Loading…

Variable Name: START_STOP_CALIBRATION

Variable ID: 0x407A

Variable Label: Start/Stop Calibration

Variable Help: Parameters serve to calibrate the gas detector via the Modbus RTU interface. A detailed description of the parameters and the calibration procedure is available from your local Draeger service organization.

Variable Handling: R/W

Variable Class: Dynamic | Correction

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 Start of the zero adjustment
0x0001 Start of the span calibration
0x0002 Measurement of the calibration gas concentration
0x0003 Measurement of the target gas concentration
0x0004 Save the calibration
0x0005 Exit calibration mode without saving

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

None

Variable Write Command List

Command Number Command ID Command Name
149 407C start_perform_stop_calibration