Loading…

Variable Name: C4W_DensityAdjustMode_1

Variable ID: 0x4027

Variable Label: Density adjustment mode

Variable Help: Select the density adjustment mode. Additional information: Before performing the density adjustment, the following conditions must be met: - The density adjustment must be performed under operating conditions and there must be little variation in the operating conditions. - The reference media must be gas-free or pressurized so that any gas they contain is compressed. - The reference density measurements must be performed at the temperature of the medium in the process. Otherwise the density adjustment will be inaccurate. - The correction made can be undone by selecting the "Restore original" option in the "Execute density adjustment" parameter.

Variable Handling: R

Variable Class: Device

Variable Size: 2

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x00EE 1 point adjustment
0x00EF 2 point adjustment

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
173 42E8 read_index_parameter_173

Variable Write Command List

Command Number Command ID Command Name
178 42E6 write_index_parameter_178