correct_inp_lrv_method | 4066 | Correct Input LRV | Correct Input LRV- Calibrates zero point (Lower Range Value) based on the current input value. | msrc_UNKNOWN | Details |
correct_inp_urv_method | 4067 | Correct Input URV | Correct Input URV- Calibrates span (Upper Range Value) based on the current input value. | msrc_UNKNOWN | Details |
reset_correct_method | 4068 | Reset Corrects | Reset Corrects- Discards calibration data. | msrc_UNKNOWN | Details |
set_lrv_actual_method | 4065 | Set LRV(Actual) | Set LRV(Actual)- Sets the Lower Range Value so that the present output value is used as the preset value. | msrc_UNKNOWN | Details |
cutoff_mode_method | 4062 | Change Cutoff Mode | Change Cutoff Mode- Selects the cutoff mode and sets the cutoff point. Valid when Square Root is selected in Transfer Function. | msrc_UNKNOWN | Details |
self_test | 4055 | Status | Status- Displays the result of self-diagnosis of the device. | msrc_UNKNOWN | Details |
clear_status_record_method | 4058 | Clear Status Records | Clear status records- Clears the Status-Record in device. | msrc_UNKNOWN | Details |
write_protect_method | 279B | Change SW Write Protect | Change Software Write Protect- Selects either 'Yes' or 'No'. | msrc_UNKNOWN | Details |
applied_zero_method | 4063 | Apply Zero Value | Apply zero value- Set the Lower Range Value at the input pressure. | msrc_UNKNOWN | Details |
applied_span_method | 4064 | Apply Span Value | Apply span value- Sets the Upper Range Value at the input pressure. | msrc_UNKNOWN | Details |
restore_factory_setting_method | 4069 | Restore Factory Setting | Restore factory setting- Return to the factory default calibration date. | msrc_UNKNOWN | Details |
open_contact_output_method | 406A | Open | Open- Simulates the contact output. Select [Open] to put the device in simulation mode, with contact output circuit is open. | msrc_UNKNOWN | Details |
closed_contact_output_method | 406B | Closed | Closed- Simulates the contact output. Select [Closed] to put the device in simulation mode, with contact output circuit is closed. | msrc_UNKNOWN | Details |
clear_contact_output_method | 406C | Clear | Clear- Simulates the contact output. Selects [Clear] to cancel simulation mode. | msrc_UNKNOWN | Details |
read_sw_version_method | 4041 | Software Version | Software Version- Displays the software version of device. | msrc_UNKNOWN | Details |
scaled_dac_trim | 1005 | Scaled D/A Trim | Digital to Analog Scaled Trim- Allows the calibration of the Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
burnout_simulation_method | 40B9 | Failure Alarm Simulation Mode | Failure Alarm Simulation Mode- Simulates the output when the device is in a critical state. | msrc_UNKNOWN | Details |
burnout_simulation_clear_method | 40BD | Clear | Clear- Simulates the output when the device is in a critical state. Select [Clear] to cancel simulation mode. | msrc_UNKNOWN | Details |
zero_trim | 1006 | Zero Trim | Zero trim- Make the sensor input the new zero input reference | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
transmitter_loop_test | 1023 | Loop Test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
transmitter_dac_trim | 1024 | D/A Trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
burn_eeprom | 100C | Masterizza EEPROM | Burn Electrically Erasable Read Only Memory- Operation where the Field Device will copy what is in its Random Access Memory into the Electrically Erasable Read Only Memory, thus making changes by the Host permanent | msrc_UNKNOWN | Details |