Variable Name: PFS_PulseWidth_1
Variable ID: 0x4185
Variable Label: Pulse width
Variable Help: Specify the duration of the output pulse. Additional information: The maximum pulse rate is defined by fmax = 1 / (2 × pulse width). The interval between two pulses (P) is at least as long as the specified pulse width (B). The maximum flow is defined by Qmax = fmax × pulse value. If the flow exceeds these limit values, the measuring device displays the diagnostic message "443 Pulse output @1 faulty". Example: - Pulse value: 0.1 g - Pulse width: 0.1 ms - fmax: 1 / (2 × 0.1 ms) = 5 kHz - Qmax: 5 kHz × 0.1 g = 0.5 kg/s
Variable Handling: R/W
Variable Class:
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|.2f
Variable Edit Format: |en|.2f
Variable Unit:
Variable Max: 2000
Variable Min: 0.05
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
173 | 42E8 | read_index_parameter_173 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
178 | 42E6 | write_index_parameter_178 |