Variable Name: C4W_DensityAdjustMode_1
Variable ID: 0x4027
Variable Label: Density adjustment mode
Variable Help: Select the density adjustment mode. Additional information: Before performing the density adjustment, the following conditions must be met: - The density adjustment must be performed under operating conditions and there must be little variation in the operating conditions. - The reference media must be gas-free or pressurized so that any gas they contain is compressed. - The reference density measurements must be performed at the temperature of the medium in the process. Otherwise the density adjustment will be inaccurate. - The correction made can be undone by selecting the "Restore original" option in the "Execute density adjustment" parameter.
Variable Handling: R
Variable Class: Device
Variable Size: 2
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x00EE | 1 point adjustment |
0x00EF | 2 point adjustment |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
173 | 42E8 | read_index_parameter_173 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
178 | 42E6 | write_index_parameter_178 |