Variable Name: START_STOP_CALIBRATION
Variable ID: 0x407A
Variable Label: Start/Stop Calibration
Variable Help: Parameters serve to calibrate the gas detector via the Modbus RTU interface. A detailed description of the parameters and the calibration procedure is available from your local Draeger service organization.
Variable Handling: R/W
Variable Class: Dynamic | Correction
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | Start of the zero adjustment |
0x0001 | Start of the span calibration |
0x0002 | Measurement of the calibration gas concentration |
0x0003 | Measurement of the target gas concentration |
0x0004 | Save the calibration |
0x0005 | Exit calibration mode without saving |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
None
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
149 | 407C | start_perform_stop_calibration |