0 |
|
1 |
|
2 |
Serial number |
3 |
Figure |
4 |
Figure: Status signals |
5 |
Figure: Application |
6 |
Figure: Adjustment |
7 |
Figure: Position correction |
8 |
Figure: Display |
9 |
Figure: Current output |
10 |
Figure: Current output 2 |
11 |
Figure: Linearization |
12 |
Figure: Low flow suppression |
13 |
Figure: Totalizer |
14 |
DD version |
15 |
1.03.02-01 |
16 |
Measurement loop name |
17 |
Name of the measurement loop |
18 |
Sensor |
19 |
Device name |
20 |
VEGADIF 80 series |
21 |
Sensor acc. to WHG |
22 |
Use as a part of an overfill protection system according to WHG |
23 |
No |
24 |
Yes |
25 |
Software version |
26 |
Version of the instrument software |
27 |
Manufacturer ID |
28 |
x |
29 |
Expanded Device Type |
30 |
Identification of the instrument type through Fieldbus |
31 |
Bustype ID |
32 |
Sensor type |
33 |
DTM ID |
34 |
Device-Revision |
35 |
Compatibility information for integration into the device catalogue |
36 |
PIN |
37 |
Determine the PIN which should be valid for the function 'Lock adjustment' |
38 |
04d |
39 |
Graph |
40 |
Selection of the presentation for the measured value image |
41 |
Measured value |
42 |
Measured value with bar graph |
43 |
Two measured values |
44 |
Indication value 1 |
45 |
Determines which measured value should be displayed |
46 |
Flow |
47 |
Totalizer 1 |
48 |
Totalizer 2 |
49 |
Process pressure |
50 |
Pressure |
51 |
Differential pressure |
52 |
Density |
53 |
Height Interface |
54 |
Filling height |
55 |
Static pressure |
56 |
Percent |
57 |
Scaled |
58 |
Current output |
59 |
Lin. percent |
60 |
Measuring cell temperature |
61 |
Electronics temperature |
62 |
Current output 2 |
63 |
Menu language |
64 |
Selection of the language for the indicating and adjustment module |
65 |
Deutsch |
66 |
English |
67 |
Français |
68 |
Español |
69 |
Português |
70 |
Italiano |
71 |
Nederlands |
72 |
Pycckuu |
73 |
Chinese |
74 |
Japanese |
75 |
Polski |
76 |
Cesky |
77 |
Türkçe |
78 |
Lighting |
79 |
Activation of the background lighting of the indicating and adjustment module |
80 |
Off |
81 |
On |
82 |
Indication value 2 |
83 |
0 % corresponds to |
84 |
Scaled value for 0 % |
85 |
.0f |
86 |
.1f |
87 |
.2f |
88 |
.3f |
89 |
.4f |
90 |
100 % corresponds to |
91 |
Scaled value for 100 % |
92 |
Scaling unit |
93 |
Measuring unit for the scaled value |
94 |
mm |
95 |
cm |
96 |
m |
97 |
in |
98 |
ft |
99 |
mbar |
100 |
bar |
101 |
Pa |
102 |
kPa |
103 |
MPa |
104 |
psi |
105 |
mmH2O |
106 |
mmHg |
107 |
inH2O |
108 |
ftH2O |
109 |
inHg |
110 |
torr |
111 |
User defined |
112 |
l |
113 |
hl |
114 |
m³ |
115 |
in³ |
116 |
gal |
117 |
ft³ |
118 |
bbl |
119 |
ImpGal |
120 |
yd³ |
121 |
bbl_liq |
122 |
bush |
123 |
kg |
124 |
t |
125 |
lb |
126 |
g |
127 |
STon |
128 |
LTon |
129 |
oz |
130 |
kg/m³ |
131 |
kg/dm³ |
132 |
g/cm³ |
133 |
g/ml |
134 |
g/l |
135 |
lb/gal |
136 |
lb/in³ |
137 |
STon/yd³ |
138 |
deg Brix |
139 |
deg Baume lt |
140 |
deg Baume hv |
141 |
% sol wt |
142 |
% sol vol |
143 |
API |
144 |
pcf |
145 |
SGU |
146 |
l/s |
147 |
l/min |
148 |
l/h |
149 |
Ml/d |
150 |
m³/s |
151 |
m³/min |
152 |
m³/h |
153 |
m³/d |
154 |
gal/s |
155 |
gal/min |
156 |
gal/h |
157 |
gal/d |
158 |
Mgal/d |
159 |
ft³/s |
160 |
ft³/min |
161 |
ft³/h |
162 |
ft³/d |
163 |
bbl/s |
164 |
bbl/min |
165 |
bbl/h |
166 |
bbl/d |
167 |
ImpGal/s |
168 |
ImpGal/min |
169 |
ImpGal/h |
170 |
ImpGal/d |
171 |
g/s |
172 |
g/min |
173 |
g/h |
174 |
kg/s |
175 |
kg/min |
176 |
kg/h |
177 |
kg/d |
178 |
lb/s |
179 |
lb/min |
180 |
lb/h |
181 |
lb/d |
182 |
STon/min |
183 |
STon/h |
184 |
STon/d |
185 |
LTon/h |
186 |
LTon/d |
187 |
t/min |
188 |
t/h |
189 |
t/d |
190 |
°C |
191 |
°F |
192 |
K |
193 |
mA |
194 |
% |
195 |
- |
196 |
pF |
197 |
ct/s |
198 |
lb/ft |
199 |
kg/m |
200 |
Hz |
201 |
dB |
202 |
V |
203 |
mV |
204 |
unknown |
205 |
m/s |
206 |
m/h |
207 |
ft/s |
208 |
ft/min |
209 |
Scaling format |
210 |
Determination of the presentation of decimal points |
211 |
# |
212 |
#.# |
213 |
#.## |
214 |
#.### |
215 |
#.#### |
216 |
Selection of the application |
217 |
By selecting the application, the sensor can be adapted to the measurement task |
218 |
Level |
219 |
Interface (separating layer) |
220 |
Density compensated level |
221 |
Integration time (t) |
222 |
Damping of measured value peaks |
223 |
s |
224 |
Adjustment unit of the instrument |
225 |
Selection of the physical unit for indication and parameter adjustment |
226 |
lb/ft³ |
227 |
Temperature unit of the instrument |
228 |
Selection of the physical temperature unit for indication and parameter adjustment |
229 |
Linearization type |
230 |
Selection of the linearization curve |
231 |
Linear |
232 |
User programmable |
233 |
Horizontal cylindrical vessel |
234 |
Spherical vessel |
235 |
Square root |
236 |
Bidirectionally linear |
237 |
Bidirectional square root |
238 |
Venturi, trapezoidal weir, rectangle weir |
239 |
Palmer-Bowlus Flume |
240 |
V-Notch, triangle weir |
241 |
Vessel height D |
242 |
Total height (inner dimension) |
243 |
Socket correction h |
244 |
Distance of the sensor reference plane to the upper edge of the vessel |
245 |
1 |
246 |
2 |
247 |
3 |
248 |
4 |
249 |
5 |
250 |
6 |
251 |
7 |
252 |
8 |
253 |
9 |
254 |
10 |
255 |
11 |
256 |
12 |
257 |
13 |
258 |
14 |
259 |
15 |
260 |
16 |
261 |
17 |
262 |
18 |
263 |
19 |
264 |
20 |
265 |
21 |
266 |
22 |
267 |
23 |
268 |
24 |
269 |
25 |
270 |
26 |
271 |
27 |
272 |
28 |
273 |
29 |
274 |
30 |
275 |
31 |
276 |
32 |
277 |
33 |
278 |
34 |
279 |
Linearization table percentage values |
280 |
Lin% |
281 |
Linearization table lin. percentage values |
282 |
Unit, totalizer |
283 |
bbl (liq) |
284 |
Counting direction for negative flow |
285 |
The counting direction for negative flow can be defined with bidirectional flow |
286 |
Upward |
287 |
Downward |
288 |
Stop |
289 |
Cutoff |
290 |
Due to the low flow suppression, the output signal of the differential pressure transmitter is suppressed inspite of an existing differential pressure of '0' |
291 |
First HART value (PV) |
292 |
Selection, which measured value should be outputted as Primary Variable (PV) |
293 |
Second HART value (SV) |
294 |
Selection, which measured value should be outputted as Secondary Variable (SV) |
295 |
Third HART value (TV) |
296 |
Selection, which measured value should be outputted as Third Variable (TV) |
297 |
Fourth HART value (QV) |
298 |
Selection, which measured value should be outputted as Quaternary Variable (QV) |
299 |
Function check |
300 |
Defines, if the diagnostic message 'Function check' should be outputted or suppressed |
301 |
Maintenance required |
302 |
Defines, if the diagnostic message 'Maintenance required' should be outputted or suppressed |
303 |
Out of specification |
304 |
Defines, if the diagnostic message 'Out of specification' should be outputted or suppressed |
305 |
Min. current |
306 |
Min. current consumption of the sensor (exception: value in case of failure) |
307 |
3.8 mA |
308 |
4 mA |
309 |
Max. current |
310 |
Max. current consumption of the sensor (exception: value in case of failure) |
311 |
20.5 mA |
312 |
20 mA |
313 |
Output characteristic |
314 |
Selection of the current characteristics |
315 |
20 ... 4 ... 20 mA |
316 |
4 ... 20 ... 4 mA |
317 |
0 ... 100 % correspond to 4 ... 20 mA |
318 |
0 ... 100 % correspond to 20 ... 4 mA |
319 |
Failure mode |
320 |
Current consumption of the sensor in case of failure |
321 |
<= 3.6 mA |
322 |
>=21 mA |
323 |
Last valid measured value |
324 |
Device address |
325 |
Loop current mode |
326 |
Fixed current (4 mA) |
327 |
Analogue current output (4 ... 20 mA) |
328 |
Output mode |
329 |
Analogue current output |
330 |
Min. adjustment value |
331 |
Max. adjustment value |
332 |
Min. adjustment in % |
333 |
Percentage value for which the min. adjustment value is entered |
334 |
Max. adjustment in % |
335 |
Percentage value for which the max. adjustment value is entered |
336 |
Offset |
337 |
Zero (initial value) |
338 |
Span (final value) |
339 |
Electronic differential pressure |
340 |
Min. adjusted pressure corresponds to |
341 |
Max. adjusted pressure corresponds to |
342 |
Current output 0 % corresponds to |
343 |
Value of the output variable at 0 % |
344 |
Current output 100 % corresponds to |
345 |
Value of the output variable at 100 % |
346 |
Direct adjustment of the offset for compensation of the mounting position |
347 |
Density upper medium |
348 |
Requirement for the interface measurement is that both sensors are covered by the medium. The interface can only be detected between the two sensors. |
349 |
Density lower medium |
350 |
Distance of the sensors (A) |
351 |
The distance of the two sensors must be entered for determination of the interface out of the measured differential pressure |
352 |
The distance of the two sensors must be entered for determination of the density out of the measured differential pressure |
353 |
Hardware version |
354 |
Hardware version of the main electronics |
355 |
0 |
356 |
Software version of the main electronics |
357 |
Fieldbus Profile Rev. |
358 |
Measuring range (initial value) |
359 |
Measuring range (final value) |
360 |
Master |
361 |
Serial number Slave |
362 |
Measuring range slave (initial value) |
363 |
Measuring range slave (final value) |
364 |
Slave |
365 |
Actual electronics temperature |
366 |
Status electronics temperature |
367 |
Failure |
368 |
Upper limitation |
369 |
Lower limitation |
370 |
Simulation value |
371 |
Substitute value |
372 |
Constant value |
373 |
Status |
374 |
Status filling height |
375 |
Percentage value |
376 |
Status percent |
377 |
Lin. percentage value |
378 |
Status lin. percent |
379 |
Status scaled |
380 |
Status measuring cell temperature |
381 |
Status static pressure |
382 |
Status flow |
383 |
Status totalizer 1 |
384 |
Status totalizer 2 |
385 |
Status density |
386 |
Status height, interface |
387 |
Date of the min. electronics temperature |
388 |
Min. electronics temperature |
389 |
Date of the max. electronics temperature |
390 |
Max. electronics temperature |
391 |
Date, min. measured value |
392 |
Min. measured value |
393 |
Date, max. measured value |
394 |
Max. measured value |
395 |
Date of the min. measuring cell temperature value |
396 |
Min. measuring cell temperature |
397 |
Date of the max. measuring cell temperature value |
398 |
Max. measuring cell temperature |
399 |
Date of the min. static pressure value |
400 |
Min. static pressure |
401 |
Date of the max. static pressure value |
402 |
Max. static pressure |
403 |
Device status |
404 |
Device status acc. to NE 107 |
405 |
OK |
406 |
Malfunction |
407 |
Detail status |
408 |
Detailed status value for identification of the exact cause of error |
409 |
Status current output |
410 |
Status current output 2 |
411 |
HART Device Status |
412 |
Primary variable out of limits |
413 |
Non-primary variable out of limits |
414 |
Analog output saturated |
415 |
Analog output fixed |
416 |
More status available |
417 |
Cold start |
418 |
Configuration changed |
419 |
Field device malfunction |
420 |
Activity to be carried out |
421 |
Restore delivery status |
422 |
Restore basic settings |
423 |
Source measuring cell temperature |
424 |
System |
425 |
Diaphragm, master |
426 |
Rear, master |
427 |
Diaphragm, slave |
428 |
Rear, slave |
429 |
Source static pressure |
430 |
Display format |
431 |
#.##### |
432 |
#.###### |
433 |
#.####### |
434 |
#.######## |
435 |
Density unit |
436 |
Selection of the density unit for indication and parameter adjustment |
437 |
Measuring unit |
438 |
Selection of the flow unit for indication and parameter adjustment |
439 |
Designation of the unit |
440 |
Basic unit |
441 |
Basic unit for user-defined unit |
442 |
Factor |
443 |
Conversion factor relating to the basic unit |
444 |
SIL qualified sensor |
445 |
Unit static pressure |
446 |
Selection of the unit of the static pressure for indication and parameter adjustment |
447 |
Delay of fault signal |
448 |
d |
449 |
ms |
450 |
Invert the differential pressure |
451 |
Activate LRV |
452 |
Activate URV |
453 |
Reset calibration |
454 |
Password |
455 |
Simulation, current output |
456 |
Simulation, current output 2 |
457 |
Simulation, electronics temperature |
458 |
Simulation temperature |
459 |
Simulation active |
460 |
Simulation |
461 |
Simulation, percent |
462 |
Simulation, lin. percent |
463 |
Simulation, static pressure |
464 |
Sensor, pressure |
465 |
Simulation, flow |
466 |
Simulation, differential pressure |
467 |
Status differential pressure |
468 |
Density medium |
469 |
Activate thermoshock suppression |
470 |
Sensor calibration |
471 |
Min. measuring range |
472 |
Max. measuring range |
473 |
Electronics version |
474 |
Version of the connected supplementary electronics |
475 |
--- |
476 |
PS60HC |
477 |
PS60HK |
478 |
PS60HS |
479 |
PS60PAC |
480 |
PS60PAK |
481 |
PS60PAS |
482 |
PS60FFC |
483 |
PS60FFK |
484 |
PS60FFS |
485 |
PLICSZEBVL |
486 |
PLICSZEBVH |
487 |
PS60WLH |
488 |
PS60WLPA |
489 |
PS60WLFF |
490 |
FX80H |
491 |
FX80HSIL |
492 |
FX80PA |
493 |
FX80FF |
494 |
PLICSZEZSA |
495 |
BR80H |
496 |
BR80HSIL |
497 |
BR80PA |
498 |
BR80FF |
499 |
BR80Z |
500 |
BR80HK |
501 |
BR80HKSIL |
502 |
BR80PAK |
503 |
BR80FFK |
504 |
BR80ZK |
505 |
PS60HW |
506 |
PS60PAW |
507 |
PS60FFW |
508 |
FX80MB |
509 |
PS64HW |
510 |
BR18-2L |
511 |
BR18-2LE |
512 |
BR18-2LEB |
513 |
BR18-3L |
514 |
Character string of the instrument |
515 |
LONG-TAG |
516 |
MESSAGE |
517 |
Message stored in the instrument for free use |
518 |
Activate thermoshock suppression Master |
519 |
Activate thermoshock suppression Slave |
520 |
Simulation density |
521 |
Offset static pressure |
522 |
Display format 1 |
523 |
Selection of the requested formatting of the display value shown on the display and adjustment module |
524 |
Automatically |
525 |
Display format 2 |
526 |
Offset, differential pressure |
527 |
Setup |
528 |
Linearization |
529 |
Diagnostics |
530 |
Additional settings |
531 |
Measured values |
532 |
Info |
533 |
Figure: Setup |
534 |
Figure: Damping |
535 |
Figure: Linearization table |
536 |
Figure: Simulation |
537 |
Figure: Reset |
538 |
Figure: Scaling |
539 |
Bild: PV Abgleich |
540 |
Bild: SV Abgleich |
541 |
Figure: HART variables |
542 |
Figure: Differential pressure transducer |
543 |
Figure: Totalizer 1 |
544 |
Figure: Info |
545 |
Display |
546 |
Application |
547 |
Before changing the application, the data should be written into the device before starting the adjustment. |
548 |
Units of measurement |
549 |
Sensor mounting correction |
550 |
In case a position correction is requested, it must be saved in the device before adjustment. |
551 |
Sensor must be unpressurized! |
552 |
Adjustment |
553 |
Damping |
554 |
The sensor is in multidrop mode |
555 |
If the pressure transmitter is below the vessel bottom, then the value must be entered with negative sign. |
556 |
Low flow suppression |
557 |
Status signals |
558 |
The instrument has a self-monitoring acc. to NE 107. Define the diagnostic states to be outputted by the instrument. |
559 |
Measured value status |
560 |
Status, additional measured values |
561 |
Peak indicator measurement value |
562 |
Peak indicator static pressure |
563 |
Peak indicator, electronics temperature |
564 |
Peak indicator measuring cell temperature |
565 |
Reset |
566 |
Scaling |
567 |
PV adjustment |
568 |
The selected output variable is also assigned to the first HART value (PV). |
569 |
SV adjustment |
570 |
The selected output variable is also assigned to the second HART value (SV). |
571 |
HART variables |
572 |
Differential pressure transducer - characteristics values |
573 |
Service login |
574 |
Special parameters |
575 |
Outputs |
576 |
Additional measured values |
577 |
Read values from device |
578 |
Write values to device |
579 |
Linearization table |
580 |
Linearization curve |
581 |
Table |
582 |
Device characteristics |
583 |
Details |
584 |
Version |
585 |
Update instrument features |
586 |
Activate simulation |
587 |
Deactivate simulation |
588 |
Accept simulation value |
589 |
Identification |
590 |
Service login required |
591 |
Measured value memory (EDD) |
592 |
Parameter adjustment |
593 |
Maintenance |
594 |
Service logout |
595 |
Activate |
596 |
Automatic correction |
597 |
Reset? |
598 |
Reset 'Configuration changed' |
599 |
Reset peak indicator |
600 |
Reset totalizer |
601 |
VEGAPULS 61 |
602 |
VEGAPULS 62 |
603 |
VEGAPULS 63 |
604 |
VEGAPULS 64 |
605 |
VEGAPULS 65 |
606 |
VEGAPULS 66 |
607 |
VEGAPULS 67 |
608 |
VEGAPULS 68 |
609 |
VEGAPULS WL61 |
610 |
VEGAPULS SR68 |
611 |
VEGAPULS 69 |
612 |
VEGAFLEX 80 series |
613 |
VEGAFLEX 80 series SIL |
614 |
VEGABAR 80 series |
615 |
VEGABAR 80 series SIL |
616 |
VEGADIF 80 series SIL |
617 |
VEGA Grieshaber KG |
618 |
Siemens |
619 |
KROHNE Messtechnik GmbH |
620 |
WIKA |
621 |
Balluff GmbH |
622 |
Communication error |
623 |
Briefly separate operating voltage, exchange device |
624 |
Impermissible electronics temperature |
625 |
Check ambient temperature, isolate electronics |
626 |
Error in the calibration |
627 |
Exchange electronics or return device for repair |
628 |
Error in the device setting |
629 |
Repeat setup or reset |
630 |
Installation/setup error |
631 |
Check installation and/or parameter adjustment |
632 |
Measurement function disturbed |
633 |
Carry out a reset or briefly separate operating voltage |
634 |
Impermissible operating voltage |
635 |
Check electrical connections |
636 |
No connection to the slave |
637 |
Check connection between master and slave |
638 |
No measured value available |
639 |
Adjusted span too small |
640 |
Change adjustment according to the limit values |
641 |
Error in the linearization table |
642 |
Check linearization table |
643 |
No executable software |
644 |
Repeat update, exchange electronics or device |
645 |
Error in the electronics |
646 |
Pressure transmitter stopped answering |
647 |
Check pressure transmitter and exchange, if necessary |
648 |
General software error |
649 |
Briefly separate operating voltage |
650 |
Measured value is determined |
651 |
Wait for switch-on phase |
652 |
Check ambient temperature, isolate electronics or use a device with higher temperature range |
653 |
Error in the delivery status |
654 |
Repeat reset or load XML file with sensor data into the sensor |
655 |
Error in the non-active linearization table |
656 |
Error in the event memory |
657 |
Error on a device interface |
658 |
Check the connections, exchange the electronics or return the device for repair |
659 |
Carry out a reset and repeat setup |
660 |
Distance to level |
661 |
Sensor value |
662 |
Unlock adjustment |
663 |
Verify and lock adjustment |
664 |
Verify adjustment |
665 |
Lock adjustment |
666 |
Deactivate |
667 |
Execute action |
668 |
Common Tables |
669 |
Interoperable * |
670 |
Universal |
671 |
Common Practice |
672 |
Pressure Universal |
673 |
Pressure Common |
674 |
Menu |
675 |
Rosemount Magflow |
676 |
Rosemount Pressure |
677 |
Moore Products |
678 |
Siemens Pressure |
679 |
P2 Common |
680 |
PV |
681 |
Wireless |
682 |
Munge |
683 |
Analytical Oxygen/Combustion/Methane/CO |
684 |
Flow Totalizer - Rate Indicator |
685 |
Flow Variable Area |
686 |
Flow Coriolis Mass |
687 |
Flow Ultrasonic - Gas |
688 |
Flow Ultrasonic - Liquid |
689 |
Level Ultrasonic |
690 |
Level Radar - Microwave |
691 |
Pressure Absolute |
692 |
Level Capacitance |
693 |
Pressure Differential |
694 |
Level Hydrostatic |
695 |
Level |
696 |
Pressure Absolute/Relative |
697 |
Level Microwave |
698 |
Level Radar |
699 |
Level Density - Nuclear |
700 |
Flow Electromagnetic |
701 |
Flow Vortex |
702 |
Flow Magmeter - Rack Version |
703 |
Flow High Performance Magmeter |
704 |
Flow Mass |
705 |
Flow Ultrasonic |
706 |
Flow Magnetic |
707 |
Flow Magnetic - Custody Transfer |
708 |
Flow 2-wire Magnetic |
709 |
Flow Coriolis Mass - Custody Transfer |
710 |
Flow 2-wire Vortex |
711 |
Flow Thermal |
712 |
Temperature Head Mounted |
713 |
Temperature Compact |
714 |
Temperature DIN Rail Mounted |
715 |
Analytical pH |
716 |
Analytical Conductivity - Conductive |
717 |
Analytical Conductivity - Inductive |
718 |
Temperature Sensor |
719 |
Level Servo |
720 |
Temperature DIN Rail |
721 |
Temperature |
722 |
Analytical ORP |
723 |
Analytical pION |
724 |
Analytical IConc |
725 |
Analytical Electrode Conductivity |
726 |
Analytical Toroidal Conductivity |
727 |
Valve Positioner |
728 |
Valve Controller |
729 |
Pressure |
730 |
Pressure Multivariable |
731 |
Analytical pH/ORP |
732 |
Flow Coriolis |
733 |
Analytical Conductivity |
734 |
Valve Positioner/Pneumatic |
735 |
Valve Actuator - Electrical |
736 |
Temperature Dual Input |
737 |
Valve Actuator |
738 |
Analytical Corrosion |
739 |
Analytical Density - Nuclear |
740 |
Level Continuous Gamma |
741 |
Level Laser Ranging |
742 |
Pressure Safety |
743 |
Flow Vortex/Swirl |
744 |
Flow |
745 |
Analytical 4 Electrode Conductivity |
746 |
Analytical 2 Electrode Conductivity |
747 |
Analytical Troidal Conductivity |
748 |
Analytical pH/ORP/pIon/Ion Conc |
749 |
Analytical Oxygen |
750 |
Pressure P, dP + Multivariable |
751 |
Temperature 2-wire |
752 |
Pressure w/ PID |
753 |
Pressure Saturated Steam |
754 |
Analytical Density - Gamma Ray |
755 |
Level Gamma Ray |
756 |
Analytical Weight - Gamma Ray |
757 |
Level w/ Density - Gamma Ray |
758 |
Analytical Motorized Density |
759 |
Flow Multivariable Mass |
760 |
Temperature Multivariable |
761 |
Specialty HART-to-Analog Signal Conv |
762 |
Flow dP |
763 |
Pressure Low Power |
764 |
Pressure Certified SIL2 |
765 |
Interoperable |
766 |
Analytical Density |
767 |
Valve Positioner/Electropneumatic |
768 |
Flow Magnetic Inductive |
769 |
Analytical |
770 |
Analytical Opacity Monitor |
771 |
Analytical Oxygen/Combustion |
772 |
Analytical 2-wire Conductivity |
773 |
Analytical 2-wire Conductivity w/ pure H20 |
774 |
Analytical 2-wire Conductivity w/ % conc |
775 |
Analytical 2-wire Oxygen/Chlorine/Ozone |
776 |
Analytical 2-wire pH/ORP |
777 |
Analytical 2-wire Contacting/Toroidal |
778 |
Analytical Oxygen - High Temperature |
779 |
Analytical Conductivity w/ % conc |
780 |
Analytical pH/ORP/ISE |
781 |
Analytical Amperometric Analyzer |
782 |
Analytical Dual Input Analyzer |
783 |
Pressure Pressure/Level |
784 |
Analytical Consistency |
785 |
Analytical Microwave-Pulp Consistency |
786 |
Analytical Microwave Consistency |
787 |
Valve Digital Positioner |
788 |
Valve Smart Valve Positioner |
789 |
Temperature Differential |
790 |
Analytical Dissolved Oxygen |
791 |
Analytical pH Controller |
792 |
Analytical Conductivity/Resistivity Controller |
793 |
Analytical Conductivity/Resistivity |
794 |
Level Displacement |
795 |
Temperature w/ PID |
796 |
Valve Universal Positioner |
797 |
Level Radar - Reflex |
798 |
Pressure Gauge |
799 |
Pressure Sealed Gauge |
800 |
Valve Digital Positioner w/ PID |
801 |
Valve Intelligent Positioner |
802 |
Analytical Gas Detector |
803 |
Analytical Infrared Gas Detector |
804 |
Analytical Electrochemical Gas Detector |
805 |
Analytical Measuring Instruments |
806 |
Level Magnostrictive |
807 |
Valve On/Off Valve Controller |
808 |
Level Radar - Pulse |
809 |
Level Continuous Capactive |
810 |
Valve Damper Actuator |
811 |
Valve Electric Damper Actuator |
812 |
Flow Electronic |
813 |
Flow Vortex - Multiparameter Mass |
814 |
Analytical UV Flame Detector |
815 |
Analytical UVIR Flame Detector |
816 |
Analytical IR Flame Detector |
817 |
Analytical Fire Detector |
818 |
Analytical Gas Detector Display Interface |
819 |
Analytical Toxic Gas Detector |
820 |
Analytical Flame Detector |
821 |
Level Nuclear |
822 |
Analytical Microwave |
823 |
Analytical Density - Radiometric |
824 |
Level Radiometric |
825 |
Analytical Gas/Steam |
826 |
Level Nucleonic |
827 |
Analytical Blade Consistency |
828 |
Analytical Rotary Consistency |
829 |
Pressure Absolute/Differential |
830 |
Positioner |
831 |
Analytical Combustible Gas Detector |
832 |
Analytical Point Infrared Gas Detector |
833 |
Analytical Multi-Detector Monitor System |
834 |
Valve Positioner/Electro-hydraulic |
835 |
Valve Monitor |
836 |
Enable on token-passing data link layer only |
837 |
Enable on TDMA data link layer only |
838 |
Enable on both TDMA and token-passing data link layer |
839 |
WirelessHART Device |
840 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
841 |
Primary |
842 |
Primary Analog Channel |
843 |
Secondary |
844 |
Secondary Analog Channel |
845 |
Tertiary |
846 |
Tertiary Analog Channel |
847 |
Quaternary |
848 |
Quaternary Analog Channel |
849 |
Quinary |
850 |
Quinary Analog Channel |
851 |
Critical Power Failure |
852 |
Lock all |
853 |
Trim not supported |
854 |
Configuration cannot be changed |
855 |
Locked by gateway |
856 |
Analog Input Channel |
857 |
Analog Output Channel |
858 |
Secondary Analog Channel Saturated |
859 |
Tertiary Analog Channel Saturated |
860 |
Quaternary Analog Channel Saturated |
861 |
Quinary Analog Channel Saturated |
862 |
Secondary Analog Channel Fixed |
863 |
Tertiary Analog Channel Fixed |
864 |
Quaternary Analog Channel Fixed |
865 |
Quinary Analog Channel Fixed |
866 |
Simulation active |
867 |
The device is in simulation mode and the level output is not representative of the process. |
868 |
Non-Volatile memory failure |
869 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
870 |
Volatile memory error |
871 |
The RAM memory check is invalid or maybe corrupt |
872 |
Watchdog reset executed |
873 |
A watchdog reset has been performed |
874 |
Voltage conditions out of range |
875 |
A voltage condition is outside its allowable range |
876 |
Environmental conditions out of range |
877 |
An internal or environmental condition is beyond acceptable limits. |
878 |
Electronic failure |
879 |
A hardware problem not related to the sensor has been detected. |
880 |
Subdevice list changed |
881 |
Duplicate master detected |
882 |
Capacity Denied |
883 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
884 |
Bandwidth allocation pending |
885 |
Block transfer pending |
886 |
Continuous |
887 |
The Burst Message is published continuously at (worst case) the Minimum Update Period. |
888 |
Window |
889 |
The Burst Message is triggered when the source value deviates more than the specified trigger value |
890 |
Rising |
891 |
The Burst Message is triggered when source value Rises Above the specified trigger value. |
892 |
Falling |
893 |
The Burst Message is triggered when the source value Falls Below the specified trigger value. |
894 |
On-change |
895 |
The Burst Message is triggered when any value in the message changes. |
896 |
Battery life |
897 |
Percent range |
898 |
Loop current |
899 |
Primary variable |
900 |
Secondary variable |
901 |
Tertiary variable |
902 |
Quaternary variable |
903 |
Off |
904 |
Enable event notification on token-passing data link layer |
905 |
Enable event notification on TDMA data link layer |
906 |
Enable event notification on TDMA and token-passing data link layer |
907 |
Configuration changed event pending |
908 |
Device status event pending |
909 |
More status available event pending |
910 |
Disable |
911 |
Enable single data point trending |
912 |
Enable filtered trending |
913 |
The sample is filtered using a time constant equal to one-third the trend sample period |
914 |
Enable average trending |
915 |
The value obtained by dividing the sum of all samples in the period by the total number of samples |
916 |
Read receive time |
917 |
Write date and time |
918 |
Non-Volatile Clock |
919 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
920 |
Clock Uninitialized |
921 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
922 |
Line Power |
923 |
Device is powered via a hardwired connection to a plant power source. |
924 |
Battery Power |
925 |
Device operates solely of its (internal) battery. |
926 |
Recharg.B./Power Sc. |
927 |
Rechargeable battery power or power scavenging. The device has a short-term supply of energy (must be at least one hour) that is continuously being replenished. The device's power is being replenished by harvesting and converting energy from the environment surrounding the device (e.g. solar, vibration, heat). |
928 |
Idle |
929 |
Active search |
930 |
Looking for neighbors. |
931 |
Negotiating |
932 |
Quarantined |
933 |
Operational |
934 |
Suspended |
935 |
Deep sleep |
936 |
Ultra-low power |
937 |
Wireless Signal Found |
938 |
Network packets have been recognized |
939 |
Wireless Signal Identified |
940 |
Wireless Time Synchronized |
941 |
WirelessHart Signal Identified |
942 |
Network Admission Requested |
943 |
Set on transmission of first join request |
944 |
Join Retrying |
945 |
Set after the first join request retry (i.e. when number of join requests > 2). Cleared when device is authenticated or when Active Search started. |
946 |
Join Failed |
947 |
Set on transition from Active Search mode to Deep Sleep mode. Cleared on transition from Deep Sleep to Active Search. |
948 |
Network Security Clearance Granted |
949 |
Network Manager Session Established |
950 |
Network Joined |
951 |
Network Joined. Normal superframe and links obtained. |
952 |
Network Bandwidth Requested |
953 |
Negotiating Network Properties. Gateway session obtained. |
954 |
Join Complete |
955 |
Normal operating mode. |
956 |
None |
957 |
No restrictions |
958 |
SI Only |
959 |
Restricted to SI Unit codes only |
960 |
Open - accept all devices |
961 |
Use Whitelist for access |
962 |
Use Blacklist |
963 |
Use Whitelist and Blacklist |
964 |
Lockdown - disable join links |
965 |
Process automation device |
966 |
Discrete device |
967 |
Hybrid:Process automation+discrete |
968 |
I/O System |
969 |
WirelessHART Process automation device |
970 |
WirelessHART Discrete device |
971 |
WirelessHART Process automation+discrete |
972 |
WirelessHART Gateway |
973 |
WirelessHART Access point |
974 |
WirelessHART Process adapter |
975 |
WirelessHART Discrete adapter |
976 |
WirelessHART enabled handheld/portable maintenance tool |
977 |
Don't attempt to join |
978 |
Join now |
979 |
Attempt to join immediately on powerup or reset |
980 |
US |
981 |
United States of America |
982 |
JP |
983 |
Japan |
984 |
DE |
985 |
Germany |
986 |
FR |
987 |
France |
988 |
ES |
989 |
Spain |
990 |
RU |
991 |
Russian Federation |
992 |
CN |
993 |
People's Republic of China |
994 |
Nominal |
995 |
Device Power Status is nominal |
996 |
Low |
997 |
Device Power Status is low |
998 |
Critically Low |
999 |
Device Power Status is critically low |
1000 |
Recharging - Low |
1001 |
Recharging - High |
1002 |
Secondary Master |
1003 |
Primary Master |
1004 |
PV Analog Channel Saturated |
1005 |
PV Analog Channel Fixed |
1006 |
Device Profile |
1007 |
Device Diagnostic Status 0 |
1008 |
Device Diagnostic Status 1 |
1009 |
I/O and Subdevice Status |
1010 |
WirelessHART Status |
1011 |
Time Set Option |
1012 |
Real Time Clock Flags |
1013 |
Country |
1014 |
Country of intended device installation |
1015 |
SI Unit Control |
1016 |
Indicates if device should only display SI unit codes |
1017 |
Burst Trigger Units |
1018 |
Burst Trigger Class |
1019 |
Wired HART Enabled |
1020 |
Enabled on Wired HART device connection only |
1021 |
Wireless HART Enabled |
1022 |
Enabled on WirelessHART device connection only |
1023 |
Both Wired and WirelessHart Enabled |
1024 |
Enabled on both Wired and WirelessHART connections |
1025 |
Burst Msg Trigger Mode |
1026 |
Event Status |
1027 |
Event Notification Control |
1028 |
Operate mode Mask |
1029 |
Not Defined |
1030 |
Power Source |
1031 |
Power source of the device |
1032 |
Wireless Mode |
1033 |
Join Status |
1034 |
The current location in the joining sequence of operations. |
1035 |
Network Access Mode |
1036 |
Join Mode |
1037 |
Power Status |
1038 |
Power status of the device |
1039 |
Master Mode |
1040 |
This sets the I/O System's Data Link Layer operation on all channels as either primary or secondary master mode. |
1041 |
Trend Units |
1042 |
Trend Device Variable Unit Code |
1043 |
Trend Classification |
1044 |
Trend Device Variable Classification |
1045 |
Trend Control |
1046 |
Device Status Mask |
1047 |
Ext dev status Mask |
1048 |
Device Diagnostic Status 0 Mask |
1049 |
Device Diagnostic Status 1 Mask |
1050 |
Device Diagnostic Status 2 Mask |
1051 |
Device Diagnostic Status 3 Mask |
1052 |
AO saturated Mask |
1053 |
AO fixed Mask |
1054 |
Latched Device Status |
1055 |
Latched Ext dev status |
1056 |
Latched Operate Mode |
1057 |
Latched Device Diagnostic Status 0 |
1058 |
Latched Device Diagnostic Status 1 |
1059 |
Latched Device Diagnostic Status 2 |
1060 |
Latched Device Diagnostic Status 3 |
1061 |
Latched AO saturated |
1062 |
Latched AO fixed |
1063 |
Wireless Module Device Type |
1064 |
Device type of the Wireless Radio |
1065 |
Wireless Module Manufacturer ID |
1066 |
Manufacturer of the Wireless Radio contained in this device |
1067 |
Subdevice Device Type |
1068 |
Device type of the subdevice attached to this system. |
1069 |
Subdevice Manufacturer Id |
1070 |
Manufacturer of the subdevice attached to this system. |
1071 |
Time Stamp |
1072 |
Relative time value on network |
1073 |
8u |
1074 |
Dynamic Variables Returned for Device Variables |
1075 |
Command Response Truncated |
1076 |
Configuration Change Counter Mismatch |
1077 |
Status bytes mismatch |
1078 |
8.3f |
1079 |
Features were updated. |
1080 |
The instrument is locked! The action cannot be executed. |
1081 |
The password is wrong! |
1082 |
Do you really want to activate the measurement with electronic differential pressure? When switching over, some settings in the instrument will be reset. |
1083 |
Action successfully carried out. |
1084 |
Do you really want to deactivate the measurement with electronic differential pressure? When switching over, some settings in the instrument will be reset. |
1085 |
Do you want to carry out this action? |
1086 |
Do you really want to reset the selected peak indicators? |
1087 |
Do you really want to reset the selected totalizer? |
1088 |
%{ii}/%{varVerificationTextCount_Count}
|
1089 |
%{varVerificationName_VerificationName}
|
1090 |
%{varVerificationValue_VerificationValue}
|
1091 |
Parameter OK? |
1092 |
; |
1093 |
Number of the modified, safety-relevant parameters |
1094 |
: %{varVerificationTextCountString_TextCount}
|
1095 |
Do the parameter values and the number of parameters correspond to the actually modified parameters? |
1096 |
Number of the modified, non-safety-relevant parameters |
1097 |
When the function 'Verify and lock' is executed, instrument adjustment is locked. The safety instrumented function is thus ensured. |
1098 |
To lock the adjustment with the actual PIN, you have to use the 'Finish' key. To modify the actual PIN, you have to enter a new four digit number code. |
1099 |
|
1100 |
pin |
1101 |
Wrong PIN |
1102 |
%{varVerifyString_VerifyString}
|
1103 |
Expected character string |
1104 |
|
1105 |
'One point two three plus four point five six minus seven eight nine point zero' |
1106 |
Do the two character sequences match? |
1107 |
Serial number of the instrument |
1108 |
: %{varSerialNumberString_SerialNumber}
|
1109 |
Does the displayed serial number match with the number of the device? |
1110 |
A function test is required. The procedure is described in the operating instructions. |
1111 |
Was the function test successful? |
1112 |
Not yet terminated |
1113 |
Enter the actually valid PIN |