Loading…
Loading…
Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
show_calibration_points_method | 40D3 | Show Cal.Data | Shows Calibration DataSet | msrc_UNKNOWN | Details |
do_calibration_method | 40D4 | Calibrate | Runs Calibration | msrc_UNKNOWN | Details |
enter_cal_points_method | 40D5 | Enter Cal.Data | Enter Calibration DataSet | msrc_UNKNOWN | Details |
enter_background_method | 40D6 | Set Background | The Background count rate must be measured only when the probe is a plastic detctor. The background is the natural radiation only. The background rate has to be measured before the source is installed. | msrc_UNKNOWN | Details |
error_log_method | 40D7 | Show Error Log | Shows the last occured errors. | msrc_UNKNOWN | Details |
change_password_method | 40D8 | Change password | This is a availability to set up max. 3 password levels. The password for the full access level includes the access to configuration and calibration too. Each password can be different or equal to another password Password for access Read-Only can't be changed. | msrc_UNKNOWN | Details |
enter_password_method | 40D9 | Enter Access Password | In order to get access to other level, the right password must be entered. | msrc_UNKNOWN | Details |
abort_restore_factory_setting_method | 4160 | msrc_UNKNOWN | Details | ||
restore_factory_settings_method | 40DA | Factory Settings | Factory Reset will delete all settings and restore the factory default values. | msrc_UNKNOWN | Details |
restore_factory_settings_method_part2 | 40DB | Factory Settings | msrc_UNKNOWN | Details | |
save_cal_curve_as_basic_curve_method | 40DC | Save | Save the current calibration data in a separate register. E.g. a Basic Calibration. | msrc_UNKNOWN | Details |
load_basic_curve_method | 40DD | Load | Load a previous stored calibration, and overwrite the current one. | msrc_UNKNOWN | Details |
abort_sensor_reset_method | 4161 | msrc_UNKNOWN | Details | ||
sensor_reset_method | 40DE | Hardware Reset | Hardware Reset will restart the CPU without loosing data. | msrc_UNKNOWN | Details |
post_edit_digital_output_method | 40E0 | msrc_UNKNOWN | Details | ||
digital_output_test_method | 412D | Dig Output Test | msrc_UNKNOWN | Details | |
plateau_measurement_method | 40E1 | Measure Plateau | Measures Plateau -Be aware that Plateau measurement is only possible when the density is constant during this time. | msrc_UNKNOWN | Details |
show_plateau_points_method | 40E2 | Show Plateau | Shows Detector Plateau | msrc_UNKNOWN | Details |
reset_mass_integrator_method | 40E3 | Reset Totalizer | Resets the Mass Totalizer | msrc_UNKNOWN | Details |
change_meas_mode_method | 40E4 | Change Meas.Mode | Changes the actual Measurement Mode | msrc_UNKNOWN | Details |
trim_dac_in_1_method | 40E5 | Adjust CIn 1 | Adjusts CurrentIn 1 - Allows the user to adjust Current Input1 | msrc_UNKNOWN | Details |
trim_dac_in_2_method | 40E6 | Adjust CIn2 | Adjusts CurrentInput 2 - Allows the user to adjust Current Input 2 | msrc_UNKNOWN | Details |
trim_pt100_method | 40E7 | Adjust Pt 100 | Runs Pt 100 adjustment | msrc_UNKNOWN | Details |
refresh_dig_out_parameter_method | 414F | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_dig_out_parameter_method | 4162 | Updating......... | msrc_UNKNOWN | Details | |
refresh_dig_in_parameter_method | 414E | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_dig_in_parameter_method | 4163 | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_temp_compensation_mode_method | 4147 | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_damping_mode_method | 4146 | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_interference_detect_ctrl_method | 4150 | Updating......... | msrc_UNKNOWN | Details | |
refresh_var_loop_current_error_method | 4135 | Updating......... | msrc_UNKNOWN | Details | |
refresh_var_loop_current_error_2_method | 413B | Updating......... | msrc_UNKNOWN | Details | |
refresh_var_density_units_dependencies_method | 413D | Updating......... | msrc_UNKNOWN | Details | |
refresh_var_concentration_units_dependencies_method | 413E | Updating......... | msrc_UNKNOWN | Details | |
refresh_var_massflow_units_dependencies_method | 413F | Updating......... | msrc_UNKNOWN | Details | |
refresh_suspension_mode_and_dependencies_method | 4148 | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_var_detector_hv_ctrl_method | 4145 | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_var_temp_units_method | 414A | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_var_mass_units_method | 4149 | Updating......... | msrc_UNKNOWN | Details | |
write_and_refresh_volumetric_flow_units_method | 414B | Updating......... | msrc_UNKNOWN | Details | |
read_hardware_config_method | 4164 | Updating......... | msrc_UNKNOWN | Details | |
loop_test_analog_out_2_method | 4165 | Cout2 Test | Cout2 Test- Allows the user to manually manipulate the Analog Output2 to select constant output value. | msrc_UNKNOWN | Details |
leave_fixed_current_mode_analog_out_2_method | 4166 | 固定電流モードを終ります | 現在の固定モードを終了 - アナログ出力を定数値のモードから移行させ, 出力が入力に応じて変わるモードに戻します。 | msrc_UNKNOWN | Details |
dac_trim_analog_out_2_method | 4167 | D/A トリム | D/A トリム (デジタル/アナログ トリム) - オペレーティング エンド ポイントに対するアナログ出力と外部指示計とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
show_operating_mode_method | 4168 | Show Op.Mode | Show the actual Operation Mode | msrc_UNKNOWN | Details |
loop_test | 1002 | ループ テスト | ループ テスト - アナログ出力をユーザーが選択した値で固定出力します | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A トリム | D/A トリム (デジタル/アナログ トリム) - オペレーティング エンド ポイントに対するアナログ出力と外部指示計とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
return_to_normal | 1016 | ノーマル モードに戻します | msrc_UNKNOWN | Details | |
leave_fixed_current_mode | 1003 | 固定電流モードを終ります | 現在の固定モードを終了 - アナログ出力を定数値のモードから移行させ, 出力が入力に応じて変わるモードに戻します。 | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |