ トリム) - オペレーティング エンド ポイントに対するアナログ出力と外部指示計とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
KDGM_trim_20mA | 4040 | Trim 20mA | D/A トリム (デジタル/アナログ トリム) - オペレーティング エンド ポイントに対するアナログ出力と外部指示計とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
KDGM_curve_profile_method | 409D | Set Curve profile | Curve profile - Selecting a non-linear profile will set the 10 point plot values P30 to P39 to that curve. SPECIAL allows individual point values to be changed. | msrc_UNKNOWN | Details |
automatic_method_p13 | 409E | Set PV scale factor | PV scale factor - If the curve profile is LINEAR this is the multiplier for the PV in units per metre, else it is the maximum PV value for the curve profile selected. If P13 = NaN then P51 (Displacer Length) is used | msrc_UNKNOWN | Details |
automatic_method_p15 | 409F | Set Calibrate max | Calibrate max - the PV at which the current output is 20mA and the % output is 100%. | msrc_UNKNOWN | Details |
auto_nan | 40A1 | auto text | msrc_UNKNOWN | Details | |
automatic_method_p17 | 40A2 | Set Level for max PV | Level for max PV - The height up the displacer for the maximum PV value. If set to NaN then the full length of the displacer as defined by P51 is automatically used. | msrc_UNKNOWN | Details |
automatic_method_p25 | 40A3 | Set Process Temperature | Process temperature - defines the temperature inside the vessel being measured. If set to NaN then the temperature measured by the sensor in the instrument head is used | msrc_UNKNOWN | Details |
automatic_method_p44 | 40A4 | Set Temperature | Electronics temperature - If set to a value then the temperature will be fixed, else if set to NaN the integral temperature sensor will be used. The temperature is used to correct for system temperature effects. | msrc_UNKNOWN | Details |
KDGM_output_cal | 405E | 実入力 | 実入力 - 機器のレンジ変換を Upper レンジ値と Lower レンジ値の実際の入力で行います。 | msrc_UNKNOWN | Details |
KDGM_sensor_cal | 405D | Sensor rerange | Sensor rerange - Used to redefine the place on the displacer that represents zero PV and max PV. These are by default the top and bottom of the displacer | msrc_UNKNOWN | Details |
KDGM_enter_password | 403B | If password is OPEN then non protected parameters can be changed. Enter 62 to OPEN the password. | msrc_UNKNOWN | Details | |
loop_test | 1002 | ループ テスト | ループ テスト - アナログ出力をユーザーが選択した値で固定出力します | msrc_UNKNOWN | Details |
return_to_normal | 1016 | ノーマル モードに戻します | msrc_UNKNOWN | Details | |
leave_fixed_current_mode | 1003 | 固定電流モードを終ります | 現在の固定モードを終了 - アナログ出力を定数値のモードから移行させ, 出力が入力に応じて変わるモードに戻します。 | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A トリム | D/A トリム (デジタル/アナログ トリム) - オペレーティング エンド ポイントに対するアナログ出力と外部指示計とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |