名称 | ID | 标签 | 描述 | 类型 | 源代码 |
---|---|---|---|---|---|
show_calibration_points_method | 4081 | Show Cal. Curve | msrc_UNKNOWN | 详细信息 | |
do_calibration_method | 4082 | Sort&Check Curve | msrc_UNKNOWN | 详细信息 | |
enter_cal_point_method | 4083 | Enter Cal.Pt | msrc_UNKNOWN | 详细信息 | |
enter_background_method | 4084 | Set Background | The background must be measured when the probe is a rod detector. The background is the natural radiation only. The background must be measured, when the source is not yet installed. | msrc_UNKNOWN | 详细信息 |
enter_adjust_lower_cal_point_method | 4085 | Adjust Lo.Cal.Pt | msrc_UNKNOWN | 详细信息 | |
enter_adjust_upper_cal_point_method | 4086 | Adjust Up.Cal.Pt | msrc_UNKNOWN | 详细信息 | |
error_log_method | 4087 | Show Error Log | Shows the last occured errors. | msrc_UNKNOWN | 详细信息 |
change_password_method | 4088 | Change Password | This is a availability to set up max. 3 password levels. The password for the full access level includes the access to configuration and calibration too. Each password can be different or equal to another password. | msrc_UNKNOWN | 详细信息 |
enter_password_method | 4089 | Enter Password | In order to have access to other level, the right password must be inserted. | msrc_UNKNOWN | 详细信息 |
restore_factory_settings_method | 408A | Factory Settings | Factory Reset will delete all settings and restore the factory default values. | msrc_UNKNOWN | 详细信息 |
save_cal_curve_as_basic_curve_method | 408B | Save | Save the current calibration curve in a separate register. E.g. a Basic Calibration Curve. | msrc_UNKNOWN | 详细信息 |
load_basic_curve_method | 408C | Load | Load a previous stored calibration curve, and overwrite the current one. | msrc_UNKNOWN | 详细信息 |
sensor_reset_method | 40CF | Hardware Reset | Hardware Reset will restart the CPU without loosing data. | msrc_UNKNOWN | 详细信息 |
pre_edit_digital_output_method | 40D0 | msrc_UNKNOWN | 详细信息 | ||
post_edit_digital_output_method | 40D1 | msrc_UNKNOWN | 详细信息 | ||
plateau_measurement_method | 40D2 | Measure Plateau | msrc_UNKNOWN | 详细信息 | |
show_plateau_points_method | 40D3 | Show Plateau | msrc_UNKNOWN | 详细信息 | |
loop_test | 1002 | 回路测试 | 回路测试——允许用户将选择的模拟量输出手动调节至选择的常数输出值。 | msrc_UNKNOWN | 详细信息 |
dac_trim | 1004 | D/A修正 | 数字到模拟修正——允许对选择的模拟量输出进行标定,在模拟量输出的操作端点有外部参考。 | msrc_UNKNOWN | 详细信息 |
return_to_normal | 1016 | 中止返回正常 | msrc_UNKNOWN | 详细信息 | |
leave_fixed_current_mode | 1003 | 退出固定电流模式 | 退出固定电流模式——使选择的模拟量输出退出输出为常数值的模式,使输出返回输出跟踪输入的模式。 | msrc_UNKNOWN | 详细信息 |