Variable Name: PFS_PulseWidth_1

Variable ID: 0x4185

Variable Label: Pulse width

Variable Help: Specify the duration of the output pulse. Additional information: The maximum pulse rate is defined by fmax = 1 / (2 × pulse width). The interval between two pulses (P) is at least as long as the specified pulse width (B). The maximum flow is defined by Qmax = fmax × pulse value. If the flow exceeds these limit values, the measuring device displays the diagnostic message "443 Pulse output @1 faulty". Example: - Pulse value: 0.1 g - Pulse width: 0.1 ms - fmax: 1 / (2 × 0.1 ms) = 5 kHz - Qmax: 5 kHz × 0.1 g = 0.5 kg/s

Variable Handling: R/W

Variable Class:

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|.2f

Variable Edit Format: |en|.2f

Variable Unit:

Variable Max: 2000

Variable Min: 0.05

Variable Read Command List

Command Number Command ID Command Name
173 42E8 read_index_parameter_173

Variable Write Command List

Command Number Command ID Command Name
178 42E6 write_index_parameter_178