Loading…

Variable Name: device_specific_status_17

Variable ID: 0x1034

Variable Label: ステータス グループ 17

Variable Help: フィールド機器スペシフィック ステータス バイト 18 - 故障/警告/プロセスのステータスに関連するフィールド機器情報のデータです。

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 F100 Sensor error -> Remedy information: 1. Restart the device 2. Contact Endress+Hauser Service
0x0002 F101 Singleline text -> Remedy information: 1. Check process temperature 2. Check ambient temperature
0x0004 500 Process alert pressure -> Remedy information: 1. Check process pressure 2. Check configuration of process alert
0x0008 501 Process alert scaled variable -> Remedy information: 1. Check process conditions 2. Check scaled variable configuration
0x0010 502 Process alert temperature -> Remedy information: 1. Check process temperature 2. Check configuration of process alert
0x0020 900 High signal noise detected -> Remedy information: 1. Check impulse line 2. Check valve position 3. Check process
0x0040 901 Low signal noise detected -> Remedy information: 1. Check impulse line 2. Check valve position 3. Check process
0x0080 902 Min signal noise detected -> Remedy information: 1. Check impulse line 2. Check valve position 3. Check process

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None