Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
dDensityMinimumSpan | 4059 | Minimum Span | Valor de span mínimo: Define la menor diferencia permitida entre el valor de rango superior y el valor de rango inferior. | R | Float | 8 |
dDensityFrequency | 4233 | Frequency | Value of variable given in frequency output | R | Float | 514 |
dDensityFreqRange | 4234 | Density Frequency Limit | Pulse frequency at maximum flow rate | R/W | Float | 512 |
dDensityFreqPercent | 4235 | Frequency Percent | Percent of maximum flow rate | R | Float | 34 |
sluglo | 405D | Slug Low Limit | A programmable low limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. | R/W | Float | 8 |
slughi | 405E | Slug High Limit | A programmable high limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. | R/W | Float | 8 |
slug_delay_time | 4125 | Slug Delay Time | The time delay between the density exceeding the slug-flow thresholds and the slug-flow error being declared. | R/W | Float | 8 |
cTemperatureUnits | 4061 | Temperature Units | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit | R/W | Enum | 32 |
dTemperatureValue | 4062 | Temperature | Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention | R | Float | 34 |
dTemperatureUpperSensorLimit | 4063 | Temp URL | Valor límite superior del sensor: Define el valor máximo utilizable para el valor de rango superior. | R | Float | 8 |