Loading…

Variable Name: FLOW_PressureShockSuppression_1

Variable ID: 0x40B2

Variable Label: Pressure shock suppression

Variable Help: Enter time frame for signal suppression (= pressure shock suppression active), for example to prevent the device from registering flow movements in the pipe when a valve is closed. Additional information: Pressure schock suppression is activated when the flow rate drops below the on value for low flow cutoff. Output values when pressure shock suppression is active: - Flow: 0 - Totalizer: Last valid value Pressure shock suppression is deactivated when: - The time frame specified has elapsed and - Flow exceeds the off value for low flow cutoff

Variable Handling: R/W

Variable Class: Device

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|.1f

Variable Edit Format: |en|.1f

Variable Unit:

Variable Max: 100

Variable Min: 0

Variable Read Command List

Command Number Command ID Command Name
173 42E8 read_index_parameter_173

Variable Write Command List

Command Number Command ID Command Name
178 42E6 write_index_parameter_178