0 |
2000 Mass flo |
1 |
Tag |
2 |
Date- Gregorian calendar date, entered as mm/dd/yy, that is stored in the Field Device. This date can be used by the user in any way. There is no specific recommended use. |
3 |
No |
4 |
4.1f |
5 |
EEPROM Checksum - Core processor |
6 |
RAM Error - Core Processor |
7 |
Temperature out of range |
8 |
Data loss possible |
9 |
Burst mode enabled |
10 |
Event 1 triggered |
11 |
Event 2 triggered |
12 |
Sensor/Xmtr Communication Error |
13 |
Calibration Factor Unentered (Flocal) |
14 |
EEPROM Checksum - 2000 |
15 |
RAM Error - 2000 |
16 |
Unrecognized/Unentered Sensor Type (K1) |
17 |
EEPROM Config. Corrupt - Core Processor |
18 |
EEPROM Totals Corrupt - Core Processor |
19 |
EEPROM Program Corrupt - Core Processor |
20 |
Line RTD Overrange |
21 |
Case RTD Overrange |
22 |
Flow is in reverse direction |
23 |
Temp. Slope Cal in Progress |
24 |
Temp. Zero Cal in Progress |
25 |
Density FD Cal in Progress |
26 |
Density 2nd Point Cal in Progress |
27 |
Density 1st Point Cal in Progress |
28 |
Mech. Zero Cal in Progress |
29 |
Core Processor Boot Sector Fault |
30 |
Series 2000 software upgrade recommended |
31 |
Freq. Output Fixed |
32 |
Density 3rd Point Cal in Progress |
33 |
Density 4th Point Cal in Progress |
34 |
PV is |
35 |
SV is |
36 |
TV is |
37 |
QV is |
38 |
Off |
39 |
On |
40 |
Scaled AO trim |
41 |
Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. |
42 |
Mass flo |
43 |
.1f |
44 |
.2f |
45 |
.3f |
46 |
.4f |
47 |
.5f |
48 |
The mass flow rate below which the frequency/pulse output indicates zero flow, and below which totalizers stop counting. |
49 |
4.2f |
50 |
% |
51 |
Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. |
52 |
7d |
53 |
8.2f |
54 |
The number of cycles, in Hertz, that equal a proportional flow rate represented by the frequency/pulse output. |
55 |
The flow rate that equals a set frequency if the frequency output represents mass or volume flow; or the mass or volume total that equals a set number of pulses if the frequency output represents mass or volume total. |
56 |
This parameter allows you to scale the frequency output in terms of the number of pulses per unit of flow. |
57 |
|
58 |
Pulses/Unit |
59 |
8.3f |
60 |
This parameter allows you to scale the frequency output in terms of the number of units of flow required to generate one pulse. |
61 |
Units/Pulse |
62 |
8.6f |
63 |
Unit |
64 |
AO cutoff |
65 |
Temp |
66 |
SGU |
67 |
Dens |
68 |
.6f |
69 |
Vol flo |
70 |
The volume flow rate below which the frequency/pulse output indicates zero flow, and below which totalizers stop counting. |
71 |
10.4f |
72 |
PV |
73 |
QV |
74 |
Configuration information identifying the flange material used with the sensor. Informational purposes only - does not affect flowmeter operation. |
75 |
Flange |
76 |
Configuration information identifying any liner material used with the sensor. Informational purposes only - does not affect flowmeter operation. |
77 |
Halar |
78 |
Configuration information identifying the sensor model installed. Informational purposes only - does not affect flowmeter operation. |
79 |
Configuration information identifying the material of which the sensor is constructed. Informational purposes only - does not affect flowmeter operation. |
80 |
A standard engineering unit that serves as the mass component in a special unit. |
81 |
Base mass unit |
82 |
A standard unit of time (seconds, minutes, hours, or days) that serves as the time component in a special unit. |
83 |
Base mass time |
84 |
A programmable multiplier for the base mass unit in a special unit. |
85 |
Mass flo conv factr |
86 |
A 1 to 8 character description of a special unit of flow. |
87 |
Mass flo text |
88 |
A 1 to 8 character description of a special unit of mass, which is based on the special unit of flow to represent total. |
89 |
Mass totl text |
90 |
This paramater enables the user to determine whether the outputs and totalizers will indicate forward, reverse, bi-directional or absolute value flow. |
91 |
Absolute Value |
92 |
A programmable high limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. |
93 |
Slug high limit |
94 |
5.4f |
95 |
A programmable low limit on density, which enables the transmitter to indicate slug flow (extremes in fluid density caused by slugs of gas or accumulation of solids in a liquid flow stream) if fluid density goes outside programmed slug flow limits. |
96 |
Slug low limit |
97 |
A programmable time period, from 0.00 to 60.00 seconds, during which outputs representing flow hold their last measured value before indicating slug flow. |
98 |
Slug duration |
99 |
5.2f |
100 |
A 10-character factor that describes the sensitivity of the sensor to mass flow. The first 6 characters are the flow rate, in grams/second, per microsecond of time shift between velocity signals from the sensor. The 2nd four characters are the flow temperature coefficient. |
101 |
FlowCal |
102 |
FCF |
103 |
The mass or volume flow rate below which a milliamp output representing flow indicates zero flow. |
104 |
PV AO cutoff |
105 |
A standard engineering unit that serves as the volume component in a special unit. |
106 |
Base vol unit |
107 |
Base vol time |
108 |
A programmable multiplier for the base volume unit in a special unit. |
109 |
Vol flo conv factr |
110 |
A 1 to 8 character description of a special unit of flow. |
111 |
Vol flo text |
112 |
A 1 to 8 character description of a special unit of volume, which is based on the special unit of flow to represent total. |
113 |
Vol totl text |
114 |
A value (in Hertz) that will be indicated on the frequency output for the purpose of a loop test. |
115 |
Allows the user to Reset, Stop, or Start all totalizers. NOTE - Inventory totals are NOT reset. |
116 |
This parameter allows you to scale the frequency output by using freq=flow, pulses/unit, or units/pulse. This parameter must be sent to the device before you make scaling changes. |
117 |
FO Scale Method |
118 |
Freq=Flow |
119 |
This parameter represents the Density Temperature Gradient Coefficient (DTG) that is on the T-series sensor tag. |
120 |
DTG |
121 |
6.2f |
122 |
This parameter represents the Density Frequency Coefficient #1 (DFQ1) that is on the T-series sensor tag. |
123 |
DFQ1 |
124 |
This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. |
125 |
DFQ2 |
126 |
This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. |
127 |
FTG |
128 |
This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. |
129 |
FFQ |
130 |
The number of readings collected from the sensor during an Auto Zero operation. |
131 |
Zero time |
132 |
5d |
133 |
A programmable time period, in seconds, that defines the crossover frequency for the frequency output, where Crossover freq. = 1 / (2 x Pulse width). Below the crossover, the output has a constant ON state, and an OFF state that varies with frequency. |
134 |
Max pulse width |
135 |
5.3f |
136 |
A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. |
137 |
Temp cal factr |
138 |
A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. |
139 |
uSec |
140 |
Units for zero offset - always microseconds. |
141 |
Zero units |
142 |
1d |
143 |
A unitless value which characterizes the amount of deviation between readings during an Auto Zero operation. |
144 |
Std dev |
145 |
Output Type |
146 |
This parameter defines the number of seconds for which all outputs will hold their last measured value prior to going to the configured fault state. The default is zero seconds. |
147 |
Fault Timeout |
148 |
2d |
149 |
AO Fault Setup |
150 |
FO Fault Setup |
151 |
Dig. Comm Fault Setup |
152 |
Downscale: Milliamp outputs that indicate flow go to the configured limit, the frequency output equals 0 Hz. Upscale: Milliamp and frequency outputs go to the configured limit. Internal zero: Flow outputs indicate zero flow. |
153 |
AO Fault indicator |
154 |
Upscale |
155 |
Downscale |
156 |
FO Fault indicator |
157 |
This fault setting applies to Modbus on RS485. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected) |
158 |
Comm Fault ind. |
159 |
IntZero-All 0 |
160 |
Not-a-Number |
161 |
IntZero-Flow 0 |
162 |
This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. |
163 |
mA Fault Value |
164 |
This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. |
165 |
FO Fault Value |
166 |
6.5f |
167 |
User-entered temperature for the purpose of performing an on-line temperature calibration. |
168 |
Ref temperature |
169 |
The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. |
170 |
D1 |
171 |
The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. |
172 |
D2 |
173 |
The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. |
174 |
D3 |
175 |
The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. |
176 |
D4 |
177 |
The tube period, in microseconds, corresponding to a density equal to D1. |
178 |
K1 |
179 |
The tube period, in microseconds, corresponding to a density equal to D2. |
180 |
K2 |
181 |
The tube period, in microseconds, corresponding to a density equal to D3. |
182 |
K3 |
183 |
The tube period, in microseconds, corresponding to a density equal to D4. |
184 |
K4 |
185 |
The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. |
186 |
Temp coeff |
187 |
DT |
188 |
Event 1 |
189 |
Event 2 |
190 |
Value |
191 |
Event num |
192 |
Event1 type |
193 |
High Alarm |
194 |
Low Alarm |
195 |
Event2 type |
196 |
A programmable value of a process variable at which an output set to represent event 1 or event 2 switches states. |
197 |
Event1 setpoint |
198 |
Event2 setpoint |
199 |
Assigns a process variable to event 1 or event 2. |
200 |
Event1 var |
201 |
Event2 var |
202 |
The frequency, in Hertz, presently being indicated on the frequency output. |
203 |
Pres freq |
204 |
6.1f |
205 |
A value, between 0.8 and 1.2, that will act as a multiplier for the mass flow rate. |
206 |
Mass factor |
207 |
A value, between 0.8 and 1.2, that will act as a multiplier for the volume flow rate. |
208 |
Vol factor |
209 |
A value, between 0.8 and 1.2, that will act as a multiplier for the density. |
210 |
Dens factor |
211 |
The tube frequency of the sensor measured in Hertz. |
212 |
Tube |
213 |
The pickoff voltage of the sensor measured in Voltage. |
214 |
V |
215 |
LPO |
216 |
RPO |
217 |
The drive gain of the transmitter. |
218 |
Drive |
219 |
This flow value is not effected by the flow cutoff and is damped at 8 seconds. It is to be used for diagnostic purposes only and is not used in ANY flow calculations or totalization. |
220 |
Live zero flow |
221 |
The board temperature of the core processor. |
222 |
Board Temperature |
223 |
The case temperature of the T-Series sensor. |
224 |
Case Temp: T-Series |
225 |
A calibration constant generated during a flowing density calibration. |
226 |
FD |
227 |
7.2f |
228 |
When this parameter is enabled, you can use the display to reset the totalizer. |
229 |
Disp. Total Reset |
230 |
Disabled |
231 |
Enabled |
232 |
When this parameter is enabled, you do not need to press the scroll button to scroll through the configured process variables. |
233 |
Disp. Auto Scroll |
234 |
This paramater allows you to set the rate at which the display will scroll from one process variable to the next. Display Auto Scroll must be enabled. |
235 |
Disp. scroll rate |
236 |
When this parameter is enabled, you can use the display to access the offline maintenance menu. |
237 |
Disp. Offline Menu |
238 |
When this parameter is enabled, you can use the display to acknowledge and view the specific alarm conditions that are present. When this parameter is disabled, all alarms will be automatically acknowledged without user intervention. |
239 |
Disp. Alarm Menu |
240 |
When this parameter is enabled, an option in the alarm menu allows you to acknowledge all of the alarms at once. |
241 |
Disp. ACK All |
242 |
When this parameter is enabled you must enter a password to gain access to the offline menu. |
243 |
Disp. Offline Pswd |
244 |
This parameter defines the password that you must enter to gain access to the offline menu. The valid range is 0 to 9999. |
245 |
Offline Password |
246 |
4d |
247 |
This parameter allows you to select the process variables that will appear on the display. Up to 15 process variables can be configured to appear on the display. They can be accessed by using the scroll button. |
248 |
Disp. Var #1 is |
249 |
Tube Frequency |
250 |
Drive Gain |
251 |
Input Voltage |
252 |
Disp. Var #2 is |
253 |
Disp. Var #3 is |
254 |
Disp. Var #4 is |
255 |
Disp. Var #5 is |
256 |
Disp. Var #6 is |
257 |
Disp. Var #7 is |
258 |
Disp. Var #8 is |
259 |
Disp. Var #9 is |
260 |
Disp. Var #10 is |
261 |
Disp. Var #11 is |
262 |
Disp. Var #12 is |
263 |
Disp. Var #13 is |
264 |
Disp. Var #14 is |
265 |
Disp. Var #15 is |
266 |
This parameter defines the protocol that will be used on the RS485 physical layer. |
267 |
RS485 Protocol |
268 |
HART |
269 |
Modbus RTU |
270 |
Modbus ASCII |
271 |
This parameter defines the baud rate for the protocol that will be used on the RS485 physical layer. |
272 |
RS485 Baudrate |
273 |
1200 |
274 |
2400 |
275 |
4800 |
276 |
9600 |
277 |
19200 |
278 |
38400 |
279 |
This parameter defines the parity for the protocol that will be used on the RS485 physical layer. |
280 |
RS485 Parity |
281 |
Odd |
282 |
Even |
283 |
This parameter defines the number of stop bits for the protocol that will be used on the RS485 physical layer. |
284 |
RS485 Stop Bits |
285 |
This paramater indicates which type of sensor is being used for process measurement. |
286 |
Sensor Type |
287 |
Other |
288 |
T-Series |
289 |
This parameter represents the various types of output option boards that are available. |
290 |
Output Opt. Brd. |
291 |
Analog I/O |
292 |
Fieldbus (H1) |
293 |
Anlog 1 range vals |
294 |
View fld dev vars |
295 |
View output vars |
296 |
Test/Status |
297 |
Test Points |
298 |
Freq scaling |
299 |
Charize sensor |
300 |
Config fld dev vars |
301 |
Config outputs |
302 |
Config events |
303 |
Display Setup |
304 |
View PV-Analog 1 |
305 |
View SV |
306 |
View TV-Frequency |
307 |
View QV |
308 |
View event 1 |
309 |
View event 2 |
310 |
Auto zero |
311 |
Density cal |
312 |
FO Scaling |
313 |
Flow |
314 |
Density |
315 |
Meter factors |
316 |
Flow |
317 |
Density |
318 |
Temperature |
319 |
Analog output 1 |
320 |
Frequency output |
321 |
RS485 Setup |
322 |
Event 1 |
323 |
Event 2 |
324 |
Enable/Disable |
325 |
Dens Pt1 (Air) |
326 |
Dens Pt2 (Water) |
327 |
Dens Pt3 T-series |
328 |
Dens Pt4 T-series |
329 |
Flowing Dens (FD) |
330 |
Spcl mass units |
331 |
Spcl vol units |
332 |
Device info |
333 |
Fld dev vars |
334 |
Outputs |
335 |
Write variables |
336 |
Sensor Selection |
337 |
This selection must be made to enable calibration parameters to be visible for the selected sensor type. |
338 |
Perform auto zero |
339 |
Please refer to manual for complete calibration instructions. |
340 |
Perform Cal |
341 |
Please refer to manual for complete calibration instructions. |
342 |
Fix frequency out |
343 |
Outputs a user-programmed frequency to the frequency output. |
344 |
Fix Analog Out 1 |
345 |
Trim Analog Out 1 |
346 |
Instructs transmitter to perform internal self test. |
347 |
Start totalizer |
348 |
Enables the internal totalizers to begin counting. |
349 |
Stop totalizer |
350 |
Prevents the internal totalizers from counting. |
351 |
Reset all totals |
352 |
Resets internal totalizers. |
353 |
Reset mass total |
354 |
Reset volume total |
355 |
Status Event1 |
356 |
When set, indicates that Event 1 has reached its programmed setpoint. |
357 |
Status Event2 |
358 |
When set, indicates that Event 2 has reached its programmed setpoint. |
359 |
View status |
360 |
Reports any currently active device status information. |
361 |
Aborts Auto Zero operation and continues to use previous zero offset value. |
362 |
Exits fixed mode and returns the frequency output to its active state. |
363 |
Invalid sensor type code |
364 |
Special units mass conversion number out of range |
365 |
Frequency out of range |
366 |
Flow rate factor out of range |
367 |
Pulses/Unit out of range |
368 |
Units/Pulse out of range |
369 |
Special units volume conversion number out of range |
370 |
Density high limit out of range |
371 |
Density low limit out of range |
372 |
Slug duration out of range |
373 |
Density Temp. Gradient out-of-range |
374 |
Density Frequency Constant 1 out-of-range |
375 |
Density Frequency Constant 2 out-of-range |
376 |
Flow Temperature Constant out-of-range |
377 |
Flow Frequency Constant out-of-range |
378 |
Maximum zeroing time out of range |
379 |
Maximum pulse width out of range |
380 |
Analog fault level out of range |
381 |
Fault timeout out of range |
382 |
Invalid Output Code |
383 |
Density for calibration out of range |
384 |
FD correction factor out of range |
385 |
D3 out of range |
386 |
D4 out of range |
387 |
K3 out of range |
388 |
K4 out of range |
389 |
Low flow cutoff out of range |
390 |
Invalid Cutoff Type Code |
391 |
D1 out of range |
392 |
D2 out of range |
393 |
K1 out of range |
394 |
K2 out of range |
395 |
Temp correction factor out of range |
396 |
Setpoint out of range |
397 |
Mass rate flow factor out of range |
398 |
Density factor out of range |
399 |
Volume rate flow factor out of range |
400 |
FMU860 |
401 |
DVC |
402 |
FCX |
403 |
AS800 |
404 |
340T |
405 |
340C |
406 |
344 |
407 |
Siemens Micro K |
408 |
SITRANS L |
409 |
SIPAN PH |
410 |
SITRANS P |
411 |
75SMT |
412 |
UNIDELTA |
413 |
8320 |
414 |
SX5100 |
415 |
TSV175 |
416 |
DMU |
417 |
3780 |
418 |
BM70 |
419 |
8u |
420 |
low |
421 |
high |
422 |
tmp_av |
423 |
Sensor type: |
424 |
T-Series;Other |
425 |
Aborting due to field device error |
426 |
WARN-Loop should be removed from automatic control |
427 |
Flow must be zero
Perform calibration? |
428 |
A problem was detected |
429 |
Abort;Fix and Retry |
430 |
NOTE-Loop may be returned to automatic control |
431 |
Calibration in Progress
Zero: %{0}
Std dev: %{1} |
432 |
Auto Zero Failed |
433 |
Auto Zero Complete
Zero: %{0}
Std dev: %{1} |
434 |
Enter Density |
435 |
Perform calibration? |
436 |
Calibration in Progress... |
437 |
Density Calibration Complete |
438 |
NOTE: Flow rate must be greater than 50% of the nominal flow for the meter. |
439 |
Offset Calibration
Enter Temp in %{0} |
440 |
Temperature Offset Calibration Complete |
441 |
Slope Calibration
Enter Temp in %{0} |
442 |
Temperature Slope Calibration Complete |
443 |
Choose frequency output level |
444 |
10 kHz;Other;End |
445 |
Output |
446 |
Frequency output is fixed at %0 Hz |
447 |
Choose analog output level |
448 |
4 mA;20 mA;Other;End |
449 |
Output value: |
450 |
Analog Output 1 is fixed at %0 |
451 |
Connect reference meter |
452 |
About to set analog output 1 to %0 mA |
453 |
Enter meter value |
454 |
Is fld dev 4.00 mA output equal to reference meter? |
455 |
Yes;No |
456 |
Is fld dev 20.00 mA output equal to reference meter? |
457 |
Selftest complete |
458 |
No Device Problems seen during Selftest |
459 |
Ev1 Triggered: Yes |
460 |
Ev1 Triggered: No |
461 |
Ev2 Triggered: Yes |
462 |
Ev2 Triggered: No |
463 |
Transmitter status complete |
464 |
Transmitter status OK |
465 |
Returning fld dev to original output |
466 |
NOTE-Send PV assignment before modifying range values. |
467 |
NOTE-Send SV assignment before modifying range values. |
468 |
NOTE-Send TV assignment before modifying frequency scaling. |
469 |
NOTE-Variables which use this as the units code will be in the previous units until this value is sent to the device |
470 |
NOTE-Send scaling method selection to the device prior to performing frequency output scaling. |