Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
xmtr_reset | 403E | Reset Transmitter | Reset Transmitter - Instructs the transmitter to generate a reset. The result is similar to the effects of briefly interrupting power to the transmitter. This can be used to recover from an error condition if needed. | msrc_UNKNOWN | Details |
one_point_cal_method | 4043 | One point calib | One Point process calibration | msrc_UNKNOWN | Details |
two_point_cal_method | 4044 | Two point calib | Two Point Calibration - Calibration procedure where two different mV values are used, typically at or near the Lower Range Value and Upper Range Value of the transmitter. | msrc_UNKNOWN | Details |
exit_from_2pt_cal | 404E | Exit from 2 point cal | msrc_UNKNOWN | Details | |
end_cal_mode | 404F | End cal mode | msrc_UNKNOWN | Details | |
temp_cal_method | 4040 | Temp. Cal | Temperature calibration | msrc_UNKNOWN | Details |
edit_cal_method | 4041 | Edit Cal | Edit Calibration - Edit process sensor slope (efficiency) and offset and temp. sensor slope and offset values | msrc_UNKNOWN | Details |
reset_cal_method | 4042 | Reset Cal | Reset the process and temperature sensor calibration data to the values set at the factory | msrc_UNKNOWN | Details |
refresh_variables | 403B | Refresh variables | Read transmitter variables to refresh their values. This will update any variable that was modified via the device keypad. | msrc_UNKNOWN | Details |
loop_test | 1002 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
scaled_dac_trim | 1005 | Scaled D/A trim | Digital to Analog Scaled Trim- Allows the calibration of the Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
warning_message | 1017 | Set loop to manual | msrc_UNKNOWN | Details | |