0 |
|
1 |
Serial number |
2 |
Figure |
3 |
Figure: Display |
4 |
Figure: Probe length |
5 |
Figure: Application |
6 |
Figure: Adjustment |
7 |
Figure: Linearization |
8 |
Figure: Current output |
9 |
DD version |
10 |
1.02.01-01 |
11 |
Measurement loop name |
12 |
Sensor |
13 |
Sensor acc. to WHG |
14 |
No |
15 |
Yes |
16 |
SIL qualified sensor |
17 |
Menu language |
18 |
Selection of the language for the indicating and adjustment module |
19 |
Deutsch |
20 |
English |
21 |
Fran�ais |
22 |
Espa�ol |
23 |
Portugu�s |
24 |
Pycckuu |
25 |
Italiano |
26 |
Nederlands |
27 |
Chinese |
28 |
Japanese |
29 |
T�rk�e |
30 |
Polski |
31 |
Cesky |
32 |
Lighting |
33 |
Activation of the background lighting of the indicating and adjustment module |
34 |
Off |
35 |
On |
36 |
Indication value 1 |
37 |
Determines which measured value should be displayed |
38 |
Percent, level |
39 |
Lin. percent, level |
40 |
Filling height of the level |
41 |
Distance to level |
42 |
Scaled level |
43 |
Meas. reliability, level |
44 |
Electronics temperature |
45 |
Percentage, interface |
46 |
Lin. percentage, interface |
47 |
Filling height to the interface |
48 |
Distance to interface |
49 |
Scaled interface |
50 |
Meas. reliability, interface |
51 |
Upper layer thickness |
52 |
Dielectric constant |
53 |
Current output |
54 |
Current output 2 |
55 |
Graph |
56 |
Selection of the presentation for the measured value image |
57 |
Measured value with unit and measurement loop name |
58 |
Measured value with units, measurement loop name and bar graph |
59 |
Two measured values with unit and measurement loop name |
60 |
Software version |
61 |
Version of the instrument software |
62 |
Activity to be carried out |
63 |
Restore delivery status |
64 |
Restore basic settings |
65 |
Device name |
66 |
VEGAFLEX 80 series |
67 |
PULS061.gif |
68 |
1 |
69 |
.3f |
70 |
% |
71 |
2 |
72 |
3 |
73 |
4 |
74 |
5 |
75 |
6 |
76 |
7 |
77 |
8 |
78 |
9 |
79 |
10 |
80 |
11 |
81 |
12 |
82 |
13 |
83 |
14 |
84 |
15 |
85 |
16 |
86 |
17 |
87 |
18 |
88 |
19 |
89 |
20 |
90 |
21 |
91 |
22 |
92 |
23 |
93 |
24 |
94 |
25 |
95 |
26 |
96 |
27 |
97 |
28 |
98 |
29 |
99 |
30 |
100 |
31 |
101 |
32 |
102 |
33 |
103 |
34 |
104 |
Linearization table percentage values |
105 |
Lin% |
106 |
Linearization table lin. percentage values |
107 |
Device status |
108 |
Expanded Device Type |
109 |
x |
110 |
Processor hardware |
111 |
Device address |
112 |
Distance unit of the instrument |
113 |
Selection of the physical distance unit for indication and parameter adjustment |
114 |
mm |
115 |
m |
116 |
in |
117 |
ft |
118 |
Temperature unit of the instrument |
119 |
Selection of the physical temperature unit for indication and parameter adjustment |
120 |
�C |
121 |
K |
122 |
�F |
123 |
Manufacturer ID |
124 |
Bustype ID |
125 |
Sensor type |
126 |
Type of medium |
127 |
Selection of the type of medium |
128 |
Liquids |
129 |
Bulk solid |
130 |
Application |
131 |
Special vessel feature |
132 |
Level |
133 |
Interface (separating layer) |
134 |
Level in the vessel |
135 |
Level in the bypass/standpipe |
136 |
Interface in the vessel |
137 |
Interface in the bypass/standpipe |
138 |
Demonstration mode |
139 |
Vessel form for bulk solids |
140 |
Level in the metal vessel |
141 |
Level in the non-metallic vessel |
142 |
Properties medium/Dielectric constant |
143 |
Solvents, oil, LPG / <3 |
144 |
Chemical mixtures / 3 ... 10 |
145 |
Water-based, acids, lyes / >10 |
146 |
Dust, powder, wood chips / <1.5 |
147 |
Granules, powder, cement / 1.5 ... 3 |
148 |
Cereals, flour, sand / >3 |
149 |
Tube inside diameter |
150 |
To compensate for the propagation speed of radar signals within tubes |
151 |
.4f |
152 |
.2f |
153 |
.1f |
154 |
.0f |
155 |
DTM ID |
156 |
PIN |
157 |
Determine the PIN which should be valid for the function 'Lock adjustment' |
158 |
04d |
159 |
OK |
160 |
Function check |
161 |
Maintenance required |
162 |
Out of specification |
163 |
Malfunction |
164 |
Detail status |
165 |
First HART value (PV) |
166 |
Second HART value (SV) |
167 |
Third HART value (TV) |
168 |
Fourth HART value (QV) |
169 |
m/h |
170 |
ft/h |
171 |
db/m |
172 |
db/ft |
173 |
Password |
174 |
Hardware version |
175 |
Device revision |
176 |
Output mode |
177 |
Fixed current (4 mA) with HART |
178 |
Analogue current output with HART |
179 |
Electronics version |
180 |
--- |
181 |
PS60HC |
182 |
PS60HK |
183 |
PS60HS |
184 |
PS60PAC |
185 |
PS60PAK |
186 |
PS60PAS |
187 |
PS60FFC |
188 |
PS60FFK |
189 |
PS60FFS |
190 |
PLICSZEBVL |
191 |
PLICSZEBVH |
192 |
PS60WLH |
193 |
PS60WLPA |
194 |
PS60WLFF |
195 |
FX80H |
196 |
FX80HSIL |
197 |
FX80PA |
198 |
FX80FF |
199 |
PLICSZEZSA |
200 |
BR80H |
201 |
BR80HSIL |
202 |
BR80PA |
203 |
BR80FF |
204 |
BR80Z |
205 |
BR80HK |
206 |
BR80HKSIL |
207 |
BR80PAK |
208 |
BR80FFK |
209 |
BR80ZK |
210 |
PS60HW |
211 |
PS60PAW |
212 |
PS60FFW |
213 |
LONG-TAG |
214 |
MESSAGE |
215 |
Dielectric constant of upper medium |
216 |
Superimposed gas layer present |
217 |
Non-specific |
218 |
Integration time level (t) |
219 |
s |
220 |
Distance D |
221 |
Distance of the sensor reference plane to the min. height of the interface |
222 |
Distance C |
223 |
Distance of the sensor reference plane to the max. height of the interface |
224 |
Min. adjustment in % |
225 |
Level in percent for which the distance Distance D should be entered |
226 |
Max. adjustment in % |
227 |
Level in percent for which the distance Distance C should be entered |
228 |
Distance B |
229 |
Distance from sensor reference plane to product surface at min. level |
230 |
Distance A |
231 |
Distance from sensor reference plane to product surface at max. level |
232 |
Level in percent for which the distance 'Distance B' must be entered |
233 |
Level in percent for which the distance 'Distance A' must be entered |
234 |
corresponds to 0 % |
235 |
corresponds to 100 % |
236 |
Scaling unit |
237 |
cm |
238 |
mbar |
239 |
bar |
240 |
Pa |
241 |
kPa |
242 |
MPa |
243 |
psi |
244 |
mmH2O |
245 |
mmHg |
246 |
inH2O |
247 |
ftH2O |
248 |
inHg |
249 |
N/m� |
250 |
torr |
251 |
User defined |
252 |
l |
253 |
hl |
254 |
m� |
255 |
in� |
256 |
gal |
257 |
ft� |
258 |
bbl |
259 |
ImpGal |
260 |
yd� |
261 |
bbl_liq |
262 |
bush |
263 |
kg |
264 |
t |
265 |
lb |
266 |
g |
267 |
STon |
268 |
LTon |
269 |
oz |
270 |
kg/m� |
271 |
kg/dm� |
272 |
g/cm� |
273 |
g/ml |
274 |
g/l |
275 |
lb/gal |
276 |
lb/in� |
277 |
STon/yd� |
278 |
deg Brix |
279 |
deg Baume lt |
280 |
deg Baume hv |
281 |
% sol wt |
282 |
% sol vol |
283 |
API |
284 |
pcf |
285 |
SGU |
286 |
l/s |
287 |
l/min |
288 |
l/h |
289 |
l/d |
290 |
Ml/d |
291 |
m�/s |
292 |
m�/min |
293 |
m�/h |
294 |
m�/d |
295 |
lb/s |
296 |
lb/min |
297 |
lb/h |
298 |
lb/d |
299 |
gal/s |
300 |
gal/min |
301 |
gal/h |
302 |
gal/d |
303 |
Mgal/d |
304 |
ft�/s |
305 |
ft�/min |
306 |
ft�/h |
307 |
ft�/d |
308 |
bbl/s |
309 |
bbl/min |
310 |
bbl/h |
311 |
bbl/d |
312 |
ImpGal/s |
313 |
ImpGal/min |
314 |
ImpGal/h |
315 |
ImpGal/d |
316 |
g/s |
317 |
g/min |
318 |
g/h |
319 |
kg/s |
320 |
kg/min |
321 |
kg/h |
322 |
kg/d |
323 |
STon/s |
324 |
STon/min |
325 |
STon/h |
326 |
STon/d |
327 |
LTon/h |
328 |
LTon/d |
329 |
t/min |
330 |
t/h |
331 |
t/d |
332 |
A |
333 |
mA |
334 |
% sol/vol |
335 |
- |
336 |
pF |
337 |
ct/s |
338 |
lb/ft |
339 |
kg/m |
340 |
Hz |
341 |
dB |
342 |
mV |
343 |
unknown |
344 |
m/s |
345 |
ft/s |
346 |
ft/min |
347 |
Scaling format |
348 |
0 |
349 |
0.0 |
350 |
0.00 |
351 |
0.000 |
352 |
0.0000 |
353 |
Linearization type |
354 |
Linear |
355 |
Horizontal cylindrical vessel |
356 |
Spherical vessel |
357 |
Venturi, trapezoidal weir, rectangle weir |
358 |
Palmer-Bowlus Flume |
359 |
V-Notch, triangle weir |
360 |
User programmable |
361 |
Vessel height D |
362 |
Socket correction h |
363 |
Integration time interface (t) |
364 |
Min. current |
365 |
3.8 mA |
366 |
4 mA |
367 |
Max. current |
368 |
20.5 mA |
369 |
20 mA |
370 |
Output characteristic |
371 |
0 ... 100 % correspond to 4 ... 20 mA |
372 |
0 ... 100 % correspond to 20 ... 4 mA |
373 |
Failure mode |
374 |
>=21 mA |
375 |
Last valid measured value |
376 |
<= 3.6 mA |
377 |
Status distance to level |
378 |
Failure |
379 |
Upper limitation |
380 |
Lower limitation |
381 |
Simulation value |
382 |
Substitute value |
383 |
Constant value |
384 |
Status filling height of the level |
385 |
Carry over the adjustment of the level measurement? |
386 |
Status percent, level |
387 |
Status lin. percent, level |
388 |
Status scaled level |
389 |
Status current output |
390 |
Status meas. reliability, level |
391 |
Sensor element type |
392 |
Rod diameter 8 mm |
393 |
Cable diameter 2 mm centering weight |
394 |
Cable diameter 2 mm gravity weight |
395 |
Cable diameter 4 mm centering weight |
396 |
Cable diameter 4 mm gravity weight |
397 |
Cable diameter 4 mm |
398 |
Rod diameter 16 mm |
399 |
Cable diameter 6 mm bulk solid gravity weight |
400 |
Cable diameter 11 mm coated gravity weight |
401 |
Cable diameter 4 mm bulk solid gravity weight |
402 |
Cable diameter 6 mm coated gravity weight |
403 |
Rod diameter 16 mm high temperature |
404 |
Cable diameter 2 mm centering weight high temperature |
405 |
Cable diameter 2 mm gravity weight high temperature |
406 |
Cable diameter 4 mm centering weight high temperature |
407 |
Cable diameter 4 mm gravity weight high temperature |
408 |
Rod diameter 16 mm high temperature reference |
409 |
Cable diameter 2 mm centering weight high temperature reference |
410 |
Cable diameter 2 mm gravity weight high temperature reference |
411 |
Cable diameter 4 mm centering weight high temperature reference |
412 |
Cable diameter 4 mm gravity weight high temperature reference |
413 |
Rod diameter 8 mm hygienic |
414 |
Rod diameter 10 mm coated |
415 |
Rod diameter 12 mm |
416 |
Cable diameter 4 mm coated gravity weight |
417 |
Coax diameter 21x8 mm |
418 |
Coax diameter 21 x 6 mm high temperature |
419 |
Coax diameter 42x16 mm |
420 |
Coax diameter 42x16 mm high temperature |
421 |
Coax diameter 42x16 mm high temperature reference |
422 |
Probe length L from seal surface |
423 |
Status distance to interface |
424 |
Status filling height to the interface |
425 |
Status percent interface |
426 |
Status lin. percentage, interface |
427 |
Status scaled interface |
428 |
Current output 0 % corresponds to |
429 |
Current output 100 % corresponds to |
430 |
Status meas. reliability, interface |
431 |
Status dielectric constant |
432 |
Status upper layer thickness |
433 |
Analogue current output |
434 |
Status electronics temperature |
435 |
Selection of an activity to process the gating out of false signals |
436 |
Cancel |
437 |
Create new |
438 |
Delete |
439 |
Sounded distance from the sealing surface to the medium |
440 |
Simulation, lin. percent |
441 |
Simulation, percent |
442 |
Simulation, distance |
443 |
Simulation lin. percentage, interface |
444 |
Simulation percent, interface |
445 |
Simulation distance to interface |
446 |
Simulation, current output |
447 |
Security at the probe end |
448 |
Correction of propagation velocity |
449 |
Averaging factor on amplitude |
450 |
d |
451 |
Echo detection change-over point close range/far range |
452 |
Echo detection threshold close range positive |
453 |
Echo detection threshold far range positive |
454 |
Echo detection threshold probe end positive |
455 |
Echo detection threshold close range negative |
456 |
Echo detection threshold far range negative |
457 |
Echo detection threshold probe end negative |
458 |
'First large echo' active |
459 |
Amplitude relation 'First large echo' |
460 |
Width focussing range level |
461 |
Time for opening of the focus range, level |
462 |
Measurement reliability outside/inside focussing range |
463 |
Measured value offset |
464 |
Measurement via probe end tracking |
465 |
Application-dependent |
466 |
Dynamic false signal suppression |
467 |
Date of the min. dielectric constant |
468 |
Dielectric constant min. |
469 |
Date of the max. dielectric constant |
470 |
Dielectric constant max. |
471 |
Date of the min. measurement reliability |
472 |
Minimum measurement reliability |
473 |
Date of the max. measurement reliability |
474 |
Max. measurement reliability |
475 |
Electronics temperature (min.) |
476 |
Min. electronics temperature |
477 |
Electronics temperature (max.) |
478 |
Max. electronics temperature |
479 |
Date of the min. distance |
480 |
Min. distance |
481 |
Date of the max. distance |
482 |
Max. distance |
483 |
Probe end detection threshold |
484 |
Probe end tracking: echo measurement |
485 |
Negative probe end |
486 |
Positive probe end |
487 |
Indication value 2 |
488 |
Carry over the scaling of the level measurement? |
489 |
Limitation measuring range begin active |
490 |
Limitation measuring range begin (positive echoes) |
491 |
Standard probe |
492 |
Special probe |
493 |
Behaviour in case of loss of echo |
494 |
Failure message |
495 |
Maintenance message |
496 |
Delay 'Signal with echo loss' |
497 |
Cable breakage detection/Probe loss |
498 |
Width focussing range, interface |
499 |
Overfill safety |
500 |
Dielectric constant, medium |
501 |
Probe end tracking: Use the dielectric constant determined by the sensor? |
502 |
Reference echo measurement |
503 |
Sump detection |
504 |
Simulation, current output 2 |
505 |
Status current output 2 |
506 |
Simulation active |
507 |
Assessment echo mobility |
508 |
Store echo curve |
509 |
Echo curve of the setup |
510 |
Not available |
511 |
Determine probe length automatically |
512 |
None |
513 |
Calculate |
514 |
Apply |
515 |
Max. amplitude, interface: Switching over to level |
516 |
Selection interface echo |
517 |
First echo |
518 |
Last echo |
519 |
Largest echo |
520 |
Temperature offset, steam compensation |
521 |
Setup |
522 |
Type of linearization |
523 |
Diagnostics |
524 |
Additional settings |
525 |
Info |
526 |
Figure: Setup |
527 |
Figure: Adjustment, level |
528 |
Figure: Adjustment, interface |
529 |
Figure: Damping |
530 |
Figure: Linearization table |
531 |
Figure: Simulation |
532 |
Figure: Status signals |
533 |
Figure: Reset |
534 |
Figure: Scaling, level |
535 |
Figure: Current output (adjustment) |
536 |
Figure: HART variables |
537 |
Figure: Service login |
538 |
Figure: Info |
539 |
Display |
540 |
Probe length |
541 |
An automatic determination of the probe length requires a free, uncovered sensor. |
542 |
The guided radar measuring principle requires a metallic radiating surface on the process fitting. For this purpose, you have to use a sheet metal (� > 200 mm/8 in) with direct contact to the process fitting (without contact to the probe). Flange versions from DN 50 do not need an additional sheet metal. |
543 |
Adjustment, level |
544 |
Adjustment, interface |
545 |
Damping |
546 |
The sensor is in multidrop mode |
547 |
False signal suppression |
548 |
If the upper edge of the socket is below the upper edge of the vessel, then the value must be entered with negative signs. |
549 |
Status signals |
550 |
Proof test |
551 |
This test is part of the proof test to be carried out with SIL devices. You can find detailed information in the Safety Manual. |
552 |
Reset |
553 |
Scaling, level |
554 |
Scaling, interface |
555 |
Current output (adjustment) |
556 |
The selected output variable is also assigned to the first HART value (PV). |
557 |
Adjustment |
558 |
Current output 2 (adjustment) |
559 |
The selected output variable is also assigned to the second HART value (SV). |
560 |
HART variables |
561 |
After changing the probe type, a reset to basic settings is necessary |
562 |
Service login |
563 |
Special parameters |
564 |
Read values from device |
565 |
Write values to device |
566 |
Linearization table |
567 |
Percent |
568 |
Lin. percent |
569 |
Linearization curve |
570 |
Table |
571 |
Distance |
572 |
V |
573 |
Echo curve |
574 |
Sensitivity reference |
575 |
Echo curve (curve) |
576 |
Show echo curve |
577 |
Settings |
578 |
Load actual echo curves |
579 |
Load reference line |
580 |
Load setup curve |
581 |
Load selected curves |
582 |
Store echo curve of the setup |
583 |
Zoom range from |
584 |
Zoom range up to |
585 |
Device characteristics |
586 |
Details |
587 |
Version |
588 |
Update instrument features |
589 |
Simulation |
590 |
Activate simulation |
591 |
Deactivate simulation |
592 |
Accept simulation value |
593 |
Identification |
594 |
HART Primary Value (PV) |
595 |
Trend display |
596 |
Additional measured values |
597 |
Outputs |
598 |
Measured values |
599 |
Service login required |
600 |
Parameter adjustment |
601 |
Maintenance |
602 |
Service logout |
603 |
Determine dielectric constant |
604 |
Reset? |
605 |
Reset 'Configuration changed' |
606 |
Reset peak indicator |
607 |
Status, level |
608 |
Status interface |
609 |
Status, additional measured values |
610 |
Status, outputs |
611 |
HART Device Status |
612 |
Peak indicator, distance to the level |
613 |
Peak indicator, distance to the interface |
614 |
Peak indicator, reliability level |
615 |
Peak indicator, reliability interface |
616 |
Peak indicator, electronics temperature |
617 |
Peak indicator, dielectric constant |
618 |
VEGAPULS 61 |
619 |
VEGAPULS 62 |
620 |
VEGAPULS 63 |
621 |
VEGAPULS 65 |
622 |
VEGAPULS 66 |
623 |
VEGAPULS 67 |
624 |
VEGAPULS 68 |
625 |
VEGAPULS WL61 |
626 |
VEGAPULS SR68 |
627 |
VEGAPULS 69 |
628 |
VEGAFLEX 80 series SIL |
629 |
VEGABAR 80 series |
630 |
VEGABAR 80 series SIL |
631 |
VEGA Grieshaber KG |
632 |
Siemens |
633 |
KROHNE Messtechnik GmbH |
634 |
Unlock adjustment |
635 |
Verify and lock adjustment |
636 |
Verify adjustment |
637 |
Lock adjustment |
638 |
Start proof test |
639 |
Common Tables |
640 |
Interoperable * |
641 |
Universal |
642 |
Common Practice |
643 |
Pressure Universal |
644 |
Pressure Common |
645 |
Menu |
646 |
Rosemount Magflow |
647 |
Rosemount Pressure |
648 |
Moore Products |
649 |
Siemens Pressure |
650 |
P2 Common |
651 |
PV |
652 |
Wireless |
653 |
Munge |
654 |
Analytical Oxygen/Combustion/Methane/CO |
655 |
Flow Totalizer - Rate Indicator |
656 |
Flow Variable Area |
657 |
Flow Coriolis Mass |
658 |
Flow Ultrasonic - Gas |
659 |
Flow Ultrasonic - Liquid |
660 |
Level Ultrasonic |
661 |
Level Radar - Microwave |
662 |
Pressure Absolute |
663 |
Level Capacitance |
664 |
Pressure Differential |
665 |
Level Hydrostatic |
666 |
Level |
667 |
Pressure Absolute/Relative |
668 |
Level Microwave |
669 |
Level Radar |
670 |
Level Density - Nuclear |
671 |
Flow Electromagnetic |
672 |
Flow Vortex |
673 |
Flow Magmeter - Rack Version |
674 |
Flow High Performance Magmeter |
675 |
Flow Mass |
676 |
Flow Ultrasonic |
677 |
Flow Magnetic |
678 |
Flow Magnetic - Custody Transfer |
679 |
Flow 2-wire Magnetic |
680 |
Flow Coriolis Mass - Custody Transfer |
681 |
Flow 2-wire Vortex |
682 |
Flow Thermal |
683 |
Temperature Head Mounted |
684 |
Temperature Compact |
685 |
Temperature DIN Rail Mounted |
686 |
Analytical pH |
687 |
Analytical Conductivity - Conductive |
688 |
Analytical Conductivity - Inductive |
689 |
Temperature Sensor |
690 |
Level Servo |
691 |
Temperature DIN Rail |
692 |
Temperature |
693 |
Analytical ORP |
694 |
Analytical pION |
695 |
Analytical IConc |
696 |
Analytical Electrode Conductivity |
697 |
Analytical Toroidal Conductivity |
698 |
Valve Positioner |
699 |
Valve Controller |
700 |
Pressure |
701 |
Pressure Multivariable |
702 |
Analytical pH/ORP |
703 |
Flow Coriolis |
704 |
Analytical Conductivity |
705 |
Valve Positioner/Pneumatic |
706 |
Valve Actuator - Electrical |
707 |
Temperature Dual Input |
708 |
Valve Actuator |
709 |
Analytical Corrosion |
710 |
Analytical Density - Nuclear |
711 |
Level Continuous Gamma |
712 |
Level Laser Ranging |
713 |
Pressure Safety |
714 |
Flow Vortex/Swirl |
715 |
Flow |
716 |
Analytical 4 Electrode Conductivity |
717 |
Analytical 2 Electrode Conductivity |
718 |
Analytical Troidal Conductivity |
719 |
Analytical pH/ORP/pIon/Ion Conc |
720 |
Analytical Oxygen |
721 |
Pressure P, dP + Multivariable |
722 |
Temperature 2-wire |
723 |
Pressure w/ PID |
724 |
Pressure Saturated Steam |
725 |
Analytical Density - Gamma Ray |
726 |
Level Gamma Ray |
727 |
Analytical Weight - Gamma Ray |
728 |
Level w/ Density - Gamma Ray |
729 |
Analytical Motorized Density |
730 |
Flow Multivariable Mass |
731 |
Temperature Multivariable |
732 |
Specialty HART-to-Analog Signal Conv |
733 |
Flow dP |
734 |
Pressure Low Power |
735 |
Pressure Certified SIL2 |
736 |
Interoperable |
737 |
Analytical Density |
738 |
Valve Positioner/Electropneumatic |
739 |
Flow Magnetic Inductive |
740 |
Analytical |
741 |
Analytical Opacity Monitor |
742 |
Analytical Oxygen/Combustion |
743 |
Analytical 2-wire Conductivity |
744 |
Analytical 2-wire Conductivity w/ pure H20 |
745 |
Analytical 2-wire Conductivity w/ % conc |
746 |
Analytical 2-wire Oxygen/Chlorine/Ozone |
747 |
Analytical 2-wire pH/ORP |
748 |
Analytical 2-wire Contacting/Toroidal |
749 |
Analytical Oxygen - High Temperature |
750 |
Analytical Conductivity w/ % conc |
751 |
Analytical pH/ORP/ISE |
752 |
Analytical Amperometric Analyzer |
753 |
Analytical Dual Input Analyzer |
754 |
Pressure Pressure/Level |
755 |
Analytical Consistency |
756 |
Analytical Microwave-Pulp Consistency |
757 |
Analytical Microwave Consistency |
758 |
Valve Digital Positioner |
759 |
Valve Smart Valve Positioner |
760 |
Temperature Differential |
761 |
Analytical Dissolved Oxygen |
762 |
Analytical pH Controller |
763 |
Analytical Conductivity/Resistivity Controller |
764 |
Analytical Conductivity/Resistivity |
765 |
Level Displacement |
766 |
Temperature w/ PID |
767 |
Valve Universal Positioner |
768 |
Level Radar - Reflex |
769 |
Pressure Gauge |
770 |
Pressure Sealed Gauge |
771 |
Valve Digital Positioner w/ PID |
772 |
Valve Intelligent Positioner |
773 |
Analytical Gas Detector |
774 |
Analytical Infrared Gas Detector |
775 |
Analytical Electrochemical Gas Detector |
776 |
Analytical Measuring Instruments |
777 |
Level Magnostrictive |
778 |
Valve On/Off Valve Controller |
779 |
Level Radar - Pulse |
780 |
Level Continuous Capactive |
781 |
Valve Damper Actuator |
782 |
Valve Electric Damper Actuator |
783 |
Flow Electronic |
784 |
Flow Vortex - Multiparameter Mass |
785 |
Analytical UV Flame Detector |
786 |
Analytical UVIR Flame Detector |
787 |
Analytical IR Flame Detector |
788 |
Analytical Fire Detector |
789 |
Analytical Gas Detector Display Interface |
790 |
Analytical Toxic Gas Detector |
791 |
Analytical Flame Detector |
792 |
Level Nuclear |
793 |
Analytical Microwave |
794 |
Analytical Density - Radiometric |
795 |
Level Radiometric |
796 |
Analytical Gas/Steam |
797 |
Level Nucleonic |
798 |
Analytical Blade Consistency |
799 |
Analytical Rotary Consistency |
800 |
Pressure Absolute/Differential |
801 |
Positioner |
802 |
Analytical Combustible Gas Detector |
803 |
Analytical Point Infrared Gas Detector |
804 |
Analytical Multi-Detector Monitor System |
805 |
Valve Positioner/Electro-hydraulic |
806 |
Valve Monitor |
807 |
Enable on token-passing data link layer only |
808 |
Enable on TDMA data link layer only |
809 |
Enable on both TDMA and token-passing data link layer |
810 |
WirelessHART Device |
811 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
812 |
Primary |
813 |
Primary Analog Channel |
814 |
Secondary |
815 |
Secondary Analog Channel |
816 |
Tertiary |
817 |
Tertiary Analog Channel |
818 |
Quaternary |
819 |
Quaternary Analog Channel |
820 |
Quinary |
821 |
Quinary Analog Channel |
822 |
Critical Power Failure |
823 |
Lock all |
824 |
Trim not supported |
825 |
Configuration cannot be changed |
826 |
Locked by gateway |
827 |
Analog Input Channel |
828 |
Analog Output Channel |
829 |
Secondary Analog Channel Saturated |
830 |
Tertiary Analog Channel Saturated |
831 |
Quaternary Analog Channel Saturated |
832 |
Quinary Analog Channel Saturated |
833 |
Secondary Analog Channel Fixed |
834 |
Tertiary Analog Channel Fixed |
835 |
Quaternary Analog Channel Fixed |
836 |
Quinary Analog Channel Fixed |
837 |
Simulation active |
838 |
The device is in simulation mode and the level output is not representative of the process. |
839 |
Non-Volatile memory failure |
840 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
841 |
Volatile memory error |
842 |
The RAM memory check is invalid or maybe corrupt |
843 |
Watchdog reset executed |
844 |
A watchdog reset has been performed |
845 |
Voltage conditions out of range |
846 |
A voltage condition is outside its allowable range |
847 |
Environmental conditions out of range |
848 |
An internal or environmental condition is beyond acceptable limits. |
849 |
Electronic failure |
850 |
A hardware problem not related to the sensor has been detected. |
851 |
Subdevice list changed |
852 |
Duplicate master detected |
853 |
Capacity Denied |
854 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
855 |
Bandwidth allocation pending |
856 |
Block transfer pending |
857 |
Continuous |
858 |
The Burst Message is published continuously at (worst case) the Minimum Update Period. |
859 |
Window |
860 |
The Burst Message is triggered when the source value deviates more than the specified trigger value |
861 |
Rising |
862 |
The Burst Message is triggered when source value Rises Above the specified trigger value. |
863 |
Falling |
864 |
The Burst Message is triggered when the source value Falls Below the specified trigger value. |
865 |
On-change |
866 |
The Burst Message is triggered when any value in the message changes. |
867 |
Battery life |
868 |
Percent range |
869 |
Loop current |
870 |
Primary variable |
871 |
Secondary variable |
872 |
Tertiary variable |
873 |
Quaternary variable |
874 |
Off |
875 |
Enable event notification on token-passing data link layer |
876 |
Enable event notification on TDMA data link layer |
877 |
Enable event notification on TDMA and token-passing data link layer |
878 |
Configuration changed event pending |
879 |
Device status event pending |
880 |
More status available event pending |
881 |
Disable |
882 |
Enable single data point trending |
883 |
Enable filtered trending |
884 |
The sample is filtered using a time constant equal to one-third the trend sample period |
885 |
Enable average trending |
886 |
The value obtained by dividing the sum of all samples in the period by the total number of samples |
887 |
Read receive time |
888 |
Write date and time |
889 |
Non-Volatile Clock |
890 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
891 |
Clock Uninitialized |
892 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
893 |
Line Power |
894 |
Device is powered via a hardwired connection to a plant power source. |
895 |
Battery Power |
896 |
Device operates solely of its (internal) battery. |
897 |
Recharg.B./Power Sc. |
898 |
Rechargeable battery power or power scavenging. The device has a short-term supply of energy (must be at least one hour) that is continuously being replenished. The device's power is being replenished by harvesting and converting energy from the environment surrounding the device (e.g. solar, vibration, heat). |
899 |
Idle |
900 |
Active search |
901 |
Looking for neighbors. |
902 |
Negotiating |
903 |
Quarantined |
904 |
Operational |
905 |
Suspended |
906 |
Deep sleep |
907 |
Ultra-low power |
908 |
Wireless Signal Found |
909 |
Network packets have been recognized |
910 |
Wireless Signal Identified |
911 |
Wireless Time Synchronized |
912 |
WirelessHart Signal Identified |
913 |
Network Admission Requested |
914 |
Set on transmission of first join request |
915 |
Join Retrying |
916 |
Set after the first join request retry (i.e. when number of join requests > 2). Cleared when device is authenticated or when Active Search started. |
917 |
Join Failed |
918 |
Set on transition from Active Search mode to Deep Sleep mode. Cleared on transition from Deep Sleep to Active Search. |
919 |
Network Security Clearance Granted |
920 |
Network Manager Session Established |
921 |
Network Joined |
922 |
Network Joined. Normal superframe and links obtained. |
923 |
Network Bandwidth Requested |
924 |
Negotiating Network Properties. Gateway session obtained. |
925 |
Join Complete |
926 |
Normal operating mode. |
927 |
None |
928 |
No restrictions |
929 |
SI Only |
930 |
Restricted to SI Unit codes only |
931 |
Open - accept all devices |
932 |
Use Whitelist for access |
933 |
Use Blacklist |
934 |
Use Whitelist and Blacklist |
935 |
Lockdown - disable join links |
936 |
Process automation device |
937 |
Discrete device |
938 |
Hybrid:Process automation+discrete |
939 |
I/O System |
940 |
WirelessHART Process automation device |
941 |
WirelessHART Discrete device |
942 |
WirelessHART Process automation+discrete |
943 |
WirelessHART Gateway |
944 |
WirelessHART Access point |
945 |
WirelessHART Process adapter |
946 |
WirelessHART Discrete adapter |
947 |
WirelessHART enabled handheld/portable maintenance tool |
948 |
Don't attempt to join |
949 |
Join now |
950 |
Attempt to join immediately on powerup or reset |
951 |
US |
952 |
United States of America |
953 |
JP |
954 |
Japan |
955 |
DE |
956 |
Germany |
957 |
FR |
958 |
France |
959 |
ES |
960 |
Spain |
961 |
RU |
962 |
Russian Federation |
963 |
CN |
964 |
People's Republic of China |
965 |
Nominal |
966 |
Device Power Status is nominal |
967 |
Low |
968 |
Device Power Status is low |
969 |
Critically Low |
970 |
Device Power Status is critically low |
971 |
Recharging - Low |
972 |
Recharging - High |
973 |
Secondary Master |
974 |
Primary Master |
975 |
PV Analog Channel Saturated |
976 |
PV Analog Channel Fixed |
977 |
Device Profile |
978 |
Device Diagnostic Status 0 |
979 |
Device Diagnostic Status 1 |
980 |
I/O and Subdevice Status |
981 |
WirelessHART Status |
982 |
Time Set Option |
983 |
Real Time Clock Flags |
984 |
Country |
985 |
Country of intended device installation |
986 |
SI Unit Control |
987 |
Indicates if device should only display SI unit codes |
988 |
Burst Trigger Units |
989 |
Burst Trigger Class |
990 |
Wired HART Enabled |
991 |
Enabled on Wired HART device connection only |
992 |
Wireless HART Enabled |
993 |
Enabled on WirelessHART device connection only |
994 |
Both Wired and WirelessHart Enabled |
995 |
Enabled on both Wired and WirelessHART connections |
996 |
Burst Msg Trigger Mode |
997 |
Event Status |
998 |
Event Notification Control |
999 |
Operate mode Mask |
1000 |
Not Defined |
1001 |
Power Source |
1002 |
Power source of the device |
1003 |
Wireless Mode |
1004 |
Join Status |
1005 |
The current location in the joining sequence of operations. |
1006 |
Network Access Mode |
1007 |
Join Mode |
1008 |
Power Status |
1009 |
Power status of the device |
1010 |
Master Mode |
1011 |
This sets the I/O System's Data Link Layer operation on all channels as either primary or secondary master mode. |
1012 |
Trend Units |
1013 |
Trend Device Variable Unit Code |
1014 |
Trend Classification |
1015 |
Trend Device Variable Classification |
1016 |
Trend Control |
1017 |
uint16_var_blk_char |
1018 |
uint8_var_blk_char |
1019 |
Device Status Mask |
1020 |
Ext dev status Mask |
1021 |
Device Diagnostic Status 0 Mask |
1022 |
Device Diagnostic Status 1 Mask |
1023 |
Device Diagnostic Status 2 Mask |
1024 |
Device Diagnostic Status 3 Mask |
1025 |
AO saturated Mask |
1026 |
AO fixed Mask |
1027 |
Latched Device Status |
1028 |
Latched Ext dev status |
1029 |
Latched Operate Mode |
1030 |
Latched Device Diagnostic Status 0 |
1031 |
Latched Device Diagnostic Status 1 |
1032 |
Latched Device Diagnostic Status 2 |
1033 |
Latched Device Diagnostic Status 3 |
1034 |
Latched AO saturated |
1035 |
Latched AO fixed |
1036 |
Wireless Module Device Type |
1037 |
Device type of the Wireless Radio |
1038 |
Wireless Module Manufacturer ID |
1039 |
Manufacturer of the Wireless Radio contained in this device |
1040 |
Subdevice Device Type |
1041 |
Device type of the subdevice attached to this system. |
1042 |
Subdevice Manufacturer Id |
1043 |
Manufacturer of the subdevice attached to this system. |
1044 |
8u |
1045 |
Dynamic Variables Returned for Device Variables |
1046 |
Command Response Truncated |
1047 |
Configuration Change Counter Mismatch |
1048 |
Status bytes mismatch |
1049 |
8.3f |
1050 |
Output Status |
1051 |
Output Status Byte |
1052 |
Uninitialized |
1053 |
Normal - one sensor |
1054 |
Normal - two sensors |
1055 |
Normal - three sensors |
1056 |
Shorted |
1057 |
Shorted- shorted field wiring. |
1058 |
Limited high |
1059 |
Limited high- block can not respond to upward movement of value. |
1060 |
Open |
1061 |
Open- open field wiring. |
1062 |
Limited low |
1063 |
Limited low- block can not respond to upward movement of target. |
1064 |
Initialization done |
1065 |
Initialization done- block initialization is complete. |
1066 |
Not selected |
1067 |
Not selected- target value provided by block was not selected. |
1068 |
No path |
1069 |
No path- there is no path to the process. |
1070 |
Fail |
1071 |
Fail- hardware failure. |
1072 |
Bad |
1073 |
Bad- block is out of service or has bad input. |
1074 |
Not from process |
1075 |
Not from process- value is not directly affected by any process information. |
1076 |
Local override |
1077 |
Local override- block has LO mode. |
1078 |
Not Cascade Mode |
1079 |
Not RCas Mode |
1080 |
Not ROut Mode |
1081 |
No communication |
1082 |
No communication- (NO_COM). |
1083 |
uint32_var_blk_char |
1084 |
int8_var_blk_char |
1085 |
new itest redef1 label |