Loading…

Variable Name: PRES_AlertScaledVariableBehaviour_1

Variable ID: 0x40CD

Variable Label: Diagnostic behavior

Variable Help: Select event behavior 'Logbook entry only': no digital or analog transmission of the message 'Warning': Current output unchanged. Message is output digitally (default). 'Alarm': Current output assumes the set alarm current. Regardless of the setting, the message appears on the display. If the permissible conditions are reached again, the warning is no longer available in the instrument.

Variable Handling: R/W

Variable Class: Device

Variable Size: 2

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x8213 Alarm
0x8211 Warning
0x00B9 Logbook entry only

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
173 4324 read_index_parameter_173

Variable Write Command List

Command Number Command ID Command Name
178 4322 write_index_parameter_178