名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
C4W_C_ArrayElement_6_1 | 4022 | C5 | Displays the current coefficients for density. | 读 | 浮点 | 2048 |
C4W_CalibrationFactor_1 | 4023 | Calibration factor | Displays the current calibration factor for the flow rate measuring sensor. Additional information: The factory setting for the calibration factor can be found on the sensor's nameplate. | 读 | 浮点 | 2048 |
C4W_Damping_ArrayElement_1_1 | 4024 | Oscillation damping 0 | Displays the current oscillation damping. Oscillation damping is an indicator for the sensor's current demand for excitation power. | 读 | 浮点 | 2 |
C4W_Damping_ArrayElement_2_1 | 4025 | Oscillation damping 1 | Displays the current oscillation damping. Oscillation damping is an indicator for the sensor's current demand for excitation power. | 读 | 浮点 | 2 |
C4W_DampingMax_1 | 4026 | Maximum damping partial filled pipe det. | 读/写 | 浮点 | 2048 | |
C4W_DensityAdjustMode_1 | 4027 | Density adjustment mode | Select the density adjustment mode. Additional information: Before performing the density adjustment, the following conditions must be met: - The density adjustment must be performed under operating conditions and there must be little variation in the operating conditions. - The reference media must be gas-free or pressurized so that any gas they contain is compressed. - The reference density measurements must be performed at the temperature of the medium in the process. Otherwise the density adjustment will be inaccurate. - The correction made can be undone by selecting the "Restore original" option in the "Execute density adjustment" parameter. | 读 | 枚举 | 2048 |
C4W_DensityAdjustSetValue1_1 | 4028 | Density setpoint 1 | Enter the density setpoint (1 point adjustment) or the first density setpoint (2 point adjustment). | 读 | 浮点 | 2048 |
C4W_DensityAdjustSetValue2_1 | 4029 | Density setpoint 2 | Enter the second density setpoint. | 读 | 浮点 | 2048 |
C4W_DensityAdjustState_1 | 402A | Density adjustment state | Shows the current state of the density adjustment | 读 | 枚举 | 2048 |
C4W_DualModeActive_1 | 402B | Multi-frequency activation | Switch the dual mode of the sensor on and off | 读 | 枚举 | 2048 |