Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
danger_max | 4079 | Danger for max working pressure | msrc_UNKNOWN | Details | |
back_to_normal | 4002 | Back to normal | msrc_UNKNOWN | Details | |
azero_trim | 4052 | ゼロ トリム | ゼロ トリム - センサーに基準ゼロを入力して補正します。 | msrc_UNKNOWN | Details |
full_trim | 405B | Full trim | Full trim method - The full trimming allows the performance of high or low trim. | msrc_UNKNOWN | Details |
static_trim | 407A | Static trim | Static trim method - The static trimming allows a single point trim of Static Pressure | msrc_UNKNOWN | Details |
set_four_twenty_ma | 4057 | Rerange 4-20mA | Perform a device rerange to 4 - 20mA by applying source pressure input. | msrc_UNKNOWN | Details |
before_updown | 4078 | Before up down scale | msrc_UNKNOWN | Details | |
before_poll | 4077 | Before polling | msrc_UNKNOWN | Details | |
put_in_manual | 4001 | Loop in manual | msrc_UNKNOWN | Details | |
self_test | 4059 | セルフ テスト | Commands the DPK pressure electronics to perform a selftest. | msrc_UNKNOWN | Details |
loop_test | 1002 | ループ テスト | ループ テスト - アナログ出力をユーザーが選択した値で固定出力します | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A トリム | D/A トリム (デジタル/アナログ トリム) - オペレーティング エンド ポイントに対するアナログ出力と外部指示計とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
scaled_dac_trim | 1005 | スケーリング D/A トリム | スケーリング D/A トリム - オペレーティング エンド ポイントに対する選択したアナログ出力と外部指示系とのキャリブレーションを行います。 | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | 固定電流モードを終ります | 現在の固定モードを終了 - アナログ出力を定数値のモードから移行させ, 出力が入力に応じて変わるモードに戻します。 | msrc_UNKNOWN | Details |
return_to_normal | 1016 | ノーマル モードに戻します | msrc_UNKNOWN | Details | |
send_warning | 1399 | Send warning | Send Warning- Yields a units code notice to the User. | msrc_UNKNOWN | Details |