Loading…
Name ID Label Description Type Access Class
custom_unit_mult 4044 Custom Unit Multiplier Multiplier used to calculate the custom unit output from the level measurement. R/W Float 2048
Algorithm 4045 Target Algorithm The method used to select the echo pulse used for the level measurement. Largest selects the largest echo. First selects the first echo above the detection threshold. Height/Distance selects the echo with the largest height divided by distance ratio. R/W Enum 2048
speed_of_sound 4046 Speed of Sound Speed of sound at 0 degrees C used to calculate the level distance. R/W Float 2048
echo_list_mode 4047 Echo List Mode Selects whether the echo list is shown in terms of distance or level. R/W Enum 2048
echo_missing 4048 Echo Missing Length of time echos have been missing. R Unsigned 2051
echo_strength 4049 Echo Strength Relative strength of the echo used to calculate level. R Unsigned 2051
num_echos 404A Number of Echos Number of echos detected. R Unsigned 2049
echo_rejection_status 404B Echo Rejection Status R/W Enum 2050
echo_profile_maturity 404C Echo Profile Maturity Percentage of echo profile obtained. R Unsigned 2051
echo_list_length 404D Echo List Length Number of echos in echo profile list. R Unsigned 2048
Name ID Label Description Type Access Class