Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
mtdPreReadChangeCounter | 4024 | Changes counter | Number of changes on the field device that where done by the user | msrc_UNKNOWN | Details |
self_test | 4006 | Selftest | Limited self test- Checks the device about: PV out of Limits, Analog output saturated, Fixed current output, Fault in electronics, invalid sensor code number, sensor code number do not match, measurement bridge malfunction, Error in temperature measurement, malfunction in memory area, error in program flow. | msrc_UNKNOWN | Details |
master_reset | 4007 | Masterreset | Master Reset- Forces the field device to restart. | msrc_UNKNOWN | Details |
factory_data_reset | 4008 | Factorydata reset | factory data reset- Every user changeable variable will be reseted to ist factory data value. | msrc_UNKNOWN | Details |
mtdClearPvMinMax | 404E | Clear PV min max | Reset minimum maximum tracking values. | msrc_UNKNOWN | Details |
mtdClearTempMinMax | 404F | Clear T min max | Reset temperature minimum maximum tracking values. | msrc_UNKNOWN | Details |
mtdLowerRangeTrim | 4050 | Trim lrv | 圧力センサー Lower トリム - 圧力センサーをキャリブレートします。 | msrc_UNKNOWN | Details |
mtdUpperRangeTrim | 4051 | Trim urv | 圧力センサー Upper トリム - 圧力センサーをキャリブレートします。 | msrc_UNKNOWN | Details |
mtdErrorCorrection | 4011 | Correct Error! | to be defined... | msrc_UNKNOWN | Details |
applied_rerange | 1007 | 実入力 | 実入力 - 機器のレンジ変換を Upper レンジ値と Lower レンジ値の実際の入力で行います。 | msrc_UNKNOWN | Details |
loop_test | 1002 | ループ テスト | ループ テスト - アナログ出力をユーザーが選択した値で固定出力します | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | 固定電流モードを終ります | 現在の固定モードを終了 - アナログ出力を定数値のモードから移行させ, 出力が入力に応じて変わるモードに戻します。 | msrc_UNKNOWN | Details |
return_to_normal | 1016 | ノーマル モードに戻します | msrc_UNKNOWN | Details | |
dac_trim | 1004 | D/A トリム | D/A トリム (デジタル/アナログ トリム) - オペレーティング エンド ポイントに対するアナログ出力と外部指示計とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |