0 |
82TC CONC |
1 |
Toroidal Concentration Transmitter |
2 |
ABB Instrumentation |
3 |
EC1 HI.PV.AD |
4 |
Process Variable A/D over range |
5 |
EC2 LO.PV.AD |
6 |
Process Variable A/D under range |
7 |
EC3 PH.PCB |
8 |
pH/ORP/pION input PCB with Conductivity/concentration firmware |
9 |
EC6 TE.PCB |
10 |
2-electrode conductivity input PCB with Toroidal conductivity/concentration firmware |
11 |
EC7 EC.PCB |
12 |
Electrode conductivity input PCB with Toroidal conductivity/concentration firmware |
13 |
PC6 HI.LOOP |
14 |
Current loop above upper range |
15 |
PC7 LO.LOOP |
16 |
Current loop below lower range value |
17 |
PC8 HI.PV |
18 |
PV above transmitter range |
19 |
PC9 LO.PV |
20 |
PV below transmitter range |
21 |
PC10 HI.TEMP |
22 |
Temperature above transmitter range |
23 |
PC11 LO.TEMP |
24 |
Temperature below transmitter range |
25 |
PC12 HI.T.AD |
26 |
Open or missing Temperature Sensor |
27 |
PC13 LO.T.AD |
28 |
Shorted Temperature Sensor |
29 |
PC2O BAD.SEE |
30 |
Bad SEEPROM or bad input PCB assembly |
31 |
PC21 NO.F.CAL |
32 |
Missing factory cal or nonfunctional SEEPROM |
33 |
PC22 BLNK.EE |
34 |
Blank microprocessor EEPROM |
35 |
PC25 ROM.SUM |
36 |
Incorrect EPROM checksum |
37 |
PC30 R0.F.CAL |
38 |
Out of range or missing factory cal for conductivity circuit range zero |
39 |
PC31 R1.F.CAL |
40 |
Out of range or missing factory cal for conductivity circuit range one |
41 |
PC32 R2.F.CAL |
42 |
Out of range or missing factory cal for conductivity circuit range two |
43 |
PC33 R3.F.CAL |
44 |
Out of range or missing factory cal for conductivity circuit range three |
45 |
PC34 R4.F.CAL |
46 |
Out of range or missing factory cal for conductivity circuit range four |
47 |
PC45 BA.F.CAL |
48 |
Out of range or missing factory cal for 3-k ohm Balco temp.compensator |
49 |
PC46 PT.F.CAL |
50 |
Out of range or missing factory cal for PT100 temp. compensator |
51 |
PC48 PK.F.CAL |
52 |
Out of range or missing factory cal for PT1000 temp. compensator |
53 |
PC50 R0.CHKS |
54 |
Incorrect or missing conductivity circuit range zero checksum |
55 |
PC51 R1.CHKS |
56 |
Incorrect or missing conductivity circuit range one checksum |
57 |
PC52 R2.CHKS |
58 |
Incorrect or missing conductivity circuit range two checksum |
59 |
PC53 R3.CHKS |
60 |
Incorrect or missing conductivity circuit range three checksum |
61 |
PC54 R4.CHKS |
62 |
Incorrect or missing conductivity circuit range four checksum |
63 |
PC65 BA.CHKS |
64 |
Incorrect or missing 3-k ohm Balco temp. compensator checksum |
65 |
PC66 PT.CHKS |
66 |
Incorrect or missing PT100 temp. compensator checksum |
67 |
PC68 PK.CHKS |
68 |
Incorrect or missing PT1000 temp. compensator checksum |
69 |
PV was unstable during PV cal |
70 |
PV value was unstable during PV calibration |
71 |
PV was drifting during PV cal |
72 |
PV value was drifting during PV calibration |
73 |
PV LO OFFSET |
74 |
PV calibration Offset too low |
75 |
PV LO SLOPE |
76 |
PV calibration Slope too low |
77 |
PV HI OFFSET |
78 |
PV calibration Offset too high |
79 |
PV HI SLOPE |
80 |
PV calibration Slope too high |
81 |
Temp Offset Warning |
82 |
Temperature calibration Offset unacceptable |
83 |
Temp Slope Warning |
84 |
Temperature calibration Slope unacceptable |
85 |
In PV Cal mode |
86 |
A PV zero or span point calibration is being performed on the device |
87 |
PV eng. unit |
88 |
ppb |
89 |
Parts per Billion |
90 |
|
91 |
Concentration - Field Device Variable Digital value that changes magnitude with respect to time with no user intervention. |
92 |
.3f |
93 |
.2f |
94 |
8.2f |
95 |
11.3f |
96 |
8u |
97 |
Temp Unit |
98 |
Temperature- In Auto temp comp mode, it is the Temp read with the temp sensor. In Manual temp comp mode, it is the user entered temp value |
99 |
In calibration security mode |
100 |
Transmitter Busy - Cannot Process Command |
101 |
10.3f |
102 |
PV slope |
103 |
PV sensor slope value |
104 |
PV offset |
105 |
PV sensor offset value. |
106 |
Temperature sensor Slope |
107 |
Temperature sensor offset |
108 |
Temperature sensor offset value. |
109 |
Span point value |
110 |
Span calibration point value for PV calibration |
111 |
Temp. calibration value |
112 |
Temperature calibration value |
113 |
Zero point value |
114 |
Zero calibration point value |
115 |
Concentration cal |
116 |
HART Parameters |
117 |
Offline Configuration Parameters |
118 |
PV Range Setup |
119 |
Reset Transmitter |
120 |
Reset Transmitter - Instructs the transmitter to generate a reset. The result is similar to the effects of briefly interrupting power to the transmitter. This can be used to recover from an error condition if needed. |
121 |
Zero point cal |
122 |
Calibration of the process sensor input via a Zero-Point offset calibration |
123 |
Span point cal |
124 |
Calibration of the process sensor input via a Span-Point slope calibration |
125 |
Read cal values |
126 |
Temp. Cal |
127 |
Calibration of Temperature sensor input via one-point smart calibration that adjusts offset and slope |
128 |
Edit Cal |
129 |
Manual adjustment of process sensor and temperature offset and slope values |
130 |
Reset Cal |
131 |
Reset the process and temperature sensor calibration data to the values set at the factory |
132 |
Refresh variables |
133 |
Read transmitter variables to refresh their values. This will update any variable that was modified via the device keypad. |
134 |
FMU860 |
135 |
DT |
136 |
DVC |
137 |
FCX |
138 |
AS800 |
139 |
340T |
140 |
340C |
141 |
344 |
142 |
Siemens Micro K |
143 |
SITRANS L |
144 |
SIPAN PH |
145 |
SITRANS P |
146 |
75SMT |
147 |
UNIDELTA |
148 |
8320 |
149 |
SX5100 |
150 |
TSV175 |
151 |
DMU |
152 |
3780 |
153 |
BM70 |
154 |
low |
155 |
high |
156 |
tmp_av |
157 |
Transmitter will be Reset! Continue? |
158 |
Transmitter Reset is in process, please wait. |
159 |
Reading current values of all transmitter variables, please wait. |
160 |
Transmitter Reset successful |
161 |
Access denied
PV cal in progress |
162 |
Zero cal point:
%{4} %{3}
Press OK to continue |
163 |
Press OK when PV is stable
%0 %{3} |
164 |
Waiting for calibration |
165 |
PV was DRIFTING during PV cal |
166 |
PV was UNSTABLE during PV cal |
167 |
Calculated
Slope:%{1}
Offset:%{2} |
168 |
REJECT;ACCEPT |
169 |
PV cal completed
Slope: %{1}
Offset: %{2} |
170 |
Calibration was not successful. Do you want to retry? |
171 |
Enter Span cal point |
172 |
Enter Temperature: |
173 |
TEMP OFFSET WARNING |
174 |
TEMP SLOPE WARNING |
175 |
Edit PV Slope |
176 |
Edit PV Offset |
177 |
Edit Temp. Slope |
178 |
Edit Temp. Offset |
179 |
Press OK to Reset Calibration |
180 |
Reading transmitter variables, please wait. |