Variable Name: var_ds_fCurrentUpperLimit

Variable ID: 0x4045

Variable Label: Current Upper Limit

Variable Help: Current Upper Limit Upper limit of the current range for the 4-20 mA current output. For adjustment purposes or to safely identify an over or under measuring range condition, the current range available for the measurement signal is extended beyond the standard range of 4 mA and 20mA. According to Namur specifications (NE 43), the upper current value must not be higher than 20.5mA. The limits can be set within the range from 20 mA to 20.5 mA.

Variable Handling: R/W

Variable Class: AnalogOut

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|.2f

Variable Edit Format: |en|.2f

Variable Unit:

Variable Max:

Variable Min:

Variable Read Command List

Command Number Command ID Command Name
178 4214 cmd_rd_loop_limits

Variable Write Command List

Command Number Command ID Command Name
179 4215 cmd_wr_loop_limits