Loading…

Method Name: leopold_lagco_flume_method_max_flowrate

Method ID: 0x41FF

Method Label: Calculate Max Flowrate

Method Help: Calculates Maximum Flowrate (Qmax) value


  • {
  • float flowratecalculated;
  • float headconvertedleopold;
  • char disp_string1[135];
  • int slen1;
  • slen1 = 135;
  • _set_device_status(0xFF,0);
  • _set_xmtr_device_status(0xFF,0);
  • _set_all_resp_code(0);
  • _set_xmtr_all_resp_code(0);
  • _get_dictionary_string(19726337,disp_string1,slen1);
  • ACKNOWLEDGE(disp_string1);
  • _add_abort_method(4118);
  • if (Distance_Obj_unit == 44)
  • {
  • headconvertedleopold = span_Obj_value_0 * 0.3048;
  • }
  • else
  • if (Distance_Obj_unit == 45)
  • {
  • headconvertedleopold = span_Obj_value_0 * 1;
  • }
  • else
  • if (Distance_Obj_unit == 47)
  • {
  • headconvertedleopold = span_Obj_value_0 * 0.0254;
  • }
  • else
  • if (Distance_Obj_unit == 48)
  • {
  • headconvertedleopold = span_Obj_value_0 * 0.01;
  • }
  • else
  • if (Distance_Obj_unit == 49)
  • {
  • headconvertedleopold = span_Obj_value_0 * 0.001;
  • }
  • if (flume_diameter_ft == 1 || flume_diameter_mm == 1 || flume_diameter_cm == 1 || flume_diameter_in == 1 || flume_diameter_meter == 1)
  • {
  • flowratecalculated = 221.25 * pow(headconvertedleopold,1.547);
  • }


Please login for more