Loading…

Variable Name: CO_OperatingMode_1

Variable ID: 0x4059

Variable Label: Current range output

Variable Help: Select the current range for the measured value output and the upper and lower fault condition signal level. Additional information: - The measured value range is specified in the "Lower range value output @1" parameter and the "Upper range value output @1" parameter. - If the measured value lies outside the scaled measured value range, diagnostic message "441 Current output @1 faulty" is generated. - In the event of a device alarm, the current output adopts the behavior specified in the "Failure behavior current output @1" parameter.

Variable Handling: R/W

Variable Class: Device

Variable Size: 2

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x800C 4...20 mA NE (3.8...20.5 mA)
0x800D 4...20 mA US (3.9...20.8 mA)
0x800A 4...20 mA (4... 20.5 mA)
0x8088 Fixed value

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
173 42E8 read_index_parameter_173

Variable Write Command List

Command Number Command ID Command Name
178 42E6 write_index_parameter_178