Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
LF600_self_test | 411E | Self test | Self test of conveter | msrc_UNKNOWN | Details |
LF600_loop_test | 411F | Fixed pls out | Fixed pulse test; Fixed pulse output. When you use fixed pulse output function, set digital output function to PULSE OUTPUT. | msrc_UNKNOWN | Details |
LF600_cnt_ctrl | 4120 | Cnt ctrl | Counter control; Counter START or STOP, RESET | msrc_UNKNOWN | Details |
static_flow_zero | 405B | Zero adj | Still water zero adjustment is possible only when the flow rate value is within the range of +/-1.25m/s. | msrc_UNKNOWN | Details |
flow_cal_0 | 4099 | Flow cal 0% | Using the converter's internal calibration circuit, 0% flow rate calibration can be performed. | msrc_UNKNOWN | Details |
flow_cal_50 | 40C1 | Flow cal 50% | Using the converter's internal calibration circuit, 50% flow rate calibration can be read. | msrc_UNKNOWN | Details |
flow_cal_100 | 409A | Flow cal 100% | Using the converter's internal calibration circuit, 100% flow rate calibration can be performed. | msrc_UNKNOWN | Details |
ex_curr_dsp | 409B | Exc curr dspl | You can monitor the exciting current value. | msrc_UNKNOWN | Details |
LF600_D_size | 4121 | Meter size | Meter size of detector | msrc_UNKNOWN | Details |
LF600_flange_types | 4122 | Flange type | msrc_UNKNOWN | Details | |
LF600_pass_word | 4123 | Password | The password function is provided to protest the settings and adjustment of the functions affecting the flow rate measurement. | msrc_UNKNOWN | Details |
limit_of_change | 405F | Limit of change | The rate-of-change limit is used to control sudden changes of the converter's flow rate signal output when excessive noise is contained in the flow rate signal. The rate-of-change limit (set in percent value to the span of measuring range) and control limit time (set in second) are used, and if the flow rate signal sampling value exceeds the rate-of-change limit value based on the previous average value of the flow rate signal, the converter rejects the sampling value and outputs the average value including the maximum value of the rate-of-change value. Rate-of-change limit 0 to 30% / 50ms (in increments of 0.1%) Control limit time 0 to 20s (in increments of 1s) | msrc_UNKNOWN | Details |
LF600_manual_zero | 4124 | Manual zero | Manual zero adjustment value can be set within the range equivalent to +/-1m/s (+/-10% of the maximum range 10m/s). | msrc_UNKNOWN | Details |
preset_counter | 4061 | Preset value | Preset count can be set within the range of 0 to 99999999. | msrc_UNKNOWN | Details |
LF600_alarm_set | 4125 | Alarm set1 | Setting of high alarm or low alarm, empty alarm | msrc_UNKNOWN | Details |
LF600_alarm_set2 | 4126 | Alarm set2 | Setting of high/high(HH) alarm or low/low(LL) alarm | msrc_UNKNOWN | Details |
alarm_upper_value | 4062 | High alarm | High limit of the flow rate, at which an alarm is generated, can be set as % value of the span flow rate of the set maximum range. High alarm values for flow rate can be set within the range of -10% to 110% (percentage to Range 1) in increments of 0.1%. | msrc_UNKNOWN | Details |
alarm_lower_value | 4063 | Low alarm | Low limit of the flow rate, at which an alarm is generated, can be set as % value of the span flow rate of the set maximum range. Low alarm values for flow rate can be set within the range of -10% to 110% (percentage to Range 1) in increments of 0.1%. | msrc_UNKNOWN | Details |
alarm_upper_value2 | 4127 | HH alarm | High/high limit of the flow rate, at which an alarm is generated, can be set as % value of the span flow rate of the set maximum range. High/high alarm values for flow rate can be set within the range of -10% to 110% (percentage to Range 1) in increments of 0.1%. | msrc_UNKNOWN | Details |
alarm_lower_value2 | 4128 | LL alarm | Low/low limit of the flow rate, at which an alarm is generated, can be set as % value of the span flow rate of the set maximum range. Low/low alarm values for flow rate can be set within the range of -10% to 110% (percentage to Range 1) in increments of 0.1%. | msrc_UNKNOWN | Details |
LF600_press_keypad_pre_warning | 4129 | Set loop to manual | msrc_UNKNOWN | Details | |
LF600_press_keypad_post_warning | 412A | NOTE-Loop may be returned to automatic control | msrc_UNKNOWN | Details | |
LF600_multi_range_hysteresys | 412B | Range hys | Range hysteresis is the dead band used when multiple ranges are switched. You can set the hysteresis within the range of 0 to 25% in increments of 0.1%. | msrc_UNKNOWN | Details |
LF600_count_rate | 412C | Count rate | Count rate must be set so that the pulse output at 100% output is within the range below: 3.6 to 36000000 pulse/h (0.001 to 10000 pulse/s). | msrc_UNKNOWN | Details |
pulse_width | 4060 | Pulse width | The pulse width must be set to a value within the range of 0.3ms to 500ms. The pulse width must be set to 40% or less of the period of pulse frequency at 100% output. If the pulse width is set to 0, it will be automatically set to 40% of the period of pulse frequency at 100% output. In this case, the pulse width setting mode remains in the Manual mode. If the calculation result exceeds 100ms, it will be forcibly set to 100ms. | msrc_UNKNOWN | Details |
main_unit | 40C5 | Main unit | Display main unit - Main unit on LCD of converter. | msrc_UNKNOWN | Details |
sub_unit | 40C6 | Sub unit | Display sub unit - Sub unit on LCD of converter. | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
transmitter_loop_test | 1023 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
transmitter_dac_trim | 1024 | D/A trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |