Loading…
Loading…
Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
InputScalingAbortMethod | 4048 | msrc_UNKNOWN | Details | ||
self_test | 4049 | Status | Self Test- Directs the THZ2/TDZ2/TCM electronics to perform a selftest. Reports any errors that occured. | msrc_UNKNOWN | Details |
sensor_type_method | 404A | Sensor Type | msrc_UNKNOWN | Details | |
give_unit_notice | 401F | msrc_UNKNOWN | Details | ||
SelectUnitsMethod | 404B | EGU Selection | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit | msrc_UNKNOWN | Details |
OutputTrimResetMethod | 404C | Output Trim Reset | User can reset output trim to factory trim. | msrc_UNKNOWN | Details |
device_variable_trim | 1080 | Sensor Trim | Trim variabile dispositivo - Consente di regolare linearmente la misura di taratura di una variabile dispositivo. Il punto di trim che viene inviato al dispositivo deve rappresentare il valore della variabile di processo attualmente applicata. | msrc_UNKNOWN | Details |
device_variable_trim_reset | 1083 | Sensor Trim Reset | Reset trim variabile dispositivo - Consente di impostare la variabile dispositivo sul suo trim di fabbrica predefinito. | msrc_UNKNOWN | Details |
loop_test | 1002 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
dac_trim | 1004 | Output Trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
applied_rerange | 1007 | Input Capture | Applied values- Perform a device rerange by applying the desired Upper Range Value and Lower Range Values to the device input | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |