Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
unit_warning | 4001 | unit warning | Units code warning | msrc_UNKNOWN | Details |
xmtr_reset | 402E | Reset Transmitter | Reset Transmitter - Instructs the transmitter to generate a reset. The result is similar to the effects of briefly interrupting power to the transmitter. This can be used to recover from an error condition if needed. | msrc_UNKNOWN | Details |
zero_point_cal_method | 4033 | Zero point cal | Calibration of the process sensor input via a Zero-Point offset calibration | msrc_UNKNOWN | Details |
span_point_cal_method | 4034 | Span point cal | Calibration of the process sensor input via a Span-Point slope calibration | msrc_UNKNOWN | Details |
read_cal_values | 403E | Read cal values | msrc_UNKNOWN | Details | |
temp_cal_method | 4030 | Temp. Cal | Calibration of Temperature sensor input via one-point smart calibration that adjusts offset and slope | msrc_UNKNOWN | Details |
edit_cal_method | 4031 | Edit Cal | Manual adjustment of process sensor and temperature offset and slope values | msrc_UNKNOWN | Details |
reset_cal_method | 4032 | Reset Cal | Reset the process and temperature sensor calibration data to the values set at the factory | msrc_UNKNOWN | Details |
refresh_variables | 402B | Refresh variables | Read transmitter variables to refresh their values. This will update any variable that was modified via the device keypad. | msrc_UNKNOWN | Details |
loop_test | 1002 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
scaled_dac_trim | 1005 | Scaled D/A trim | Digital to Analog Scaled Trim- Allows the calibration of the Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
warning_message | 1017 | Set loop to manual | msrc_UNKNOWN | Details | |