Variable Name: FLOW_PressureShockSuppression_1
Variable ID: 0x40B2
Variable Label: Pressure shock suppression
Variable Help: Enter time frame for signal suppression (= pressure shock suppression active), for example to prevent the device from registering flow movements in the pipe when a valve is closed. Additional information: Pressure schock suppression is activated when the flow rate drops below the on value for low flow cutoff. Output values when pressure shock suppression is active: - Flow: 0 - Totalizer: Last valid value Pressure shock suppression is deactivated when: - The time frame specified has elapsed and - Flow exceeds the off value for low flow cutoff
Variable Handling: R/W
Variable Class: Device
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|.1f
Variable Edit Format: |en|.1f
Variable Unit:
Variable Max: 100
Variable Min: 0
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
173 | 42E8 | read_index_parameter_173 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
178 | 42E6 | write_index_parameter_178 |