0 |
密度气体粘度计 |
1 |
型号 |
2 |
动作 |
3 |
固定 |
4 |
实时 |
5 |
模式 |
6 |
写保护 |
7 |
Disabled |
8 |
启用 |
9 |
触发模式 |
10 |
The type of process change that will initiate the trigger. |
11 |
事件通知控制: |
12 |
初级变量超出限定范围 |
13 |
非初级变量超出限定范围 |
14 |
初级变量毫安输出饱和 |
15 |
初级变量毫安输出固定 |
16 |
g/mol |
17 |
µ秒 |
18 |
MJ/Kg |
19 |
MJ/Nm3 |
20 |
BTU/lb |
21 |
BTU/scf |
22 |
MJ/min |
23 |
MJ/day |
24 |
BTU/min |
25 |
BTU/day |
26 |
长位号 |
27 |
日期 |
28 |
说明 |
29 |
变送器软件 |
30 |
.1f |
31 |
硬件 |
32 |
序列号 |
33 |
通用 |
34 |
现场设备 |
35 |
A000:未定义 |
36 |
A001:EEPROM 错误 |
37 |
A002:RAM 错误 |
38 |
A003:传感器无响应 |
39 |
A004:温度超限 |
40 |
A005: 速度超限 |
41 |
A006:需要特征化 |
42 |
A007: 预留 |
43 |
A008:密度超限 |
44 |
A009:变送器初始化/预热 |
45 |
A010:校准失败 |
46 |
A011:调零校准失败:低 |
47 |
A012:调零校准失败:高 |
48 |
A013:调零校准失败:不稳定 |
49 |
A014:变送器故障 |
50 |
A015: 预留 |
51 |
A016: 传感器 RTD 故障 |
52 |
A017: 传感器 RTD 故障 |
53 |
A018: 预留 |
54 |
A019: 预留 |
55 |
A020: 缺少标定系数 |
56 |
A021: 变送器/传感器/软件不匹配 |
57 |
A022: 预留 |
58 |
A023: 预留 |
59 |
A024: 预留 |
60 |
A025: 预留 |
61 |
A026: 预留 |
62 |
A027:安全漏洞 |
63 |
A028: 预留 |
64 |
A029: PIC/子板通讯故障 |
65 |
A030:错误的电子板类型 |
66 |
A031:电量不足 |
67 |
A032: 预留 |
68 |
A033: 传感器信号不足 |
69 |
A034: 预留 |
70 |
A035: 预留 |
71 |
A036: Viscosity Overrange |
72 |
A037: 传感器检查失败 |
73 |
A038: 周期超限 |
74 |
A039:未定义 |
75 |
A100: 毫安输出 1 饱和 |
76 |
A101: 毫安输出 1 固定 |
77 |
A102:驱动超限 |
78 |
A103: 预留 |
79 |
A104: 标定进行中 |
80 |
A105: Two-Phase Flow |
81 |
A106:启用阵发模式 |
82 |
A107:发生电源复位 |
83 |
A108: 预留 |
84 |
A109: 预留 |
85 |
A110: 预留 |
86 |
A111: 预留 |
87 |
A112: 预留 |
88 |
A113: 毫安输出 2 饱和 |
89 |
A114:毫安输出 2 固定 |
90 |
A115: 无外部输入或轮询数据 |
91 |
A116: 温度超限 (API) |
92 |
A117: 密度超限 (API) |
93 |
A118: 离散输出固定 |
94 |
A119: 预留 |
95 |
A120: 曲线拟合失败(浓度) |
96 |
A121: 外推警报(浓度) |
97 |
A122: 压力超限 (API) |
98 |
A123: 预留 |
99 |
A124: 预留 |
100 |
A125: 预留 |
101 |
A126: 预留 |
102 |
A127: 预留 |
103 |
A128: 预留 |
104 |
A129: 预留 |
105 |
A130: 预留 |
106 |
A131: 预留 |
107 |
A132:传感器仿真激活 |
108 |
A133: EEPROM 错误(显示屏) |
109 |
A134: 管-壳温差超限 |
110 |
A135: 周期中断激活 |
111 |
A136: 显示类型不正确 |
112 |
A137: 传感器检查可疑 |
113 |
A138: 预留 |
114 |
A139:未定义 |
115 |
传感器检查进行中 |
116 |
SG/RD/MW 标定进行中 |
117 |
密度偏差标定进行中 |
118 |
粘度比例标定进行中 |
119 |
标定检查 |
120 |
调零校准正在进行 |
121 |
温度斜率校准正在进行 |
122 |
温度偏移校准正在进行 |
123 |
增强事件 1 激活 |
124 |
增强事件 2 激活 |
125 |
增强事件 3 激活 |
126 |
增强事件 4 激活 |
127 |
增强事件 5 激活 |
128 |
离散输出 1 激活 |
129 |
预留 |
130 |
Velocity Switch Active |
131 |
未定义 |
132 |
Dynamic Variables Returned for Device Variables |
133 |
Command Response Truncated |
134 |
|
135 |
|
136 |
|
137 |
|
138 |
% 范围 |
139 |
MAO1 |
140 |
初级变量百分比范围 |
141 |
初级变量量程上限值 |
142 |
初级变量量程下限值 |
143 |
4.2f |
144 |
更新速率 |
145 |
默认更新速率 |
146 |
触发水平 |
147 |
The value of the process variable that will initiate the trigger. Applicable only if Trigger Mode = Rising or Falling. |
148 |
阵发变量 0 |
149 |
触发时间间隔 |
150 |
The amount of time that the meter will wait, after an alert is active, before initiating event notification. |
151 |
重试速率 |
152 |
Event notification will continue until acknowledged by a host or until this interval expires, whichever comes first. |
153 |
A105:团状流 |
154 |
选项 |
155 |
带状态的设备变量 |
156 |
读取其他设备状态 |
157 |
选择无效。 |
158 |
速度开关设定点值超出范围 |
159 |
变送器的特殊指令错误(无效 CM 曲线) |
160 |
无效浓度单位代码 |
161 |
无效温度单位代码 |
162 |
无效密度单位代码 |
163 |
参考温度超出范围 |
164 |
水参考温度超出范围 |
165 |
水参考密度超出范围 |
166 |
斜率调整超出范围 |
167 |
斜率偏移超出范围 |
168 |
推测警报限定超出范围 |
169 |
拟合最高阶超出范围 |
170 |
线圈或启动寄存器超出范围。 |
171 |
值超出范围。 |
172 |
温度斜率标定超出范围 |
173 |
温度偏差标定超出范围 |
174 |
模拟故障水平超出限制。 |
175 |
热膨胀系数超出范围 |
176 |
备选压力超出范围。 |
177 |
启用 API 计算时出错。 |
178 |
启用超出范围的浓度测量代码 |
179 |
启用密度外推下限时出错。 |
180 |
启用密度外推上限时出错。 |
181 |
启用温度外推下限时出错。 |
182 |
启用温度外推上限时出错。 |
183 |
曲线索引无效 |
184 |
数据索引无效 |
185 |
Invalid Viscosity Range Select Index |
186 |
Invalid Referred Viscosity Calculation Method |
187 |
Invalid Referred Viscosity Maximum Fit Order |
188 |
遗留指数 1 无效。 |
189 |
遗留指数 2 无效。 |
190 |
无效浓度测量模式代码 |
191 |
密度专用基本单位超出范围。 |
192 |
动态粘度专用基本单位超出范围。 |
193 |
运动粘度专用基本单位超出范围。 |
194 |
Invalid Compressibility Method |
195 |
SGERG value Out of Range |
196 |
用户水密度超出范围。 |
197 |
用户定义的计算输出基本密度来源无效。 |
198 |
高度无效。 |
199 |
Invalid Gas Type |
200 |
浓度方程类型无效。 |
201 |
应用过程太高。 |
202 |
应用过程太低。 |
203 |
Enable External Volume Flow Error |
204 |
Enable External Mass Flow Error |
205 |
Density at Reference (API) |
206 |
浓度 (CM) |
207 |
Density at Reference (CM) |
208 |
管-壳温差 |
209 |
速度 |
210 |
用户定义的计算输出 |
211 |
分子量 |
212 |
相对密度 |
213 |
Specific Gravity (Gas) |
214 |
Base Density (Gas) |
215 |
Net Mass Flow Rate (CM) |
216 |
Net Volume Flow Rate (CM) |
217 |
Specific Gravity (CM) |
218 |
动态粘度 |
219 |
运动粘度 |
220 |
参考粘度 |
221 |
Referred Viscosity (Secondary) |
222 |
可压缩性 |
223 |
电流(毫安输出 1) |
224 |
电流(毫安输出 2) |
225 |
Pickoff Amplitude |
226 |
传感器周期 (A) |
227 |
传感器周期 |
228 |
质量系数 |
229 |
外壳温度 |
230 |
特定温度和浓度液体密度 |
231 |
浓度 |
232 |
浓度 1 |
233 |
密度 1 |
234 |
浓度 2 |
235 |
密度 2 |
236 |
浓度 3 |
237 |
密度 3 |
238 |
浓度 4 |
239 |
密度 4 |
240 |
浓度 5 |
241 |
密度 5 |
242 |
参考温度和特定浓度液体密度 |
243 |
参考温度 |
244 |
参考密度 1 |
245 |
参考密度 2 |
246 |
参考密度 3 |
247 |
参考密度 4 |
248 |
参考密度 5 |
249 |
浓度 6 |
250 |
参考密度 6 |
251 |
Subtest |
252 |
检定周期差异 |
253 |
检定周期稳定性(测试) |
254 |
流体温度稳定性(测试) |
255 |
外壳温度稳定性(测试) |
256 |
驱动增益(测试) |
257 |
结果 |
258 |
Fluids and Samples |
259 |
Fluid 1 and Sample 1 |
260 |
Fluid 1 and Sample 2 |
261 |
Fluid 2 and Sample 1 |
262 |
Fluid 2 and Sample 2 |
263 |
Fluid 3 and Sample 1 |
264 |
Fluid 3 and Sample 2 |
265 |
Fluid 4 and Sample 1 |
266 |
Fluid 4 and Sample 2 |
267 |
Fluid 5 and Sample 1 |
268 |
Fluid 5 and Sample 2 |
269 |
Fluid 6 and Sample 1 |
270 |
Fluid 6 and Sample 2 |
271 |
Fluid 7 and Sample 1 |
272 |
Fluid 7 and Sample 2 |
273 |
Fluid 8 and Sample 1 |
274 |
Fluid 8 and Sample 2 |
275 |
粘度 |
276 |
Fluid 1 |
277 |
Sample 1 |
278 |
Sample 2 |
279 |
Fluid 2 |
280 |
Fluid 3 |
281 |
Fluid 4 |
282 |
Fluid 5 |
283 |
Fluid 6 |
284 |
Fluid 7 |
285 |
Fluid 8 |
286 |
Referred Viscosity at Specified Temperature |
287 |
Limits |
288 |
Calibration Gas Low |
289 |
Calibration Gas Medium |
290 |
Calibration Gas High |
291 |
比重 |
292 |
Sensor Time Period (µs) |
293 |
Molecular Weight (g/mol) |
294 |
高级 HART 组态 |
295 |
重置矩阵数据 |
296 |
Deletes all data in the concentration matrices. |
297 |
External Device Polling |
298 |
锁定/解锁设备 |
299 |
API 参考设置 |
300 |
输入矩阵数据 |
301 |
Define or modify the matrix being configured. |
302 |
毫安输出 |
303 |
恢复工厂组态 |
304 |
定位设备 |
305 |
读取 Modbus 数据 |
306 |
写入 Modbus 数据 |
307 |
Discrete Output Loop Test |
308 |
外部输入 |
309 |
Configure parameters required for pressure, external temperature, and other external variables. |
310 |
特殊单位 |
311 |
Configure the parameters required to convert an existing measurement unit to a different unit. |
312 |
启用/禁用应用程序 |
313 |
Enable or disable optional applications. |
314 |
标定系数 |
315 |
已知密度检定 |
316 |
用户定义的计算 |
317 |
Create a new process variable from current process data and user-defined constants. |
318 |
密度偏差标定 |
319 |
设备复位 |
320 |
A005: Velocity Overrange |
321 |
Start Calibration |
322 |
Viscosity Scaling Factor Calibration |
323 |
Gas Meter Calibration |
324 |
Perform gas meter calibration. |
325 |
Base Density |
326 |
Configure the parameters and inputs required for base density measurement. |
327 |
Enable/Disable Compressibility Calculations |
328 |
|
329 |
Results |
330 |
Select Device Measurement |
331 |
概述 |
332 |
阵发模式 |
333 |
阵发信息 1 |
334 |
信息内容 |
335 |
组态更新速率 |
336 |
阵发信息 2 |
337 |
阵发信息 3 |
338 |
事件通知 |
339 |
警报位选择 |
340 |
标识 |
341 |
修订版本 |
342 |
应用 |
343 |
MID 信息 |
344 |
安全 |
345 |
组态 |
346 |
CDM 100 格式 |
347 |
7835/45 Format |
348 |
Displays the calibration factors in legacy format. |
349 |
Density (Uncompensated) Range 1 |
350 |
Density (Uncompensated) Range 2 |
351 |
Temperature Compensation |
352 |
Pressure Compensation Range 1 |
353 |
Pressure Compensation Range 2 |
354 |
Pressure Compensation Range 3 |
355 |
Pressure Compensation Range 4 |
356 |
Density (Calculated) |
357 |
Density (Uncompensated) |
358 |
Ultra Low Range (0.50 – 10.46 cP) |
359 |
Low Range (10.46 – 108.16 cP) |
360 |
Medium Range (108.16 – 1087.43 cP) |
361 |
High Range (1087.43 – 13345.64 cP) |
362 |
Viscosity Compensation for Density (Low) |
363 |
Viscosity Compensation for Density (High) |
364 |
Velocity of Sound Compensation |
365 |
Calibration Coefficients Set 1 |
366 |
Calibration Coefficients Set 2 |
367 |
Calibration Coefficients Set 3 |
368 |
Calibration Coefficients Set 4 |
369 |
Calibration Information |
370 |
Time Period Data Set 1 |
371 |
Time Period Data Set 2 |
372 |
Time Period Data Set 3 |
373 |
Time Period Data Set 4 |
374 |
Gas Data Set 1 |
375 |
Gas Data Set 2 |
376 |
Gas Data Set 3 |
377 |
Gas Data Set 4 |
378 |
可选设置 |
379 |
Mass (Calculated) |
380 |
API 参考 |
381 |
浓度测量配置 |
382 |
Configure Matrix |
383 |
Reference Conditions |
384 |
特定温度和浓度下 |
385 |
参考温度和浓度下 |
386 |
温度等温线 |
387 |
等温线 1 |
388 |
等温线 2 |
389 |
等温线 3 |
390 |
等温线 4 |
391 |
等温线 5 |
392 |
等温线 6 |
393 |
调整浓度测量过程变量 |
394 |
参考密度 |
395 |
Calculation Constants |
396 |
Configure External Inputs |
397 |
Gas Measurement |
398 |
Setup Compressibility |
399 |
Energy Content Measurement |
400 |
Volume |
401 |
Gas Composition |
402 |
Energy |
403 |
Gas Calibration |
404 |
Mass |
405 |
显示行为 |
406 |
显示菜单 |
407 |
过程变量 |
408 |
速度开关 |
409 |
外部装置轮询 |
410 |
外部轮询 1 |
411 |
外部轮询 2 |
412 |
外部轮询 3 |
413 |
外部轮询 4 |
414 |
Mapped Variable Units |
415 |
输入/输出故障动作 |
416 |
警报严重级别 |
417 |
增强事件 |
418 |
增强事件 1 |
419 |
增强事件 2 |
420 |
增强事件 3 |
421 |
增强事件 4 |
422 |
增强事件 5 |
423 |
浓度测量警报 |
424 |
维修工具 |
425 |
变量 |
426 |
Calculated Variables |
427 |
API Variable |
428 |
Ignition Index |
429 |
毫安输出 1 (MAO1) |
430 |
毫安输出 2 (MAO2) |
431 |
标定 |
432 |
重置/恢复 |
433 |
F:EEPROM 错误 |
434 |
F:RAM 错误 |
435 |
F:传感器无响应 |
436 |
查看诊断变量 |
437 |
F:温度超限 |
438 |
Alert Diagnostic Variables |
439 |
F: Velocity Overrange |
440 |
F:需要特征化 |
441 |
查看标定系数 |
442 |
F:密度超限 |
443 |
F:变送器初始化/预热 |
444 |
F:校准失败 |
445 |
F:调零校准失败:低 |
446 |
F:调零校准失败:高 |
447 |
F:调零校准失败:不稳定 |
448 |
F:变送器故障 |
449 |
F:传感器 RTD 故障 |
450 |
M:传感器 RTD 故障 |
451 |
F: Calibration Factors Missing |
452 |
F: Transmitter/Sensor/Software Mismatch |
453 |
F:安全漏洞 |
454 |
F:PIC/子板通讯故障 |
455 |
F:错误的电子板类型 |
456 |
F: Insufficient Pickoff Signal |
457 |
F: Viscosity Overrange |
458 |
F: Sensor Check Failed |
459 |
F: Time Period Overrange |
460 |
A: mA Output 1 Saturated |
461 |
A: mA Output 1 Fixed |
462 |
M:驱动超限 |
463 |
A:正在进行校准 |
464 |
M: Two-Phase Flow |
465 |
A:启用阵发模式 |
466 |
M:发生电源复位 |
467 |
A:毫安输出 2 饱和 |
468 |
A:毫安输出 2 固定 |
469 |
M:无外部输入或轮询数据 |
470 |
M: Temperature Overrange (API) |
471 |
M: Density Overrange (API) |
472 |
A:离散输出固定 |
473 |
A: Curve Fit Failure (Concentration) |
474 |
M: Extrapolation Alert (Concentration) |
475 |
View Concentration Measurement Configuration |
476 |
M: Pressure Overrange (API) |
477 |
A:传感器仿真激活 |
478 |
M: EEPROM Error (Display) |
479 |
M: Tube-Case Temperature Difference Overrange |
480 |
M: Incorrect Display Type |
481 |
A: Enhanced Event 1 Active |
482 |
A: Enhanced Event 2 Active |
483 |
A: Enhanced Event 3 Active |
484 |
A: Enhanced Event 4 Active |
485 |
A: Enhanced Event 5 Active |
486 |
A: Velocity Switch Active |
487 |
M:初级变量毫安输出饱和 |
488 |
A:初级变量毫安输出固定 |
489 |
User-Defined 1 |
490 |
Programmable Constants |
491 |
Calculation Variables |
492 |
Calculation Variable (a) |
493 |
Calculation Variable (b) |
494 |
Calculation Variable (c) |
495 |
Calculation Variable (d) |
496 |
Calculation Variable (e) |
497 |
Calculation Variable (f) |
498 |
User-Defined 2 |
499 |
User-Defined 3 |
500 |
诊断变量 |
501 |
External Variables |
502 |
Setup Calibration Gas Data |
503 |
Matrix Referral |
504 |
Viscosity at Specific Temp |
505 |
ASTM D341 Single |
506 |
Multi-curve ASTM |
507 |
8.2f |
508 |
8d |
509 |
更新周期 |
510 |
温度分类 |
511 |
温度类型 |
512 |
状态 |
513 |
温度 LS |
514 |
温度类型状态 |
515 |
温度更新周期 |
516 |
.2f |
517 |
.3f |
518 |
.4f |
519 |
.5f |
520 |
.6f |
521 |
High Density Cutoff |
522 |
Damping |
523 |
密度分类 |
524 |
密度类型 |
525 |
密度 LS |
526 |
密度类型状态 |
527 |
密度更新周期 |
528 |
The referred density value after the density offset has been applied. |
529 |
API 密度分类 |
530 |
API 密度类型 |
531 |
API 密度 LS |
532 |
API 密度类型状态 |
533 |
API 密度更新周期 |
534 |
参考条件下的密度 (CM) |
535 |
ED 密度参考分类 |
536 |
ED 密度参考类型 |
537 |
ED 密度参考 LS |
538 |
ED 密度参考类型状态 |
539 |
ED 密度参考更新周期 |
540 |
比重 (CM) |
541 |
Specific Gravity (CM) Upper Sensor Limit |
542 |
Specific Gravity (CM) Lower Sensor Limit |
543 |
Specific Gravity (CM) Minimum Span |
544 |
ED 密度 SGU 分类 |
545 |
ED 密度 SGU 类型 |
546 |
ED 密度 SGU PDQ |
547 |
ED 密度 SGU LS |
548 |
ED 密度 SGU 类型状态 |
549 |
ED 密度 SGU 更新周期 |
550 |
Mass Flow Rate Unit |
551 |
ED 净质量流量分类 |
552 |
ED 净质量流量类型 |
553 |
ED 净质量流量 PDQ |
554 |
ED 净质量流量 LS |
555 |
ED 净质量流量类型状态 |
556 |
ED 净质量流量更新周期 |
557 |
Volume Flow Rate Unit |
558 |
10.4f |
559 |
ED 净体积流量分类 |
560 |
ED 净体积流量类型 |
561 |
ED 净体积流量 PDQ |
562 |
ED 净体积流量 LS |
563 |
ED 净体积流量类型状态 |
564 |
ED 净体积流量更新周期 |
565 |
The concentration value after the density offset has been applied. |
566 |
5.4f |
567 |
ED 浓度分类 |
568 |
ED 浓度类型 |
569 |
ED 浓度 LS |
570 |
ED 浓度类型状态 |
571 |
ED 浓度更新周期 |
572 |
Drive Gain Unit |
573 |
Drive Gain is the amount of current required to drive the vibrating element, expressed as a percentage of the maximum allowable drive gain. |
574 |
6.2f |
575 |
驱动分类 |
576 |
驱动类型 |
577 |
驱动 LS |
578 |
驱动类型状态 |
579 |
驱动更新周期 |
580 |
管壳温度分类 |
581 |
管壳温度类型 |
582 |
管壳温度 PDQ |
583 |
管壳温度 LS |
584 |
管壳温度类型状态 |
585 |
外壳温度更新周期 |
586 |
Pickoff Amplitude represents the distance that the vibrating element is traveling from its resting state. |
587 |
8.3f |
588 |
管分类 |
589 |
管类型 |
590 |
管 PDQ |
591 |
管 LS |
592 |
管类型状态 |
593 |
管更新周期 |
594 |
Board Temperature is the temperature of the device's electronic board stack. |
595 |
电子板温度分类 |
596 |
电子板温度类型 |
597 |
电子板温度 PDQ |
598 |
电子板温度 LS |
599 |
电子板温度类型状态 |
600 |
板温度更新周期 |
601 |
Input Voltage is the current power input. |
602 |
输入电压分类 |
603 |
输入电压类型 |
604 |
输入电压 PDQ |
605 |
输入电压 LS |
606 |
输入电压类型状态 |
607 |
输入电压更新周期 |
608 |
Pressure (Absolute) Unit |
609 |
Pressure (Gauge) Unit |
610 |
External Pressure (Absolute) |
611 |
External Pressure (Gauge) |
612 |
Current pressure value from an external device. |
613 |
压力分类 |
614 |
压力类型 |
615 |
压力 PDQ |
616 |
压力 LS |
617 |
压力类型状态 |
618 |
压力更新周期 |
619 |
Current temperature value from an external device. |
620 |
外部温度分类 |
621 |
外部温度类型 |
622 |
外部温度 PDQ |
623 |
外部温度 LS |
624 |
外部温度类型状态 |
625 |
外部温度更新周期 |
626 |
User-Defined Calculation Output Unit |
627 |
The result of the user-defined calculation. |
628 |
用户定义的计算输出传感器上限 |
629 |
用户定义的计算输出传感器下限 |
630 |
用户定义的计算输出最小量程 |
631 |
用户定义的计算输出分类 |
632 |
用户定义的计算输出类型 |
633 |
用户定义的计算输出 PDQ |
634 |
用户定义的计算输出 LS |
635 |
用户定义的计算输出类型状态 |
636 |
用户定义的计算输出更新周期 |
637 |
Sensor Time Period Unit |
638 |
传感器周期 (A) 分类 |
639 |
传感器周期 (A) 类型 |
640 |
传感器周期 (A) PDQ |
641 |
传感器周期 (A) LS |
642 |
传感器周期 (A) 类型状态 |
643 |
传感器周期更新周期 |
644 |
The difference between the temperature of the sensor case and the temperature of the sensor tubes. |
645 |
管-壳温差分类 |
646 |
管-壳温差类型 |
647 |
管-壳温差 LS |
648 |
管-壳温差类型状态 |
649 |
管-壳温差更新周期 |
650 |
动态粘度单位 |
651 |
动态粘度传感器上限 |
652 |
动态粘度传感器下限 |
653 |
动态粘度最小量程 |
654 |
动态粘度分类 |
655 |
动态粘度类型 |
656 |
动态粘度 LS |
657 |
动态粘度类型状态 |
658 |
动态粘度更新周期 |
659 |
运动粘度单位 |
660 |
运动粘度传感器上限 |
661 |
运动粘度传感器下限 |
662 |
运动粘度最小量程 |
663 |
运动粘度分类 |
664 |
运动粘度类型 |
665 |
运动粘度 LS |
666 |
运动粘度类型状态 |
667 |
运动粘度更新周期 |
668 |
Referred Viscosity Unit |
669 |
参考粘度传感器上限 |
670 |
参考粘度传感器下限 |
671 |
参考粘度最小量程 |
672 |
参考粘度分类 |
673 |
参考粘度类型 |
674 |
参考粘度 PDQ |
675 |
参考粘度 LS |
676 |
参考粘度类型状态 |
677 |
参考粘度更新周期 |
678 |
二级参考粘度传感器上限 |
679 |
二级参考粘度传感器下限 |
680 |
二级参考粘度最小量程 |
681 |
二级参考粘度分类 |
682 |
二级参考粘度类型 |
683 |
二级参考粘度 PDQ |
684 |
二级参考粘度 LS |
685 |
二级参考粘度类型状态 |
686 |
二级参考粘度更新周期 |
687 |
Calculated Carbon Aromaticity Index |
688 |
CCAI 传感器上限 |
689 |
CCAI 传感器下限 |
690 |
CCAI 最小量程 |
691 |
CCAI 分类 |
692 |
CCAI 类型 |
693 |
CCAI PDQ |
694 |
CCAI LS |
695 |
CCAI 类型状态 |
696 |
CCAI 更新周期 |
697 |
Calculated Ignition Index |
698 |
CII 传感器上限 |
699 |
CII 传感器下限 |
700 |
CII 最小量程 |
701 |
CII 分类 |
702 |
CII 类型 |
703 |
CII PDQ |
704 |
CII LS |
705 |
CII 类型状态 |
706 |
CII 更新周期 |
707 |
质量系数传感器上限 |
708 |
质量系数传感器下限 |
709 |
质量系数最小量程 |
710 |
质量系数分类 |
711 |
质量系数类型 |
712 |
质量系数 PDQ |
713 |
质量系数 LS |
714 |
质量系数类型状态 |
715 |
质量系数更新周期 |
716 |
气体比重传感器上限 |
717 |
气体比重传感器下限 |
718 |
气体比重最小量程 |
719 |
气体比重 |
720 |
气体比重分类 |
721 |
气体比重类型 |
722 |
气体比重 LS |
723 |
气体比重类型状态 |
724 |
气体比重更新周期 |
725 |
Relative Density Upper Sensor Limit |
726 |
Relative Density Minimum Span |
727 |
相对密度分类 |
728 |
相对密度类型 |
729 |
相对密度 LS |
730 |
相对密度类型状态 |
731 |
相对密度更新周期 |
732 |
分子量单位 |
733 |
分子量传感器上限 |
734 |
分子量传感器下限 |
735 |
分子量最小量程 |
736 |
分子量分类 |
737 |
分子量类型 |
738 |
分子量 LS |
739 |
分子量类型状态 |
740 |
分子量更新周期 |
741 |
可压缩性分类 |
742 |
可压缩性类型 |
743 |
可压缩性 PDQ |
744 |
可压缩性 LS |
745 |
可压缩性类型状态 |
746 |
可压缩性更新周期 |
747 |
沃布指数 |
748 |
The ratio of the calorific value of a gas to its specific gravity. Typically measured in BTU/SCF or MJ/SCM. |
749 |
沃布指数传感器上限 |
750 |
沃布指数传感器下限 |
751 |
沃布指数最小量程 |
752 |
沃布指数分类 |
753 |
沃布指数类型 |
754 |
沃布指数 PDQ |
755 |
沃布指数 LS |
756 |
沃布指数类型状态 |
757 |
沃布指数更新周期 |
758 |
热值单位 |
759 |
热值 |
760 |
The amount of heat released during the combustion of a specified amount of gas. Measured in units of energy per units of the gas. |
761 |
热值传感器上限 |
762 |
热值传感器下限 |
763 |
热值最小量程 |
764 |
热值分类 |
765 |
热值类型 |
766 |
热值 PDQ |
767 |
热值 LS |
768 |
热值类型状态 |
769 |
热值更新周期 |
770 |
CO2 百分比单位 |
771 |
CO2 百分比 |
772 |
CO2 百分比传感器上限 |
773 |
CO2 百分比传感器下限 |
774 |
CO2 百分比最小量程 |
775 |
CO2 百分比分类 |
776 |
CO2 百分比类型 |
777 |
CO2 百分比 PDQ |
778 |
CO2 百分比 LS |
779 |
CO2 百分比类型状态 |
780 |
CO2 百分比更新周期 |
781 |
External Volume Flow Rate |
782 |
Current volume flow rate from an external device. |
783 |
来自电磁仪表的体积 |
784 |
体积流量分类 |
785 |
体积流量类型 |
786 |
体积流量 LS |
787 |
体积流量类型状态 |
788 |
体积流量更新周期 |
789 |
气体基本密度 |
790 |
气体基本密度分类 |
791 |
气体基本密度类型 |
792 |
气体基本密度 PDQ |
793 |
气体基本密度 LS |
794 |
气体基本密度类型状态 |
795 |
气体基本密度更新周期 |
796 |
Mass Flow Rate (Calculated) |
797 |
Current mass flow rate as calculated from external volume flow rate. |
798 |
质量流量分类 |
799 |
质量流量类型 |
800 |
质量流量 LS |
801 |
质量流量类型状态 |
802 |
质量流量更新周期 |
803 |
N2 百分比单位 |
804 |
N2 百分比 |
805 |
N2 百分比传感器上限 |
806 |
N2 百分比传感器下限 |
807 |
N2 百分比最小量程 |
808 |
N2 百分比分类 |
809 |
N2 百分比类型 |
810 |
N2 百分比 PDQ |
811 |
N2 百分比 LS |
812 |
N2 百分比类型状态 |
813 |
N2 百分比更新周期 |
814 |
H2 百分比单位 |
815 |
H2 百分比 |
816 |
H2 百分比传感器上限 |
817 |
H2 百分比传感器下限 |
818 |
H2 百分比最小量程 |
819 |
H2 百分比分类 |
820 |
H2 百分比类型 |
821 |
H2 百分比 PDQ |
822 |
H2 百分比 LS |
823 |
H2 百分比类型状态 |
824 |
H2 百分比更新周期 |
825 |
CO 百分比单位 |
826 |
CO 百分比 |
827 |
CO 百分比传感器上限 |
828 |
CO 百分比传感器下限 |
829 |
CO 百分比最小量程 |
830 |
CO 百分比分类 |
831 |
CO 百分比类型 |
832 |
CO 百分比 PDQ |
833 |
CO 百分比 LS |
834 |
CO 百分比类型状态 |
835 |
CO 百分比更新周期 |
836 |
参考温度时的体积流量分类 |
837 |
参考温度时的体积流量类型 |
838 |
参考温度时的体积流量 PDQ |
839 |
参考温度时的体积流量 LS |
840 |
参考温度时的体积流量类型状态 |
841 |
参考温度时的体积流量更新周期 |
842 |
External Mass Flow Rate |
843 |
Current mass flow rate from an external device. |
844 |
Energy Flow Rate Unit |
845 |
MJ/hr |
846 |
BTU/hr |
847 |
能量流量 |
848 |
The energy of the process gas flowing through the meter per unit of time. Measured in units of energy per units of time. |
849 |
能量流量传感器上限 |
850 |
能量流量传感器下限 |
851 |
能量流量最小量程 |
852 |
能量流量分类 |
853 |
能量流量类型 |
854 |
能量流量 PDQ |
855 |
能量流量 LS |
856 |
能量流量类型状态 |
857 |
能量流量更新周期 |
858 |
速度单位 |
859 |
The velocity value below which the transmitter will report zero velocity. |
860 |
速度临界值 |
861 |
速度传感器上限 |
862 |
速度阻尼 |
863 |
Viscosity Damping |
864 |
速度传感器下限 |
865 |
速度最小量程 |
866 |
速度分类 |
867 |
速度类型 |
868 |
速度 LS |
869 |
速度类型状态 |
870 |
速度更新周期 |
871 |
The time required for the vibrating element to complete one vibration cycle. |
872 |
Sensor Time Period Upper Sensor Limit |
873 |
Sensor Time Period Lower Sensor Limit |
874 |
Sensor Time Period Minimum Span |
875 |
传感器周期分类 |
876 |
传感器周期类型 |
877 |
传感器周期 PDQ |
878 |
传感器周期 LS |
879 |
传感器周期类型状态 |
880 |
参考条件下的密度 (API) |
881 |
Volume from Mag/Vortex Meter |
882 |
Calculated Mass Flow from Mag Meter Input |
883 |
Volume Flow at Reference Temperature |
884 |
Mass from Coriolis Meter |
885 |
二级变量百分比范围 |
886 |
二级变量量程上限值 |
887 |
二级变量量程下限值 |
888 |
MAO2 |
889 |
仪表法兰 |
890 |
ANSI 300 |
891 |
ANSI 600 |
892 |
PN40 |
893 |
ANSI 900 |
894 |
PN100 |
895 |
ANSI 150 |
896 |
卫生夹接头 |
897 |
PN16 |
898 |
ANSI 1500 |
899 |
锥形座压接 |
900 |
未知 |
901 |
接液材料 |
902 |
Hastelloy C |
903 |
钛 |
904 |
316L SST |
905 |
The Product Code of the device as ordered from the factory. |
906 |
产品代码 |
907 |
1d |
908 |
Controls the effect of forward and reverse flow on transmitter outputs and digital communications. See the transmitter configuration manual for more information. |
909 |
流向 |
910 |
前向 |
911 |
反向 |
912 |
双向 |
913 |
两相流上限 |
914 |
两相流下限 |
915 |
板类型 |
916 |
气体密度计 |
917 |
音叉密度计/粘度计 |
918 |
管式密度计 |
919 |
两相流超时 |
920 |
5.2f |
921 |
Velocity Switch Setpoint |
922 |
滞后 (0.001-0.1) |
923 |
4.3f |
924 |
变量代码 22 |
925 |
变量代码 23 |
926 |
变量代码 24 |
927 |
变量代码 25 |
928 |
变量代码 26 |
929 |
变量代码 27 |
930 |
变量代码 28 |
931 |
变量代码 29 |
932 |
变量代码 30 |
933 |
变量代码 31 |
934 |
变量代码 32 |
935 |
变量代码 33 |
936 |
变量代码 34 |
937 |
变量代码 35 |
938 |
变量代码 36 |
939 |
变量代码 37 |
940 |
变量代码 38 |
941 |
数值所要显示的小数点位数. |
942 |
Sensor Time Period (A) |
943 |
正在配置的矩阵 |
944 |
Select the matrix to be configured or modified. |
945 |
曲线 1 |
946 |
曲线 2 |
947 |
曲线 3 |
948 |
曲线 4 |
949 |
曲线 5 |
950 |
曲线 6 |
951 |
参考温度 |
952 |
水参考温度 |
953 |
温度单位 |
954 |
水参考密度 |
955 |
密度单位 |
956 |
斜率 |
957 |
偏移 |
958 |
限定范围 |
959 |
曲线拟合最高阶 |
960 |
浓度单位 |
961 |
浓度单位串 |
962 |
矩阵名称 |
963 |
曲线拟合结果 |
964 |
准确性 |
965 |
Concentration Equation Type |
966 |
The multiplier (meter factor) to be used to adjust the concentration process variable. |
967 |
The value to be added to the concentration process variable. |
968 |
2 |
969 |
3 |
970 |
4 |
971 |
5 |
972 |
The label to be used for the concentration measurement unit. |
973 |
Label |
974 |
A user-defined label to be used for the concentration measurement unit. |
975 |
Displays the name of the matrix. |
976 |
不良 |
977 |
空 |
978 |
0 |
979 |
1 |
980 |
5d |
981 |
Source |
982 |
速度比率开关 |
983 |
标定进行中 |
984 |
极性 |
985 |
The state to which the discrete output will be set if the transmitter detects a fault with Alert Severity = Fault. If None is selected, the behavior of the discrete output will be controlled by Discrete Output Source. |
986 |
斜率 |
987 |
偏移 |
988 |
3d |
989 |
当前零值 |
990 |
故障动作 |
991 |
高刻度端 (21.0-21.5 mA) |
992 |
低刻度端 (3.2-3.6 mA) |
993 |
故障动作 |
994 |
故障动作 |
995 |
This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 3.2 to 3.6 mA for downscale or 21.0 to 21.5 for upscale. |
996 |
故障等级 |
997 |
故障等级 |
998 |
2d |
999 |
轮询控制 1 |
1000 |
初级轮询主机 |
1001 |
二级轮询主机 |
1002 |
轮询控制 3 |
1003 |
轮询控制 4 |
1004 |
轮询变量 1 |
1005 |
来自科里奥利仪表的质量流量 |
1006 |
轮询变量 2 |
1007 |
轮询变量 3 |
1008 |
轮询变量 4 |
1009 |
外部设备位号 3 |
1010 |
外部设备位号 4 |
1011 |
Pressure (Absolute) Input |
1012 |
Pressure (Gauge) Input |
1013 |
禁用 |
1014 |
启用 |
1015 |
Fixed or Polled Pressure (Absolute) |
1016 |
Fixed or Polled Pressure (Gauge) |
1017 |
RTD 线缆电阻 |
1018 |
Meter RTD Resistance |
1019 |
管线 RTD 电阻 |
1020 |
供电时间 |
1021 |
Power On Time is the number of seconds since the last power cycle. |
1022 |
表 23E |
1023 |
表 24E |
1024 |
/degC |
1025 |
4.1f |
1026 |
已禁用 |
1027 |
已启用 |
1028 |
The pressure to which referred density data will be corrected. |
1029 |
Reference Pressure (Gauge) |
1030 |
API K0 |
1031 |
API K1 |
1032 |
外部温度 |
1033 |
Fixed or Polled Temperature |
1034 |
Enabled: Process measurement will be adjusted for the effect of temperature on the vibrating element.
Disabled: Process measurement will not be adjusted for the effect of temperature. |
1035 |
外部温度模数 |
1036 |
External Temperature for Viscosity |
1037 |
Enabled: The external (polled) temperature value will be used in gas measurement.
Disabled: The onboard (RTD) temperature data will be used in gas measurement. |
1038 |
External Temperature for Gas |
1039 |
菜单密码 |
1040 |
密码(4 位数字) |
1041 |
4d |
1042 |
变量 1 |
1043 |
变量 2 |
1044 |
变量 3 |
1045 |
变量 4 |
1046 |
变量 5 |
1047 |
变量 6 |
1048 |
变量 7 |
1049 |
变量 8 |
1050 |
变量 9 |
1051 |
变量 10 |
1052 |
变量 11 |
1053 |
变量 12 |
1054 |
变量 13 |
1055 |
变量 14 |
1056 |
变量 15 |
1057 |
刷新率 (100 - 10,000) |
1058 |
英语 |
1059 |
德语 |
1060 |
法语 |
1061 |
西班牙语 |
1062 |
Modbus ASCII 支持 |
1063 |
MID |
1064 |
工程师到订单编号 |
1065 |
变送器固件校验和 |
1066 |
The result of the integrity check calculated for the meter. |
1067 |
MID 软件版本 |
1068 |
The version of the software component that implements MID security. |
1069 |
活动矩阵 |
1070 |
浓度测量 |
1071 |
Displays the status of the concentration measurement application. |
1072 |
温度高 |
1073 |
Enable or disable the extrapolation alerts as desired. Note: If you are using matrix switching, you must enable the extrapolation alerts that will trigger matrix switching. |
1074 |
密度低 |
1075 |
密度高 |
1076 |
温度低 |
1077 |
变量 |
1078 |
事件类型 |
1079 |
密度 6 |
1080 |
温度等温线 1 |
1081 |
温度等温线 2 |
1082 |
温度等温线 3 |
1083 |
温度等温线 4 |
1084 |
温度等温线 5 |
1085 |
温度等温线 6 |
1086 |
得出的变量 |
1087 |
Density at Reference |
1088 |
Specific Gravity |
1089 |
Mass Concentration(Dens) |
1090 |
Mass Concentration(SG) |
1091 |
Volume Concentration(Dens) |
1092 |
Volume Concentration(SG) |
1093 |
Concentration(Dens) |
1094 |
Concentration(SG) |
1095 |
速度上限 |
1096 |
速度下限 |
1097 |
The calibration factors calculated for this meter, in an updated format. |
1098 |
A1 |
1099 |
A2 |
1100 |
A3 |
1101 |
A4 |
1102 |
A5 |
1103 |
A6 |
1104 |
A7 |
1105 |
A8 |
1106 |
A9 |
1107 |
The adjustment to be applied to the pressure value. This value is added to the external pressure value. |
1108 |
Pressure (Absolute) Offset |
1109 |
Pressure (Gauge) Offset |
1110 |
The adjustment to be applied to the density process variable. This value is added to the measured density. |
1111 |
密度偏差 |
1112 |
The adjustment to be applied to the density process variable. The measured density is multiplied by this value. |
1113 |
仪表系数 |
1114 |
VOS 校正 |
1115 |
The unit from which the special unit will be calculated. |
1116 |
基础单位 |
1117 |
The value by which the base unit will be divided, to report data in the special unit. |
1118 |
转换系数 |
1119 |
The name to be used for the special unit. |
1120 |
标签 |
1121 |
Enabled: The meter will calculate and report process variables based on the values specified here.
Disabled: The meter will calculate and report process variables based on live data. |
1122 |
When Sensor Simulation is enabled, Velocity will be fixed at this value. |
1123 |
The amount of time required for one vibration cycle. |
1124 |
The temperature of the process gas or fluid. |
1125 |
A |
1126 |
The value to be used for this constant in the equation. |
1127 |
B |
1128 |
X |
1129 |
C |
1130 |
Y |
1131 |
E |
1132 |
F |
1133 |
常量 |
1134 |
The constant value to be used for this variable in the equation. |
1135 |
User-Defined Calculation Output Type |
1136 |
基本密度来源 |
1137 |
计算得出的基本密度 |
1138 |
用户基本密度 |
1139 |
The process variable to be used for this variable in the equation. |
1140 |
t |
1141 |
a |
1142 |
b |
1143 |
c |
1144 |
d |
1145 |
e |
1146 |
f |
1147 |
ρw (density of water) |
1148 |
Density of water at reference temperature and reference pressure. |
1149 |
The label to be used for the result of the user-defined calculation. |
1150 |
Calculation Output Label |
1151 |
实验室密度 |
1152 |
The density of your process fluid at line temperature, as measured in a laboratory or by a very accurate device. |
1153 |
开始密度偏差标定 |
1154 |
用户传感器检查 |
1155 |
基本密度传感器检查 |
1156 |
Time Period (Vacuum) at 20 degC |
1157 |
Time Period (Air) at 20 degC |
1158 |
健康检查结果 |
1159 |
可疑 |
1160 |
未运行 |
1161 |
进行中 |
1162 |
No Factory Calibration |
1163 |
传感器周期 (A) 平均值 |
1164 |
传感器周期 (A) 稳定性 |
1165 |
温度检查平均值 |
1166 |
温度检查稳定性 |
1167 |
传感器检查平均值 |
1168 |
质量平均值 |
1169 |
温差平均值 |
1170 |
温差稳定性 |
1171 |
仪表海拔 |
1172 |
The altitude or elevation of this meter. 0 = sea level. |
1173 |
0 英尺 |
1174 |
1000 英尺 |
1175 |
2000 英尺 |
1176 |
3000 英尺 |
1177 |
4000 英尺 |
1178 |
5000 英尺 |
1179 |
6000 英尺及更高 |
1180 |
500 米 |
1181 |
1000 米 |
1182 |
2000 米及更高 |
1183 |
健康检查失败 |
1184 |
传感器周期平均值错误 |
1185 |
传感器周期最大偏差错误 |
1186 |
温度最大偏差错误 |
1187 |
DG 平均值错误 |
1188 |
外壳温度最大偏差错误 |
1189 |
开始传感器检查 |
1190 |
Measurement Type |
1191 |
音叉类型 |
1192 |
Viscomaster |
1193 |
Viscomaster Dynamic |
1194 |
Aquametro Basic |
1195 |
接口类型 |
1196 |
TPS |
1197 |
模拟 |
1198 |
远程访问 |
1199 |
尖头型 |
1200 |
短 |
1201 |
长 |
1202 |
CM SG |
1203 |
CM Baume |
1204 |
CM Twaddle |
1205 |
CM Plato |
1206 |
CM Brix |
1207 |
Matrix Switching |
1208 |
通道 A |
1209 |
毫安输出 1 |
1210 |
通道 B |
1211 |
毫安输出 2 |
1212 |
密度遗留指数 1 |
1213 |
设置 1 |
1214 |
设置 2 |
1215 |
K0 |
1216 |
K1 |
1217 |
K2 |
1218 |
The calibration factors calculated for this meter, in legacy format. |
1219 |
密度遗留指数 2 |
1220 |
设置 3 |
1221 |
设置 4 |
1222 |
K20A |
1223 |
K20B |
1224 |
K21A |
1225 |
K21B |
1226 |
K18 |
1227 |
K19 |
1228 |
K22 |
1229 |
K23 |
1230 |
驱动增益平均值 |
1231 |
驱动增益稳定性 |
1232 |
外壳温度平均值 |
1233 |
外壳温度稳定性 |
1234 |
选择要进行组态的选项: |
1235 |
完成退出 |
1236 |
Select ‘Yes’ to save the burst configuration settings. |
1237 |
保存组态: |
1238 |
Version of the current Device Description. |
1239 |
A020: 未输入标定系数 |
1240 |
A021: 传感器类型未被认出/未输入 |
1241 |
A029: 内部通讯故障 |
1242 |
A030: 硬件/软件不兼容 |
1243 |
A033: 管道/音叉不振动 |
1244 |
A038: 周期超出范围 |
1245 |
A100:MAO1 饱和 |
1246 |
A114: 毫安输出 2 固定 |
1247 |
A122: API:压力超出标准范围 |
1248 |
A133: UI EEPROM 错误 |
1249 |
A134: 外壳-仪表温差超出限制 |
1250 |
A136: 显示类型不正确 |
1251 |
PASS |
1252 |
FAIL |
1253 |
KDV Calibration Index |
1254 |
传感器周期稳定性 |
1255 |
Sensor Time Period obtained during factory test. |
1256 |
用户平均周期 |
1257 |
A113: Secondary mAO Saturated |
1258 |
A114: Secondary mAO Fixed |
1259 |
A118: Discrete Output Fixed |
1260 |
A036: 速度超出限制 |
1261 |
KV Index |
1262 |
KV4 |
1263 |
KV5 |
1264 |
KV6 |
1265 |
Density calibration factors. |
1266 |
K0 |
1267 |
K1 |
1268 |
K2 |
1269 |
Temperature calibration factors. |
1270 |
K18 |
1271 |
K19 |
1272 |
Velocity of Sound of Calibration Fluid |
1273 |
Velocity of Sound of Measured Fluid |
1274 |
Gas K19 Index |
1275 |
Density calibration factors, used to establish the relationship between density and sensor response. |
1276 |
6.6f |
1277 |
Temperature compensation factors, used to adjust measurement for the effects of temperature on the device. |
1278 |
Velocity of sound compensation factors, used to adjust measurement for the effects of pressure waves in the process gas. |
1279 |
K3 |
1280 |
K4 |
1281 |
The molecular weight of air. The default value (28.96469 g/mol) is appropriate for most applications. |
1282 |
Molecular Weight of Air |
1283 |
Calibration Gas Constant |
1284 |
Measurement Gas Constant |
1285 |
Calibration Temperature Constant |
1286 |
Measurement Temperature Constant |
1287 |
Viscosity Value Range |
1288 |
Ultra Low (0.50 – 10.46 cP) |
1289 |
Low (10.46 – 108.16 cP) |
1290 |
Medium (108.16 – 1087.43 cP) |
1291 |
High (1087.43 – 13345.64 cP) |
1292 |
V0 |
1293 |
V1 |
1294 |
V2 |
1295 |
Range Scale at selected range |
1296 |
Viscosity range switch setpoint index |
1297 |
Ultra Low to Low |
1298 |
Low to Medium |
1299 |
Medium to High |
1300 |
Dynamic Viscosity Scaling Factor |
1301 |
Viscosity Hysteresis |
1302 |
Compression Filter parameter |
1303 |
Viscosity Offset |
1304 |
The density of the process gas at base temperature and base pressure. |
1305 |
Density ASTM Coeff A |
1306 |
Density ASTM Coeff B |
1307 |
Start Viscosity Scaling Factor Calibration |
1308 |
cP |
1309 |
Laboratory Dynamic Viscosity |
1310 |
Referred Viscosity Calculation Method |
1311 |
Referred Viscosity Maximum Fit Order for 6x6 matrix |
1312 |
Referred Temperature 1 |
1313 |
Referred Temperature 2 |
1314 |
Fit Accuracy |
1315 |
Fit Result |
1316 |
Good |
1317 |
Failed |
1318 |
Matrix Data Unit |
1319 |
Number of ASTM Reference Curves |
1320 |
Compressibility Method |
1321 |
The measurement standard to be used to measure compressibility. |
1322 |
SGERG88 |
1323 |
AGA NX19 |
1324 |
AGA NX19 MOD |
1325 |
AGA NX19 MOD3H |
1326 |
Base Pressure (Absolute) |
1327 |
Base Pressure (Gauge) |
1328 |
The compressibility of the process gas at base temperature and base pressure. |
1329 |
Base Compressibility |
1330 |
The temperature to which density data will be corrected (the reference temperature). |
1331 |
Base Temperature |
1332 |
The density of air at base temperature and base pressure. |
1333 |
Base Density of Air |
1334 |
The percentage of carbon dioxide in the process gas. |
1335 |
The percentage of nitrogen in the process gas. |
1336 |
The percentage of hydrogen in the process gas. |
1337 |
The percentage of carbon monooxide in the process gas. |
1338 |
SGERG Method |
1339 |
SGERG_ALL |
1340 |
SGERG_NOT_N2 |
1341 |
SGERG_NOT_CO2 |
1342 |
SGERG_NOT_CV |
1343 |
SGERG_NOT_RD |
1344 |
Calibration Number |
1345 |
Low Range Time Period Gas K0 |
1346 |
Mid Range Time Period Gas K1 |
1347 |
High Range Time Period Gas K2 |
1348 |
Low Range SG Gas K0 |
1349 |
Mid Range SG Gas K1 |
1350 |
High Range SG Gas K2 |
1351 |
Gas |
1352 |
The sensor time period used to calculate the K0 calibration point. |
1353 |
Sensor Time Period (Low) |
1354 |
The sensor time period used to calculate the K1 calibration point. |
1355 |
Sensor Time Period (Medium) |
1356 |
The sensor time period used to calculate the K2 calibration point. |
1357 |
Sensor Time Period (High) |
1358 |
The specific gravity used to calculate the K0 calibration point. |
1359 |
The relative density used to calculate the K0 calibration point. |
1360 |
The molecular weight used to calculate the K0 calibration point. |
1361 |
Specific Gravity (Low) |
1362 |
Relative Density (Low) |
1363 |
Molecular Weight Low (g/mol) |
1364 |
5.6f |
1365 |
The specific gravity used to calculate the K1 calibration point. |
1366 |
The relative density used to calculate the K1 calibration point. |
1367 |
The molecular weight used to calculate the K1 calibration point. |
1368 |
Specific Gravity (Medium) |
1369 |
Relative Density (Medium) |
1370 |
Molecular Weight Medium (g/mol) |
1371 |
The specific gravity used to calculate the K2 calibration point. |
1372 |
The relative density used to calculate the K2 calibration point. |
1373 |
The molecular weight used to calculate the K2 calibration point. |
1374 |
Specific Gravity (High) |
1375 |
Relative Density (High) |
1376 |
Molecular Weight High (g/mol) |
1377 |
The name of the gas to be used to determine the K0 calibration point. |
1378 |
Hydrogen |
1379 |
Helium |
1380 |
CO2 |
1381 |
Nitrogen |
1382 |
CO |
1383 |
Oxygen |
1384 |
Argon |
1385 |
Air |
1386 |
Hydrogen Sulphide |
1387 |
Methane |
1388 |
Ethane |
1389 |
Propane |
1390 |
Butane |
1391 |
Custom Gas 1 |
1392 |
Custom Gas 2 |
1393 |
Custom Gas 3 |
1394 |
Custom Gas 4 |
1395 |
Other |
1396 |
The name of the gas to be used to determine the K1 calibration point. |
1397 |
The name of the gas to be used to determine the K2 calibration point. |
1398 |
Start/Abort Gas Calibration |
1399 |
Abort |
1400 |
Start |
1401 |
Save Gas Calibration 1 |
1402 |
Save Gas Calibration 2 |
1403 |
Save Gas Calibration 3 |
1404 |
Start Gas Calibration |
1405 |
Active Calibration |
1406 |
The calibration factors to be used for measurement. |
1407 |
Calibration Format |
1408 |
Displays whether this calibration is based on two calibration gases or three calibration gases. |
1409 |
2-Point Calibration |
1410 |
3-Point Calibration |
1411 |
Indicates whether or not the data for this calibration is acceptable. |
1412 |
Calibration Point |
1413 |
A measure of the variability of the Sensor Time Period value, used to indicate the stage or the quality of the calibration process. |
1414 |
ns |
1415 |
Stability |
1416 |
Volume Flow Source |
1417 |
Enabled: Volume flow data will be obtained from an external device. Requires polling setup.
Disabled: No volume flow data will be obtained. |
1418 |
The current volume flow rate from the external device. |
1419 |
External Flow |
1420 |
Mass Flow Source |
1421 |
Enabled: Mass flow data will be obtained from an external device. Requires polling setup.
Disabled: No mass flow data will be obtained. |
1422 |
The current mass flow rate from the external device. |
1423 |
Referred Viscosity Curve Isotherm Index |
1424 |
Referred Viscosity Temperature Isotherm Value |
1425 |
Referred Viscosity ISO at temperature 0 |
1426 |
Referred Viscosity ISO at temperature 1 |
1427 |
Referred Viscosity ISO at temperature 2 |
1428 |
Referred Viscosity ISO at temperature 3 |
1429 |
Referred Viscosity ISO at temperature 4 |
1430 |
Referred Viscosity ISO at temperature 5 |
1431 |
Centistokes |
1432 |
Viscosity 1 |
1433 |
Viscosity 2 |
1434 |
Viscosity 3 |
1435 |
Viscosity 4 |
1436 |
Viscosity 5 |
1437 |
Viscosity 6 |
1438 |
Referred Viscosity ASTM Curve Index |
1439 |
Temperature 1 |
1440 |
Temperature 2 |
1441 |
The compressibility of the process gas. Typically measured using a gas chromatograph. |
1442 |
Line Compressibility |
1443 |
Tag, model code, revision, security, and related data. |
1444 |
修订号 |
1445 |
Current values of process variables and outputs. |
1446 |
NOTE: If pressure is assigned, the value will be reported as gauge pressure. |
1447 |
NOTE: If pressure is assigned, the value will be reported as absolute pressure. |
1448 |
初级变量 |
1449 |
二级变量 |
1450 |
External Variable |
1451 |
Trend charts for common process and diagnostic variables. |
1452 |
Displays a trend chart for net mass flow rate (CM). |
1453 |
Displays a trend chart for net volume flow rate (CM). |
1454 |
Displays a trend chart for specific gravity (CM). |
1455 |
Displays a trend chart for density. |
1456 |
Displays a trend chart for temperature. |
1457 |
Displays a trend chart for velocity. |
1458 |
Displays a trend chart for concentration (CM). |
1459 |
Displays a trend chart for density at reference (CM). |
1460 |
Displays a trend chart for density at reference (API). |
1461 |
Displays a trend chart for specific gravity (Gas). |
1462 |
Displays a trend chart for molecular weight. |
1463 |
Displays a trend chart for relative density. |
1464 |
Displays a trend chart for base density (Gas). |
1465 |
Displays a trend chart for dynamic viscosity. |
1466 |
Displays a trend chart for kinematic viscosity. |
1467 |
Displays a trend chart for referred viscosity. |
1468 |
Displays a trend chart for referred viscosity (Secondary). |
1469 |
Displays a trend chart for sensor time period (A). |
1470 |
Displays a trend chart for sensor time period. |
1471 |
Displays a trend chart for case temperature. |
1472 |
Displays a trend chart for tube-case temp difference. |
1473 |
Displays a trend chart for input voltage. |
1474 |
Displays a trend chart for board temperature. |
1475 |
Displays a trend chart for left pickoff amplitude. |
1476 |
Displays a trend chart for right pickoff amplitude. |
1477 |
Displays a trend chart for pickoff amplitude. |
1478 |
Displays a trend chart for drive gain. |
1479 |
Displays a trend chart for quality factor. |
1480 |
初始设置 |
1481 |
组态标定系数。 |
1482 |
Configure channels, outputs and mapped variables. |
1483 |
执行已知密度检定。 |
1484 |
Configure outputs and mapped variables. |
1485 |
Perform Gas Meter Calibration. |
1486 |
View calibration data. |
1487 |
Configure the parameters required to measure the energy content of the gas. |
1488 |
Note: Gas Composition values can be read in from an external device if External Polling has been configured. |
1489 |
Configure the parameters required for external pressure, temperature, gas composition, and other process variables. |
1490 |
The meter supports a maximum of four polling connections. |
1491 |
Configure the parameters required for fundamental gas measurement. |
1492 |
Configure the parameters required to measure gas compressibility. |
1493 |
Calculation Methods |
1494 |
Note: The SGERG-88 and AGA NX19 MOD3H compressibility methods require Calorific Value. You must set up energy content measurement for these methods. |
1495 |
Note: Gas Composition values can be read in from an external device if External Polling has been configured. Percent H2 value can be entered only if Compressibility Method is set to SGERG-88. |
1496 |
选择要显示在显示屏上的过程变量。 |
1497 |
指定每个过程变量的显示精度。 |
1498 |
View the output channel types. |
1499 |
组态毫安输出 1 和毫安输出 2(如果可用)。 |
1500 |
组态轮询以获取外部过程数据。 |
1501 |
将变量映射到“初级变量 (PV)”、“二级变量 (SV)”、“三级变量 (TV)”和“四级变量 (QV)”。 |
1502 |
设置通道 B 的通道类型。 |
1503 |
组态离散输出。 |
1504 |
View the channel types. |
1505 |
仪表信息 |
1506 |
组态阵发信息和事件通知。 |
1507 |
组态 HART 通讯。 |
1508 |
组态阵发变量 |
1509 |
Configure the contents of Burst Message 1. |
1510 |
设备变量 |
1511 |
PV、SV、TV、QV |
1512 |
装置锁定时,只有当前连接的 HART 主机可组态或标定装置。 装置未锁定时,所有 HART 主机均可组态或标定装置。 其他连接类型不受影响。 |
1513 |
数字通讯 |
1514 |
输入警报代码并将警报的严重级别更改为 Fault(故障)、Information(通知)或 Ignore(忽略)。 |
1515 |
输入警报代码并查看警报的当前严重级别。 |
1516 |
外推类型 |
1517 |
同时确认所有警报。 |
1518 |
变送器检测到与传感器的通讯问题。 |
1519 |
推荐措施:
1. 关闭并重新打开仪表电源。
2. 联系高准公司。 |
1520 |
变送器未从传感器接收一个或多个基本电信号。 |
1521 |
推荐措施:
1. 检查驱动增益和传感器电压。
2. 检查有无短路。
3.检查传感器管的完整性。 |
1522 |
RTD 电阻超出传感器的范围。 |
1523 |
推荐措施:
1. 检查有无短路。
2. 验证温度特征或标定参数。
3. 对照装置报告的值检查过程条件。
4. 联系高准公司。 |
1524 |
测得的速度超过传感器的最大速度限制。 |
1525 |
推荐措施:
1. 如果存在其他警报,首先解决这些警报条件。 如果当前警报持续存在,请继续执行推荐措施。
2. 对照装置报告的值检查过程条件。
3. 检查两相流。 |
1526 |
未输入标定系数,或传感器类型错误,或标定系数不适用于该传感器类型。 |
1527 |
推荐措施:
1. 验证所有特征或标定参数。 参考传感器位号或仪表的标定数据表。
2. 检查有无短路。
3. 联系高准公司。 |
1528 |
管道密度超过 10 g/cm3。 |
1529 |
推荐措施:
1. 如果存在其他警报,首先解决这些警报条件。 如果当前警报持续存在,请继续执行推荐措施。
2. 检查流量管中是否存在有空气、未充液、有异物、有涂层或其他过程问题。
3. 检查两相流。
4. 如果伴有警报 A003,检查有无短路。
5. 验证所有特征或标定参数。 参考传感器位号或仪表的标定数据表。
6. 检查驱动增益和传感器电压。
7. 联系高准公司。 |
1530 |
变送器处于通电模式。 |
1531 |
推荐措施:
1. 等待仪表完成其通电过程。 警报应当自动清除。
2. 验证并确保管中充满过程流体。
3. 验证并确保变送器获得充足的电能。
4. 如果未获得,纠正问题,关闭并重新打开仪表电源。
5. 如果已获得,则表示变送器存在内部电源问题。 更换变送器。 |
1532 |
可能的原因有多种,例如标定过程中流经传感器的流体过多,或者调零结果超出范围。 此警报直到关闭并重新打开仪表电源后才会清除。 |
1533 |
推荐措施:
1. 如果此警报在调零过程中出现(仅限于 CDM 仪表),验证并确保没有流体流过传感器,关闭并重新打开仪表电源,然后重新执行该程序。 |
1534 |
可能的原因有多种,例如标定过程中流经传感器的流体过多(特别是反向流),或者调零结果过低。 此警报伴有 A010。 此警报直到关闭并重新打开仪表电源后才会清除。 |
1535 |
推荐措施:
1. 验证没有流体流过传感器,关闭并重新打开仪表电源,然后重新执行该程序。 |
1536 |
可能的原因有多种,例如标定过程中流经传感器的流体过多(特别是正向流),或者调零结果过高。 此警报伴有 A010。 此警报直到关闭并重新打开仪表电源后才会清除。 |
1537 |
标定期间过程太不稳定。 此警报直到关闭并重新打开仪表电源后才会清除。 |
1538 |
推荐措施:
1. 清除或减少机电噪声源(例如,泵、振动、管道应力),关闭并重新打开仪表电源,然后重新执行该程序。 |
1539 |
推荐措施:
1. 确保所有接线盒盖都安装正确。
2. 确保所有变送器线路都符合规范,且所有缆线屏蔽层都正确端接。
3. 检查所有组件的接地情况。
4. 评估环境中的高电磁干扰 (EMI) 源,必要时重新部署变送器或线路。
5. 联系高准公司。 |
1540 |
推荐措施:
1. 对照装置报告的值检查过程条件。
2. 检查有无短路。
3. 联系高准公司。 |
1541 |
计算出的仪表和壳体 RTD 的电阻值超出限制。 |
1542 |
推荐措施:
1. 对照装置报告的值检查过程条件。 温度应在 -200 ºF 至 +400 ºF 之间。
2. 检查有无短路。
3. 联系高准公司。 |
1543 |
部分标定系数未输入或不正确。 |
1544 |
推荐措施:
1. 验证所有特征或标定参数。 参考传感器位号或仪表的标定数据表。 |
1545 |
组态的板类型与实际的板不匹配,或组态的传感器类型与实际的传感器不匹配。 |
1546 |
推荐措施:
1. 验证所有特征或标定参数。 参考传感器位号或仪表的标定数据表。
2. 确保安装了正确的板。 |
1547 |
变送器从安全模式切换到非安全模式。 在非安全模式下,变送器组态可更改。 |
1548 |
推荐措施:
1. 检查 HART 设备 ID。对于 MID 应用,它必须为非零值。 如果设置为 0,请联系厂家。
2. 将变送器设置为安全模式以清除警报。 请注意,可能需要执行其他程序以使装置恢复安全模式。 |
1549 |
内部电子元件故障。 这可能指示变送器与显示模块之间存在通讯丢失。 |
1550 |
推荐措施:
1. 关闭并重新打开仪表电源。
2. 更换显示模块。
3. 联系高准公司。 |
1551 |
附加软件与编程好的电子板类型不兼容。 |
1552 |
来自传感器的信号不足。 这表示传感器管或振动元件不振动。 |
1553 |
推荐措施:
1. 检查流量管中是否存在有空气、未充液、有异物、有涂层或其他过程问题。
2. 通过监测密度值并与从预期密度值获得的结果进行比较,检查流体分离情况。
3. 确保传感器方位符合应用要求。 即使流量管已充满,使用两相流或三相流的设置也可能会产生这种警报。 |
1554 |
The measured viscosity is outside the calibrated range. |
1555 |
Recommended Actions:
1. Check your process conditions against the values reported by the device.
2. Contact Micro Motion. |
1556 |
已知密度检定失败。 |
1557 |
推荐措施:
1. 检查子测试结果并执行推荐措施。
2. 重试测试。
3. 联系高准公司。 |
1558 |
周期信号超出传感器类型的限制。 |
1559 |
推荐措施:
1. 对照装置报告的值检查过程条件。 |
1560 |
计算得出的毫安输出值超出组态范围。 |
1561 |
推荐措施:
1. 检查“上限值”和“下限值”的设置。
2. 检查过程条件。 实际条件可能会超出组态输出时的正常条件。
3. 检查流量管中是否存在有空气、未充液、有异物、有涂层或其他过程问题。
4. 验证并确保已根据您的应用正确组态测量单位。
5. 清洗流量管。 |
1562 |
HART 地址设置为非零值,或毫安输出组态为发送常量值。 |
1563 |
推荐措施:
1. 检查输出是否处于回路测试模式。 如果是,取消输出固定。
2. 如果适合,退出毫安输出调整。
3. 检查 HART 地址。 如果 HART 地址为非零值,您可能需要启用毫安输出操作(回路电流模式)参数。
4. 通过数字通讯检查输出是否设置为常量值。 |
1564 |
推荐措施:
1. 检查驱动增益和传感器电压。
2. 检查有无短路。
3. 检查流量管中是否存在有空气、未充液、有异物、有涂层或其他过程问题。
4. 通过监测密度值并与从预期密度值获得的结果进行比较,检查流体分离情况。
5. 确保传感器方位符合应用要求。 即使流量管已充满,使用两相流或三相流的设置也可能会产生这种警报。 |
1565 |
推荐措施:
1. 等待程序完成。
2. 对于零点标定(仅限于 CDM 仪表),您可以中断标定、将调零时间参数设置为更低的值并重新开始标定。 |
1566 |
The density has exceeded the user-defined two-phase (density) limits. |
1567 |
Recommended Actions:
1. Check that meter is completely full.
2. Check that two-phase limits are set to appropriate values.
3. Verify density cutoff is at least 0.2 g/cc.
4. If you want to change the alert severity, go to the Configure-> Alert Setup-> Alert Severity screen. |
1568 |
HART 猝发模式已启用。 |
1569 |
推荐措施:
1. 无需任何措施。
2. 如果需要,将警报强度重新组态为“忽略”。 |
1570 |
毫安输出组态为发送常量值。 |
1571 |
推荐措施:
1. 检查输出是否处于回路测试模式。 如果是,取消输出固定。
2. 如果适合,退出毫安输出调整。
3. 通过数字通讯检查输出是否设置为常量值。 |
1572 |
与外部测量装置的连接发生故障。 无外部数据可用。 |
1573 |
推荐措施:
1. 验证并确保外部装置正常工作。
2. 验证并确保变送器与外部装置之间的线路完好。
3. 验证 HART 轮询组态。 |
1574 |
测得的温度超出 API 表的范围。 |
1575 |
推荐措施:
1. 对照装置报告的值检查过程条件。
2. 验证 API 参考应用程序的组态和相关参数。 |
1576 |
测得的密度超出 API 表的范围。 |
1577 |
推荐措施:
1. 对照装置报告的值检查过程条件。
2. 验证 API 参考应用程序的组态和相关参数。 |
1578 |
离散输出组态为发送常量值。 |
1579 |
推荐措施:
1. 检查输出是否处于回路测试模式。 如果是,取消输出固定。 |
1580 |
变送器无法从当前数据计算出有效的浓度矩阵。 |
1581 |
推荐措施:
1. 验证浓度测量应用程序的组态。 |
1582 |
管道密度或管道温度超出浓度矩阵以及组态外推限制的范围。 |
1583 |
推荐措施:
1. 对照装置报告的值检查过程条件。
2. 验证浓度测量应用程序的组态。 |
1584 |
管道压力超出 API 表以及外推限制的范围。 |
1585 |
推荐措施:
1. 对照装置报告的值检查过程条件。
2. 验证 API 参考应用程序的组态和相关参数。 |
1586 |
显示模块中出现内存错误。 |
1587 |
推荐措施:
1. 关闭并重新打开仪表电源。
2. 更换显示模块。
3. 联系高准公司。 |
1588 |
推荐措施:
1. 禁用传感器模拟。 |
1589 |
外壳温度与流体温度之间的差值超过 10 °C。 |
1590 |
推荐措施:
1. 检查有无过程紊乱。 管道温度的变化可能不会立即反映到外壳温度。
2. 检查有无短路。
3. 安装加热夹套。
4. 联系高准公司。 |
1591 |
装置中安装了错误的显示模块。 在危险区域中这可能会造成安全违规事件。 |
1592 |
推荐措施:
1. 将安装的显示模块更换为正确的显示模块。 |
1593 |
推荐措施:
1. 验证并确保毫安输出范围设置为适当的值。 |
1594 |
建议措施:
1. 检查 HART 地址。如果非零,设置毫安输出动作参数,允许实时毫安输出。
2. 如果正在进行毫安输出调整,则需退出毫安输出调整。
3. 如果正在进行毫安输出回路测试,则需退出毫安输出回路测试。
4. 检查是否已通过数字通讯或本地显示器将输出设置为固定值。 |
1595 |
建议措施:
1. 这是一个即时报警,故障严重级别警报之一活动时会产生该报警。先解决其它活动报警。 |
1596 |
增强事件 1 已触发。 |
1597 |
推荐措施:
1. 无需任何措施。 |
1598 |
增强事件 2 已触发。 |
1599 |
增强事件 3 已触发。 |
1600 |
增强事件 4 已触发。 |
1601 |
增强事件 5 已触发。 |
1602 |
Velocity switch has been triggered. |
1603 |
No Active Alerts |
1604 |
标定毫安输出以匹配参考仪表。 |
1605 |
如果设备配备有显示屏,则在显示屏上显示图案以便于识别设备。 |
1606 |
这种模式将在显示屏上显示 60 秒。 |
1607 |
单击“下一步”开始。 |
1608 |
这种模式当前显示在装置显示屏上。 |
1609 |
显示屏会在 60 秒后恢复正常工作状态。
要在 60 秒之内恢复正常工作,按设备显示屏上的任意按钮。 |
1610 |
执行密度偏差标定。 |
1611 |
执行温度标定。 |
1612 |
Perform Viscosity Scaling Factor Calibration. |
1613 |
从装置内存读取 Modbus 数据。 |
1614 |
向装置内存写入 Modbus 数据。 |
1615 |
关闭并重新打开装置电源,此时不会影响装置组态。 |
1616 |
使用出厂组态集覆盖现有装置组态。 |
1617 |
Manually reset the configuration changed flag if the host is not able to reset the flag automatically. |
1618 |
验证毫安输出 1 的线路和电流电平。 |
1619 |
验证毫安输出 2 的线路和电流电平。 |
1620 |
验证离散输出的线路和极性。 |
1621 |
模拟值 |
1622 |
组态阵发信息 1 |
1623 |
需要启用阵发信息 1 后才可继续。
此信息将以连续的传输速率阵发。
请选择以下阵发选项: |
1624 |
组态阵发信息 2 |
1625 |
请选择以下阵发选项: |
1626 |
组态阵发信息 3 |
1627 |
阵发信息 1 组态如下所示。是否保存设置? |
1628 |
阵发速率组态 |
1629 |
阵发信息 2 组态如下所示。是否保存设置? |
1630 |
阵发信息 3 组态如下所示。是否保存设置? |
1631 |
主机系统必须能接收阵发信息和事件通知。请先确认再继续组态操作。 |
1632 |
事件通知通过发送阵发信息帮助确定设备的异常状况。 |
1633 |
在屏幕上选择状态变化会触发事件通知的警报。 |
1634 |
需要时调整“触发时间间隔”和“重试速率”的值。
触发时间间隔是警报为触发事件通知而被激活的时间。
重试速率是通过主机确认前事件通知持续显示的时间。 |
1635 |
高级 HART 组态成功完成。 |
1636 |
信息 2 |
1637 |
扩展设备状态 |
1638 |
信息 3 |
1639 |
警告 |
1640 |
已调整触发更新速率。 |
1641 |
错误 |
1642 |
选择了无效的阵发触发模式。 |
1643 |
阵发条件冲突 |
1644 |
矩阵数据 |
1645 |
毫安输出 1 |
1646 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort. |
1647 |
毫安输出 2 |
1648 |
型号 |
1649 |
确保“温度单位”设置为外部温度装置使用的单位。 |
1650 |
Gas Inputs |
1651 |
Percent Carbon Dioxide |
1652 |
Percent Nitrogen |
1653 |
Percent Carbon Monoxide |
1654 |
Percent Hydrogen |
1655 |
“初级变量”自动分配为毫安输出 1。 |
1656 |
“二级变量”自动分配为毫安输出 2(如果可用)。 |
1657 |
“三级变量”仅用于 HART 通讯。 |
1658 |
“四级变量”仅用于 HART 通讯。 |
1659 |
映射变量单位 |
1660 |
初级变量单位 |
1661 |
二级变量单位 |
1662 |
三级变量单位 |
1663 |
四级变量单位 |
1664 |
已知密度检定结果 |
1665 |
已知密度检定失败。 |
1666 |
通过已知密度检定。 |
1667 |
检定周期信号平均值超出可接受的范围。
推荐措施: 确保仪表洁净干燥。 然后重试已知密度检定程序。 |
1668 |
检定周期信号最大偏差超出可接受的范围。
推荐措施: 确保仪表洁净干燥。 然后重试已知密度检定程序。 |
1669 |
管道温度最大偏差 > 1 degC
推荐措施: 验证并确保仪表的环境温度稳定,且仪表
温度在测试位置达到稳定。 然后重试已知密度检定程序。 |
1670 |
驱动增益平均值超出可接受的范围。
推荐措施: 确保仪表洁净干燥。 然后重试已知密度检定程序。 |
1671 |
外壳温度最大偏差 > 1 degC
推荐措施: 验证并确保仪表的环境温度稳定,且仪表
温度在测试位置达到稳定。 然后重试已知密度检定程序。 |
1672 |
用户定义的计算 1 |
1673 |
对于每个计算变量,分配一个过程变量或输入一个常量。 |
1674 |
NOTE: The meter performs this calculation using internal measurement units. If a constant represents a process variable, you must enter its value in the internal measurement unit. If a constant will be used to modify a process variable, you must use the internal measurement units to derive the constant. |
1675 |
用户定义的计算 2 |
1676 |
对于变量 t,分配一个过程变量或输入一个常量。 |
1677 |
用户定义的计算 3 |
1678 |
选择与仪表海拔高度最接近的值。 |
1679 |
验证结果 |
1680 |
Configure the parameters required for concentration measurement. |
1681 |
Configure Concentration Measurement |
1682 |
Select the concentration matrix to be used for measurement. |
1683 |
Define or modify a concentration matrix. |
1684 |
矩阵配置 |
1685 |
Equation Selection Type |
1686 |
Adjust concentration measurement by applying an offset or a slope. |
1687 |
Important: These trim values are applied only to the concentration matrix currently selected for configuration. |
1688 |
矩阵选择 |
1689 |
测警报 |
1690 |
电子条件 |
1691 |
传感器条件 |
1692 |
过程条件 |
1693 |
配置条件 |
1694 |
Referred Matrices |
1695 |
ASTM Referral Matrices |
1696 |
Select the calibration gas and enter the Specific Gravity: |
1697 |
Select the calibration gas and enter the Relative Density: |
1698 |
Select the calibration gas and enter the Molecular Weight: |
1699 |
Calibration is in progress. Allow the sensor to stabilize for a minimum of 15 minutes, then observe the Sensor Time Period and Stability values. When the Calibration Point is Good,click Next to accept the calibration. |
1700 |
Calculated Coefficients |
1701 |
阵发变量 1 |
1702 |
阵发变量 2 |
1703 |
阵发变量 3 |
1704 |
阵发变量 4 |
1705 |
阵发变量 5 |
1706 |
阵发变量 6 |
1707 |
阵发变量 7 |
1708 |
分类 |
1709 |
触发单位 |
1710 |
阵发信息 |
1711 |
事件 |
1712 |
PV Analog Channel Saturated |
1713 |
PV Analog Channel Fixed |
1714 |
Wireless |
1715 |
WirelessHART Device |
1716 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
1717 |
Primary |
1718 |
Primary Analog Channel |
1719 |
Secondary |
1720 |
Secondary Analog Channel |
1721 |
Tertiary |
1722 |
Tertiary Analog Channel |
1723 |
Quaternary |
1724 |
Quaternary Analog Channel |
1725 |
Quinary |
1726 |
Quinary Analog Channel |
1727 |
Critical Power Failure |
1728 |
Lock all |
1729 |
Trim not supported |
1730 |
Configuration cannot be changed |
1731 |
Locked by gateway |
1732 |
Analog Input Channel |
1733 |
Analog Output Channel |
1734 |
Secondary Analog Channel Saturated |
1735 |
Tertiary Analog Channel Saturated |
1736 |
Quaternary Analog Channel Saturated |
1737 |
Quinary Analog Channel Saturated |
1738 |
Secondary Analog Channel Fixed |
1739 |
Tertiary Analog Channel Fixed |
1740 |
Quaternary Analog Channel Fixed |
1741 |
Quinary Analog Channel Fixed |
1742 |
Simulation active |
1743 |
The device is in simulation mode and the level output is not representative of the process. |
1744 |
Non-Volatile memory failure |
1745 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
1746 |
Volatile memory error |
1747 |
The RAM memory check is invalid or maybe corrupt |
1748 |
Watchdog reset executed |
1749 |
A watchdog reset has been performed |
1750 |
Voltage conditions out of range |
1751 |
A voltage condition is outside its allowable range |
1752 |
Environmental conditions out of range |
1753 |
An internal or environmental condition is beyond acceptable limits. |
1754 |
Electronic failure |
1755 |
A hardware problem not related to the sensor has been detected. |
1756 |
Subdevice list changed |
1757 |
Duplicate master detected |
1758 |
Capacity Denied |
1759 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
1760 |
Bandwidth allocation pending |
1761 |
Block transfer pending |
1762 |
Battery life |
1763 |
Percent range |
1764 |
Loop current |
1765 |
Primary variable |
1766 |
Secondary variable |
1767 |
Tertiary variable |
1768 |
Quaternary variable |
1769 |
Read receive time |
1770 |
Write date and time |
1771 |
Non-Volatile Clock |
1772 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
1773 |
Clock Uninitialized |
1774 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
1775 |
None |
1776 |
No restrictions |
1777 |
SI Only |
1778 |
Restricted to SI Unit codes only |
1779 |
Process automation device |
1780 |
Discrete device |
1781 |
Hybrid:Process automation+discrete |
1782 |
I/O System |
1783 |
WirelessHART Process automation device |
1784 |
WirelessHART Discrete device |
1785 |
WirelessHART Process automation+discrete |
1786 |
WirelessHART Gateway |
1787 |
WirelessHART Access point |
1788 |
WirelessHART Process adapter |
1789 |
WirelessHART Discrete adapter |
1790 |
WirelessHART enabled handheld/portable maintenance tool |
1791 |
Device Profile |
1792 |
US |
1793 |
United States of America |
1794 |
JP |
1795 |
Japan |
1796 |
DE |
1797 |
Germany |
1798 |
FR |
1799 |
France |
1800 |
ES |
1801 |
Spain |
1802 |
RU |
1803 |
Russian Federation |
1804 |
CN |
1805 |
People's Republic of China |
1806 |
Secondary Master |
1807 |
Primary Master |
1808 |
Device Diagnostic Status 0 |
1809 |
Device Diagnostic Status 1 |
1810 |
I/O and Subdevice Status |
1811 |
WirelessHART Status |
1812 |
Time Set Option |
1813 |
Real Time Clock Flags |
1814 |
Country |
1815 |
Country of intended device installation |
1816 |
SI Unit Control |
1817 |
Indicates if device should only display SI unit codes |
1818 |
Burst Trigger Units |
1819 |
Enable on token-passing data link layer only |
1820 |
Enable on TDMA data link layer only |
1821 |
Enable on both TDMA and token-passing data link layer |
1822 |
Burst Trigger Class |
1823 |
Continuous |
1824 |
The Burst Message is published continuously at (worst case) the Minimum Update Period. |
1825 |
Window |
1826 |
The Burst Message is triggered when the source value deviates more than the specified trigger value |
1827 |
Rising |
1828 |
The Burst Message is triggered when source value Rises Above the specified trigger value. |
1829 |
Falling |
1830 |
The Burst Message is triggered when the source value Falls Below the specified trigger value. |
1831 |
On-change |
1832 |
The Burst Message is triggered when any value in the message changes. |
1833 |
Off |
1834 |
Enable event notification on token-passing data link layer |
1835 |
Enable event notification on TDMA data link layer |
1836 |
Enable event notification on TDMA and token-passing data link layer |
1837 |
Configuration changed event pending |
1838 |
Device status event pending |
1839 |
More status available event pending |
1840 |
Wired HART Enabled |
1841 |
Enabled on Wired HART device connection only |
1842 |
Wireless HART Enabled |
1843 |
Enabled on WirelessHART device connection only |
1844 |
Both Wired and WirelessHart Enabled |
1845 |
Enabled on both Wired and WirelessHART connections |
1846 |
Burst Msg Trigger Mode |
1847 |
Event Status |
1848 |
Event Notification Control |
1849 |
Device Status Mask |
1850 |
Ext dev status Mask |
1851 |
Device Diagnostic Status 0 Mask |
1852 |
Device Diagnostic Status 1 Mask |
1853 |
Device Diagnostic Status 2 Mask |
1854 |
Device Diagnostic Status 3 Mask |
1855 |
AO saturated Mask |
1856 |
AO fixed Mask |
1857 |
Latched Device Status |
1858 |
Operate mode Mask |
1859 |
Not Defined |
1860 |
Latched Ext dev status |
1861 |
Latched Device Diagnostic Status 0 |
1862 |
Latched Device Diagnostic Status 1 |
1863 |
Latched Device Diagnostic Status 2 |
1864 |
Latched Device Diagnostic Status 3 |
1865 |
Latched AO saturated |
1866 |
Latched AO fixed |
1867 |
Latched Operate Mode |
1868 |
Flow Coriolis Mass |
1869 |
Time Stamp |
1870 |
Relative time value on network |
1871 |
8u |
1872 |
Configuration Change Counter Mismatch |
1873 |
Status bytes mismatch |
1874 |
Device Variable |
1875 |
Analog Channel Number |
1876 |
Analog Channel Number- Index into an array of Analog Channels. |
1877 |
STX Count |
1878 |
Count of STX messages received by this device |
1879 |
ACK Count |
1880 |
Count of ACK messages received by this device |
1881 |
BACK Count |
1882 |
Count of BACK messages received by this device |
1883 |
Burst Command |
1884 |
Cmd 48: Read Additional Device Status |
1885 |
Read Additional Device - This command reads additional device status information |
1886 |
Update Period |
1887 |
Update period for burst message |
1888 |
5.3f |
1889 |
Max Update Period |
1890 |
Maxium update period for burst message |
1891 |
Burst Variable Code |
1892 |
Burst Variable Slot - Device variable code assigned to the slot to be read in burst mode. |
1893 |
Burst Trigger Level |
1894 |
Time First Unack Event Triggered |
1895 |
The time when the first unacknowledged event was triggered |
1896 |
Event Notification Retry Time |
1897 |
Max Update Time |
1898 |
Burst Message Number |
1899 |
Total Burst Msgs |
1900 |
Total number of burst messages supported by this device |
1901 |
Event Number |
1902 |
Event Debounce Interval |
1903 |
Device Specific Status 0 Mask |
1904 |
Device Specific Status 1 Mask |
1905 |
Device Specific Status 2 Mask |
1906 |
Device Specific Status 3 Mask |
1907 |
Device Specific Status 4 Mask |
1908 |
Device Specific Status 5 Mask |
1909 |
Device Specific Status 14 Mask |
1910 |
Device Specific Status 15 Mask |
1911 |
Device Specific Status 16 Mask |
1912 |
Device Specific Status 17 Mask |
1913 |
Device Specific Status 18 Mask |
1914 |
Device Specific Status 19 Mask |
1915 |
Device Specific Status 20 Mask |
1916 |
Latched Device Specific Status 0 |
1917 |
Latched Device Specific Status 1 |
1918 |
Latched Device Specific Status 2 |
1919 |
Latched Device Specific Status 3 |
1920 |
Latched Device Specific Status 4 |
1921 |
Latched Device Specific Status 5 |
1922 |
Latched Device Specific Status 14 |
1923 |
Latched Device Specific Status 15 |
1924 |
Latched Device Specific Status 16 |
1925 |
Latched Device Specific Status 17 |
1926 |
Latched Device Specific Status 18 |
1927 |
Latched Device Specific Status 19 |
1928 |
Latched Device Specific Status 20 |
1929 |
Latched Cfg chng count |
1930 |
Number of Events Supported |
1931 |
Burst Variable Codes |
1932 |
Update time adjusted |
1933 |
Invalid burst message |
1934 |
Invalid device variable classification |
1935 |
Invalid burst trigger mode selection code |
1936 |
Burst condition conflict |
1937 |
Update period increased |
1938 |
Insufficient bandwidth |
1939 |
Update period or debounce interval adjusted |
1940 |
Update times adjusted |
1941 |
Update rate uncertain |
1942 |
Not all events cleared |
1943 |
NOTE-Send TV assignment before modifying range values. |
1944 |
现在可以安全地将控制回路恢复为自动控制模式。 |
1945 |
更改组态可能影响变送器操作。 确保将控制回路设置为手动操作后再继续。 |
1946 |
select |
1947 |
由于现场装置错误而中断。 |
1948 |
warning |
1949 |
select1 |
1950 |
select2 |
1951 |
select3 |
1952 |
完成 |
1953 |
输入调零时间: |
1954 |
Prepare the meter:
1. Allow the meter to warm up for at least 20 minutes after applying power.
2. Run the process fluid through the meter until the sensor temperature reaches the normal process operating temperature.
3. Stop flow through the meter by shutting the downstream valve, and then the upstream valve if available.
4. Verify that the meter is blocked in, that flow has stopped, and that the meter is completely full of process fluid.
5. Observe the drive gain, temperature, and density readings. If they are stable, check the Live Zero value. If the average value is close to 0, you should not need to zero the meter.
Click Next when meter preparation is complete. |
1955 |
调零标定进行中... |
1956 |
Press Next to reset the data for all six matrices.
Warning: All matrix data will be deleted from the device. |
1957 |
所有矩阵数据都已从装置中删除。 |
1958 |
Press Next to refresh the Concentration Measurement screen. |
1959 |
选择锁定选项: |
1960 |
永久;暂时;全部锁定 |
1961 |
不能使设备解锁。
设备被其他主机锁定。 |
1962 |
所有警报已确认。 |
1963 |
选择 API 表号: |
1964 |
选择 API 表字符: |
1965 |
A;B;D |
1966 |
A;B;D;E |
1967 |
C;E |
1968 |
输入热膨胀系数: |
1969 |
输入参考温度: |
1970 |
Enter Reference Pressure (Gauge): |
1971 |
Select the mA output to configure: |
1972 |
mA Output 1;mA Output 2;Exit |
1973 |
Keep Existing Assignment;Temperature;Density;Density at Reference (API);Density at Reference (CM);Specific Gravity (CM);Concentration (CM);Drive Gain;Case Temperature;External Pressure (Gauge);External Temperature;User-Defined Calculation Output;Tube-Case Temperature Difference;Velocity;Sensor Time Period; |
1974 |
Keep Existing Assignment;Temperature;Drive Gain;Case Temperature;External Pressure (Gauge);External Temperature;User-Defined Calculation Output;Tube-Case Temperature Difference;Velocity;Sensor Time Period; |
1975 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Temperature. |
1976 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Density. |
1977 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Density at Reference (API). |
1978 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Density at Reference (CM). |
1979 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Specific Gravity (CM). |
1980 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Concentration (CM). |
1981 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Drive Gain. |
1982 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Case Temperature. |
1983 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to External Pressure (Gauge). |
1984 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to External Temperature. |
1985 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to User-Defined Calculation Output. |
1986 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Tube-Case Temperature Difference. |
1987 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Velocity. |
1988 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Sensor Time Period. |
1989 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Temperature. |
1990 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Density. |
1991 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Density at Reference (API). |
1992 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Density at Reference (CM). |
1993 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Specific Gravity (CM). |
1994 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Concentration (CM). |
1995 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Drive Gain. |
1996 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Case Temperature. |
1997 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to External Pressure (Gauge). |
1998 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to External Temperature. |
1999 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to User-Defined Calculation Output. |
2000 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Tube-Case Temperature Difference. |
2001 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Velocity. |
2002 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Sensor Time Period. |
2003 |
您将使用出厂值覆盖设备组态。 在继续之前,将控制回路设置为手动操作,并确保现有的组态值已记录或保存在文件中。 如果设备连接丢失,请使用默认设置重新连接。 要继续吗? |
2004 |
Factory values are restored successfully. |
2005 |
连接参考仪表。 |
2006 |
Click Next to set mA Output 1 to %0 mA. |
2007 |
输入从参考仪表获得的读数: |
2008 |
装置输出与参考仪表是否匹配? |
2009 |
Click Next to set mA Output 2 to %0 mA. |
2010 |
设备被锁定,无法组态参数。 |
2011 |
将毫安输出 1 恢复为正常功能。 |
2012 |
将毫安输出 2 恢复为正常功能。 |
2013 |
Modbus 数据类型: |
2014 |
线圈;未分配 16(U16);未分配 32(U32);浮点;退出 |
2015 |
要读取的线圈地址: |
2016 |
number |
2017 |
要读取的寄存器地址: |
2018 |
线圈值为 %0,异常码为 %1。 |
2019 |
U16 值为 %0,异常码为 %1。 |
2020 |
U32 值为 %0,异常码为 %1。 |
2021 |
浮点值为 %0,异常码为 %1。 |
2022 |
要写入的线圈地址: |
2023 |
要写入的寄存器地址: |
2024 |
要写入的线圈值: |
2025 |
0;1 |
2026 |
要写入的 U16 值: |
2027 |
uValue |
2028 |
要写入的 U32 值: |
2029 |
要写入的浮点值: |
2030 |
fValue |
2031 |
选择毫安输出 1 水平: |
2032 |
毫安输出 1 固定为 %0 mA。 |
2033 |
选择毫安输出 2 水平: |
2034 |
毫安输出 2 固定为 %0 mA。 |
2035 |
选择离散输出状态: |
2036 |
关闭;打开;结束 |
2037 |
Discrete Output State is OFF. |
2038 |
Discrete Output State is ON. |
2039 |
Returning Discrete Output to normal function. |
2040 |
Select pressure (absolute) unit used by the external pressure (absolute) device: |
2041 |
Select pressure (gauge) unit used by the external pressure (gauge) device: |
2042 |
选择警报代码: |
2043 |
当前严重级别为 %0。 |
2044 |
选择警报严重级别: |
2045 |
查看警报严重级别: |
2046 |
将 %{1} 设置为 %{0}。 |
2047 |
A001:EEPROM 错误 (CP) 设置为 %0。 |
2048 |
A002:RAM 错误 (CP) 设置为 %0。 |
2049 |
A003:无传感器响应设置为 %0。 |
2050 |
A004:温度超限设置为 %0。 |
2051 |
A005:质量流量超限设置为 %0。 |
2052 |
A006:所需特征化设置为 %0。 |
2053 |
A008:密度超限设置为 %0。 |
2054 |
A009:变送器初始化预热设置为 %0。 |
2055 |
A010:校准失败设置为 %0。 |
2056 |
A011:调零校准失败:低设置为 %0。 |
2057 |
A012:调零校准失败:高设置为 %0。 |
2058 |
A013:调零校准失败:不稳定设置为 %0。 |
2059 |
A014:变送器故障设置为 %0。 |
2060 |
A016:传感器 RTD 故障设置为 %0。 |
2061 |
A017:T 系列 RTD 故障设置为 %0。 |
2062 |
A020:无流量校准值设置为 %0。 |
2063 |
A021:无 K1 值设置为 %0。 |
2064 |
A027:安全漏洞设置为 %0。 |
2065 |
A029: 内部通讯故障设置为 %0。 |
2066 |
A030: 硬件/软件不兼容设置为 %0。 |
2067 |
A033:无 LPO RPO 信号设置为 %0。 |
2068 |
A037: 传感器检查失败设置为 %0。 |
2069 |
A038: 周期超出范围设置为 %0。 |
2070 |
A100:MAO1 饱和设置为 %0。 |
2071 |
A101:MAO1 固定设置为 %0。 |
2072 |
A102:驱动超限设置为 %0。 |
2073 |
A104:正在进行校准设置为 %0。 |
2074 |
A106:启用阵发模式设置为 %0。 |
2075 |
A107:发生电源复位设置为 %0。 |
2076 |
A113:MAO2 饱和设置为 %0。 |
2077 |
A114:MAO2 固定设置为 %0。 |
2078 |
A115:无外部数据设置为 %0。 |
2079 |
A116:API:温度超限设置为 %0。 |
2080 |
A117:API:密度超限设置为 %0。 |
2081 |
A118:DO1 固定设置为 %0。 |
2082 |
A120:ED:曲线拟合失败设置为 %0。 |
2083 |
A121:ED:推测报警设置为 %0。 |
2084 |
A122: API:压力超出标准范围设置为 %0。 |
2085 |
A132:传感器仿真激活设置为 %0。 |
2086 |
A133: 本地显示固件校验和错误设置为 %0。 |
2087 |
A134: 外壳-仪表温差超出限制设置为 %0。 |
2088 |
A136: 显示类型不正确设置为 %0。 |
2089 |
Select the application option: |
2090 |
API 参考;浓度测量;无 |
2091 |
API Referral;None |
2092 |
Concentration Measurement;None |
2093 |
注: MID 已启用,无法更改。
选择应用程序选项: |
2094 |
注: MID 已禁用,无法更改。
选择应用程序选项: |
2095 |
应用程序已禁用。 您要启用它吗? |
2096 |
应用程序已启用。 您要禁用它吗? |
2097 |
API 参考为 %0。 |
2098 |
浓度测量为 %0。 |
2099 |
API application is disabled. |
2100 |
Concentration Measurement is disabled. |
2101 |
所有应用程序都已禁用。 |
2102 |
Current viscosity value range is Ultra Low Range (0.50 – 10.46 cP).
Select the viscosity value range: |
2103 |
Ultra Low Range (0.50 – 10.46 cP);Low (10.46 – 108.16 cP);Medium (108.16 – 1087.43 cP);High (1087.43 – 13345.64 cP) |
2104 |
Current viscosity value range is Low (10.46 – 108.16 cP).
Select the viscosity value range: |
2105 |
Current viscosity value range is Medium (108.16 – 1087.43 cP).
Select the viscosity value range: |
2106 |
Current viscosity value range is High (1087.43 – 13345.64 cP).
Select the viscosity value range: |
2107 |
注: 应当首先完成偏差标定,然后再进行斜率标定。
选择所需的选项: |
2108 |
温度偏差标定;温度斜率标定 |
2109 |
温度偏差标定: |
2110 |
标定进行中... |
2111 |
标定完成。 |
2112 |
温度斜率标定: |
2113 |
Known Density Verification has been aborted. |
2114 |
Prepare the meter:
1. Remove the meter from the process and place it in a protected stable environment.
2. Minimize variation in ambient temperature.
3. Eliminate vibration.
4. Ensure that a suitably rated vacuum pump is used that is able to draw a vacuum on the meter. Pull a vacuum inside the meter to a maximum pressure of 0.2 Torr.
5. Apply power to the meter.
Click Next when meter preparation is complete. |
2115 |
Prepare the meter:
1. Remove the meter from the process and place it in a protected stable environment.
2. Minimize variation in ambient temperature.
3. Eliminate vibration.
4. Ensure that the meter is clean. Check for deposition on the vibrating element or the inner walls of the device.
5. Apply power to the meter.
Click Next when meter preparation is complete. |
2116 |
已知密度检定需要 20 份样品,完成验证大约需要 20 秒。
单击“下一步”继续。 |
2117 |
Known Density Verification in progress... |
2118 |
No factory calibration present. Known Density Verification cannot be run. |
2119 |
在显示屏上显示易识别的图案有助于识别设备。
单击“确定”开始。 |
2120 |
设备正显示图案,使其可被识别。
显示屏会在 60 秒后恢复正常工作状态,按设备显示屏上的任意按钮可中断这种状态。 |
2121 |
Active Calculation: User-Defined Calculation 1 |
2122 |
无;用户定义的计算 1;用户定义的计算 2;用户定义的计算 3 |
2123 |
Active Calculation: User-Defined Calculation 2 |
2124 |
Active Calculation: User-Defined Calculation 3 |
2125 |
活动计算: 无 |
2126 |
None;User-Defined Calculation 1;User-Defined Calculation 2 |
2127 |
您已选择不使用用户定义的计算。
点击“下一步”继续。 |
2128 |
User-Defined Calculation Output value is %0 %1 |
2129 |
密度:
值 %0 %1
温度:
值 %2 %3
速度:
值 %4 %5 |
2130 |
User-Defined Calculation Output:
Value %0 |
2131 |
Concentration (CM):
Value %0 %1
Density at Reference (CM):
Value %2 %3
Temperature:
Value %4 %5 |
2132 |
User-Defined Calculation Output:
Value %0
Velocity:
Value %1 %2
Specific Gravity:
Value %3 |
2133 |
Density at Reference (API):
Value %0 %1
Temperature:
Value %2 %3
User-Defined Calculation Output:
Value %4 |
2134 |
Velocity:
Value %0 %1 |
2135 |
Temperature:
Value %0 %1
Velocity:
Value %2 %3
User-Defined Calculation Output:
Value %4 |
2136 |
单击“下一步”开始密度偏差标定。 |
2137 |
Access Restricted |
2138 |
Resetting device. Please wait. |
2139 |
设备复位成功完成。 |
2140 |
按“OK”(确认)重置全部六个矩阵数据,或按“ABORT”(中止)不重置数据。 |
2141 |
Press OK to refresh the Concentration Measurement screen. |
2142 |
Priority: Failed - Fix Now
The transmitter has detected a problem communicating with the sensor.
Recommended Actions:
1. Cycle power to the meter.
2. Contact Micro Motion. |
2143 |
Priority: Failed - Fix Now
The transmitter is not receiving one or more basic electrical signals from the sensor.
Recommended Actions:
1. Check the drive gain and the pickoff voltage.
2. Check for electrical shorts.
3. Check the integrity of the sensor tubes. |
2144 |
Priority: Failed - Fix Now
The RTD resistance is out of range for the sensor.
Recommended Actions:
1. Check for electrical shorts.
2. Verify temperature characterization or calibration parameters.
3. Check your process conditions against the values reported by the device.
4. Contact Micro Motion. |
2145 |
Priority: Failed - Fix Now
The measured velocity is greater than the maximum velocity limit of the sensor.
Recommended Actions:
1. If other alerts are present, resolve those alert conditions first. If the current alert persists, continue with the recommended actions.
2. Check your process conditions against the values reported by the device.
3. Check for two-phase flow. |
2146 |
Priority: Failed - Fix Now
Calibration factors have not been entered, or the sensor type is incorrect, or the calibration factors are incorrect for the sensor type.
Recommended Actions:
1. Verify all of the characterization or calibration parameters. See the sensor tag or the calibration sheet for your meter.
2. Check for electrical shorts.
3. Contact Micro Motion. |
2147 |
Priority: Failed - Fix Now
The line density is greater than 10 g/cm3.
Recommended Actions:
1. If other alerts are present, resolve those alerts conditions first. If the current alert persists, continue with the recommended actions.
2. Check for air in the flow tubes, tubes not filled, foreign material in the tubes, coating in the tubes, or other process problems.
3. Check for two-phase flow.
4. If accompanied by an A003 alert, check for electrical shorts.
5 Verify all of the characterization or calibration parameters. See the sensor tag or the calibration sheet for your meter.
6. Check the drive gain and the pickoff voltage.
7. Contact Micro Motion. |
2148 |
Priority: Failed - Fix Now
Transmitter is in power-up mode.
Recommended Actions:
1. Allow the meter to complete its power-up sequence. The alert should clear automatically.
2. Verify that the tubes are full of process fluid.
3. Verify that the transmitter is receiving sufficient power.
4. If it is not, correct the problem and cycle power to the meter.
5. If it is, this suggests that the transmitter has an internal power issue. Replace the transmitter. |
2149 |
Priority: Failed - Fix Now
Many possible causes, such as too much flow through the sensor during a calibration procedure, or a zero result that is out of range. This alert will not clear until you cycle power to the meter.
Recommended Actions:
1. If this alert appears during zeroing (CDM meters only), verify that there is no flow through the sensor, cycle power to the meter, then retry the procedure. |
2150 |
Priority: Failed - Fix Now
Many possible causes, such as too much flow, especially reverse flow, through the sensor during a calibration procedure, or a zero result that is too low. This alert is accompanied by A010. This alert will not clear until you cycle power to the meter.
Recommended Actions:
1. Verify that there is no flow through the sensor, cycle power to the meter, then retry the procedure. |
2151 |
Priority: Failed - Fix Now
Many possible causes, such as too much flow, especially forward flow, through the sensor during a calibration procedure, or a zero result that is too high. This alert is accompanied by A010. This alert will not clear until you cycle power to the meter.
Recommended Actions:
1. Verify that there is no flow through the sensor, cycle power to the meter, then retry the procedure. |
2152 |
Priority: Failed - Fix Now
There was too much process instability during the calibration procedure. This alert will not clear until you cycle power to the meter.
Recommended Actions:
1. Remove or reduce sources of electromechanical noise (e.g., pumps, vibration, pipe stress), cycle power to the meter, then retry the procedure. |
2153 |
Priority: Failed - Fix Now
Many possible causes.
Recommended Actions:
1. Ensure that all wiring compartment covers are installed correctly.
2. Ensure that all transmitter wiring meets specifications and that all cable shields are properly terminated.
3. Check the grounding of all components.
4. Evaluate the environment for sources of high electromagnetic interference (EMI) and relocate the transmitter or wiring as necessary.
5. Contact Micro Motion. |
2154 |
Priority: Failed - Fix Now
The value computed for the resistance of the Line RTD is outside limits.
Recommended Actions:
1. Check your process conditions against the values reported by the device.
2. Check for electrical shorts.
3. Contact Micro Motion. |
2155 |
Priority: Maintenance - Fix Soon
The value computed for the resistance of the meter and case RTDs are outside limits.
Recommended Actions:
1.Check your process conditions against the values reported by the device. Temperature should be bettween – 200 °F and +400 °F.
2. Check for electrical shorts.
3. Contact Micro Motion. |
2156 |
Priority: Failed - Fix Now
Some calibration factors have not been entered or are incorrect.
Recommended Actions:
1. Verify all of the characterization or calibration parameters. See the sensor tag or the calibration sheet for your meter. |
2157 |
Priority: Failed - Fix Now
The configured board type does not match the physical board, or the configured sensor type does not match the physical sensor.
Recommended Actions:
1. Verify all of the characterization or calibration parameters. See the sensor tag or the calibration sheet for your meter.
2. Ensure that the correct board is installed. |
2158 |
Priority: Failed - Fix Now
The transmitter has been switched from secure mode to unsecure mode. In unsecured mode, the transmitter configuration can be changed.
Recommended Actions:
1. Check the HART device ID. For MID applications, it must be non-zero. If it is set to 0, contact the factory.
2. Set the transmitter to secure to clear the alert. Note that additional procedures may be required to return the device to secured mode. |
2159 |
Priority: Failed - Fix Now
Internal electronics failure. This can indicate a loss of communication between the transmitter and the display module.
Recommended Actions:
1. Cycle power to the meter.
2. Replace the display module.
3. Contact Micro Motion. |
2160 |
Priority: Failed - Fix Now
The loaded software is not compatible with the programmed board type.
Recommended Actions:
1. Cycle power to the meter.
2. Contact Micro Motion. |
2161 |
Priority: Failed - Fix Now
The signal from the sensor pick-off(s) is insufficient. This suggests that the sensor tubes or vibrating elements are not vibrating.
Recommended Actions:
1. Check for air in the flow tubes, tubes not filled, foreign material in the tubes, coating in the tubes, or other process problems.
2. Check for fluid separation by monitoring the density value and comparing the results against expected density values.
3. Ensure that the sensor orientation is appropriate for your application. Settling from a two-phase or three-phase fluid can cause this alert even if the flow tubes are full. |
2162 |
Priority: Failed - Fix Now
The measured viscosity is outside the calibrated range.
Recommended Actions:
1. Check your process conditions against the values reported by the device.
2. Contact Micro Motion. |
2163 |
Priority: Failed - Fix Now
Known Density Verification failed.
Recommended Actions:
1. Check the subtest results and perform the recommended actions.
2. Retry the test.
3. Contact Micro Motion. |
2164 |
Priority: Failed - Fix Now
The time period signal is outside the limits for the sensor type.
Recommended Actions:
1. Check your process conditions against the values reported by the device. |
2165 |
Priority: Advisory
The calculated mA output value is outside the configured range.
Recommended Actions:
1. Check the settings of ‘Upper Range Value’ and ‘Lower Range Value.’
2. Check process conditions. Actual conditions may be outside the normal conditions for which the output is configured.
3. Check for air in the flow tubes, tubes not filled, foreign material in the tubes, coating in the tubes, or other process problems.
4. Verify that the measurement units are configured correctly for your application.
5. Purge the flow tubes. |
2166 |
Priority: Advisory
The HART address is set to a non-zero value, or the mA output is configured to send a constant value.
Recommended Actions:
1. Check whether the output is in loop test mode. If it is, unfix the output.
2. Exit mA output trim, if applicable.
3. Check the HART address. If the HART address is non-zero, you may need to enable the mA Output Action (Loop Current Mode) parameter.
4. Check whether the output has been set to a constant value via digital communication. |
2167 |
Priority: Maintenance - Fix Soon
The drive power (current/voltage) is at its maximum.
Recommended Actions:
1. Check the drive gain and the pickoff voltage.
2. Check for electrical shorts.
3. Check for air in the flow tubes, tubes not filled, foreign material in the tubes, coating in the tubes, or other process problems.
4. Check for fluid separation by monitoring the density value and comparing the results against expected density values.
5. Ensure that the sensor orientation is appropriate for your application. Settling from a two-phase or three-phase fluid can cause this alert even if the flow tubes are full. |
2168 |
Priority: Advisory
A calibration procedure is in progress.
Recommended Actions:
1. Allow the procedure to complete.
2. For zero calibration (CDM meters only), you may abort the calibration, set the zero time parameter to a lower value, and restart the calibration. |
2169 |
Priority: Maintenance - Fix Soon
The density has exceeded the user-defined two-phase (density) limits.
Recommended Actions:
1. Check that meter is completely full.
2. Check that two-phase limits are set to appropriate values.
3. Verify density cutoff is at least 0.2 g/cc.
4. If you want to change the alert severity, go to the Configure-> Alert Setup-> Alert Severity screen. |
2170 |
Priority: Advisory
HART burst mode is enabled.
Recommended Actions:
1. No action required.
2. If desired, you can reconfigure the alert severity level to Ignore. |
2171 |
Priority: Maintenance - Fix Soon
The transmitter has been restarted.
Recommended Actions:
1. No action required.
2. If desired, you can reconfigure the alert severity level to Ignore. |
2172 |
Priority: Advisory
The mA output is configured to send a constant value.
Recommended Actions:
1. Check whether the output is in loop test mode. If it is, unfix the output.
2. Exit mA output trim, if applicable.
3. Check whether the output has been set to a constant value via digital communication. |
2173 |
Priority: Maintenance - Fix Soon
The connection to an external measurement device has failed. No external data is available.
Recommended Actions:
1. Verify that the external device is operating correctly.
2. Verify the wiring between the transmitter and the external device.
3. Verify the HART polling configuration. |
2174 |
Priority: Maintenance - Fix Soon
The measured temperature is outside the range of the API table.
Recommended Actions:
1. Check your process conditions against the values reported by the device.
2. Verify the configuration of the API referral application and related parameters. |
2175 |
Priority: Maintenance - Fix Soon
The measured density is outside the range of the API table.
Recommended Actions:
1. Check your process conditions against the values reported by the device.
2. Verify the configuration of the API referral application and related parameters. |
2176 |
Priority: Advisory
The discrete output has been configured to send a constant value.
Recommended Actions:
1. Check whether the output is in loop test mode. If it is, unfix the output. |
2177 |
Priority: Advisory
The transmitter was unable to calculate a valid concentration matrix from the current data.
Recommended Actions:
1. Verify the configuration of the concentration measurement application. |
2178 |
Priority: Maintenance - Fix Soon
The line density or line temperature is outside the range of the concentration matrix plus the configured extrapolation limit.
Recommended Actions:
1. Check your process conditions against the values reported by the device.
2. Verify the configuration of the concentration measurement application. |
2179 |
Priority: Maintenance - Fix Soon
The line pressure is outside the range of the API table plus the extrapolation limit.
Recommended Actions:
1. Check your process conditions against the values reported by the device.
2. Verify the configuration of the API Referral application and related parameters. |
2180 |
Priority: Advisory
Simulation mode is enabled.
Recommended Actions:
1. Disable sensor simulation. |
2181 |
Priority: Maintenance - Fix Soon
There is a memory error in the display module.
Recommended Actions:
1. Cycle power to the meter.
2. Replace the display module.
3. Contact Micro Motion. |
2182 |
Priority: Maintenance - Fix Soon
The difference between the case temperature and the fluid temperature is greater than 10 °C.
Recommended Actions:
1. Check for process upsets. Changes to line temperature may not be immediately reflected in case temperature.
2. Check for electrical shorts.
3. Install a heat jacket.
4. Contact Micro Motion. |
2183 |
Priority: Maintenance - Fix Soon
An incorrect display module has been installed on the device. This may cause a safety violation in hazardous areas.
Recommended Actions:
1. Replace the installed display module with an appropriate display module. |
2184 |
Priority: Advisory
Enhanced Event 1 has been triggered.
Recommended Actions:
1. No action required. |
2185 |
Priority: Advisory
Enhanced Event 2 has been triggered.
Recommended Actions:
1. No action required. |
2186 |
Priority: Advisory
Enhanced Event 3 has been triggered.
Recommended Actions:
1. No action required. |
2187 |
Priority: Advisory
Enhanced Event 4 has been triggered.
Recommended Actions:
1. No action required. |
2188 |
Priority: Advisory
Enhanced Event 5 has been triggered.
Recommended Actions:
1. No action required. |
2189 |
Priority: Advisory
Velocity switch has been triggered.
Recommended Actions:
1. No action required. |
2190 |
Priority: Maintenance - Fix Soon
Non-zero HART address configured, or user has set the mA output to a fixed value.
Recommended Actions:
1. Check the HART address. If non-zero, set mA Output Action parameter to enable Live mA Output.
2. If mA output trim is being performed, exit mA output trim.
3. If mA output loop test is being performed, exit mA output loop test.
4. Check to see if the output has been fixed via digital communications or the local display. |
2191 |
Priority: Maintenance - Fix Soon
The calculated amount of current output is outside of the linear range.
Recommended Actions:
1. Verify that mA output range values are set to appropriate values. |
2192 |
Prepare the meter:
1. Remove the meter from the process and place it in a protected stable environment.
2. Minimize variation in ambient temperature.
3. Eliminate vibration.
4. Ensure that a suitably rated vacuum pump is used that is able to draw a vacuum on the meter. Pull a vacuum inside the meter to a maximum pressure of 0.2 Torr.
Click OK when meter preparation is complete. |
2193 |
Prepare the meter:
1. Remove the meter from the process and place it in a protected stable environment.
2. Minimize variation in ambient temperature.
3. Eliminate vibration.
4. Ensure that the meter is clean. Check for deposition on the vibrating element or the inner walls of the device.
Click OK when meter preparation is complete. |
2194 |
Known Density Verification requires 20 samples, and will take about 20 seconds to complete.
Click OK to continue. |
2195 |
已知密度检定进行中... |
2196 |
Know Density Verification: %0
Time Period Diff: %1
Time Period Stability: %2
Fluid Temp: %3
Drive Gain: %4
Case Temp Stability: %5 |
2197 |
Know Density Verification: %0
Time Period Diff: %1
Time Period Stability: %2
Fluid Temp: %3
Drive Gain: %4 |
2198 |
Click OK to begin density offset calibration. |
2199 |
Keep Existing Assignment;Temperature;Density;Density at Reference (API);Density at Reference (CM);Specific Gravity (CM);Net Mass Flow Rate (CM);Net Volume Flow Rate (CM);Concentration (CM);Drive Gain;Case Temperature;External Pressure (Gauge);External Temperature;User-Defined Calculation Output;Volume from Mag/Vortex Meter;Calculated Mass Flow from Mag Meter Input;Sensor Time Period; |
2200 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Net Mass Flow Rate (CM). |
2201 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Net Volume Flow Rate (CM). |
2202 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Volume from Mag/Vortex Meter. |
2203 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Calculated Mass Flow from Mag Meter Input. |
2204 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Net Mass Flow Rate (CM). |
2205 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Net Volume Flow Rate (CM). |
2206 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Volume from Mag/Vortex Meter. |
2207 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Calculated Mass Flow from Mag Meter Input. |
2208 |
Keep Existing Assignment;Temperature;Density;Density at Reference (API);Density at Reference (CM);Specific Gravity (CM);Net Mass Flow Rate (CM);Net Volume Flow Rate (CM);Concentration (CM);Drive Gain;External Pressure (Gauge);External Temperature;User-Defined Calculation Output;Dynamic Viscosity;Kinematic Viscosity;Referred Viscosity;Referred Viscosity (Secondary);Volume from Mag/Vortex Meter;Calculated Mass Flow from Mag Meter Input; |
2209 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Dynamic Viscosity. |
2210 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Kinematic Viscosity. |
2211 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Referred Viscosity. |
2212 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Referred Viscosity (Secondary). |
2213 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Dynamic Viscosity. |
2214 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Kinematic Viscosity. |
2215 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Referred Viscosity. |
2216 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Referred Viscosity (Secondary). |
2217 |
Keep Existing Assignment;Temperature;Density;Drive Gain;External Pressure (Absolute);External Temperature;User-Defined Calculation Output;Specific Gravity (Gas);Relative Density;Molecular Weight;Volume from Mag/Vortex Meter;Base Density (Gas);Calculated Mass Flow from Mag Meter Input;Volume Flow at Reference Temperature;Mass from Coriolis Meter;Sensor Time Period; |
2218 |
Keep Existing Assignment;Temperature;Drive Gain;External Pressure (Absolute);External Temperature;User-Defined Calculation Output;Specific Gravity (Gas);Relative Density;Molecular Weight;Volume from Mag/Vortex Meter;Base Density (Gas);Calculated Mass Flow from Mag Meter Input;Volume Flow at Reference Temperature;Mass from Coriolis Meter;Sensor Time Period; |
2219 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to External Pressure (Absolute). |
2220 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Specific Gravity (Gas). |
2221 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Relative Density. |
2222 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Molecular Weight. |
2223 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Base Density (Gas). |
2224 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Volume Flow at Reference Temperature. |
2225 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Mass from Coriolis Meter. |
2226 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to External Pressure (Absolute). |
2227 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Specific Gravity (Gas). |
2228 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Relative Density. |
2229 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Molecular Weight. |
2230 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Base Density (Gas). |
2231 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Volume Flow at Reference Temperature. |
2232 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Mass from Coriolis Meter. |
2233 |
Keep Existing Assignment;Temperature;Density;Density at Reference (API);Drive Gain;External Temperature;User-Defined Calculation Output;Dynamic Viscosity;Kinematic Viscosity;Referred Viscosity;Referred Viscosity (Secondary);CCAI;CII; |
2234 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to CCAI. |
2235 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to CII. |
2236 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to CCAI. |
2237 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to CII. |
2238 |
A036: Viscosity Out of Limit is set to %0. |
2239 |
Keep Existing Assignment;Temperature;Density;Net Mass Flow Rate (CM);Net Volume Flow Rate (CM);Concentration (CM);Drive Gain;External Pressure (Absolute);External Temperature;User-Defined Calculation Output;Specific Gravity (Gas);Molecular Weight;Wobbe Index;Calorific Value;Percent CO2;Volume from Mag/Vortex Meter;Base Density (Gas);Calculated Mass Flow from Mag Meter Input;Percent N2;Percent H2;Percent CO;Volume Flow at Reference Temperature;Mass from Coriolis Meter;Energy Flow;Sensor Time Period; |
2240 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Wobbe Index. |
2241 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Calorific Value. |
2242 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Percent CO2. |
2243 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Percent N2. |
2244 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Percent H2. |
2245 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Percent CO. |
2246 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Primary Variable is set to Energy Flow. |
2247 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Wobbe Index. |
2248 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Calorific Value. |
2249 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Percent CO2. |
2250 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Percent N2. |
2251 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Percent H2. |
2252 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Percent CO. |
2253 |
Any changes made on this screen will be written to the device when the Next button is clicked and could affect transmitter operation. Click Next to write changes to the device, or click Cancel to abort.
Secondary Variable is set to Energy Flow. |
2254 |
Keep Existing Assignment;Temperature;Density;Net Mass Flow Rate (CM);Net Volume Flow Rate (CM);Concentration (CM);Drive Gain;External Pressure (Absolute);External Temperature;User-Defined Calculation Output;Relative Density;Percent CO2;Volume from Mag/Vortex Meter;Base Density (Gas);Calculated Mass Flow from Mag Meter Input;Percent N2;Percent H2;Percent CO;Volume Flow at Reference Temperature;Mass from Coriolis Meter;Sensor Time Period; |
2255 |
Dynamic Viscosity:
Value %0 %1
Kinematic Viscosity:
Value %2 %3
Referred Viscosity:
Value %4 %5 |
2256 |
Referred Viscosity (Secondary):
Value %0 %1
Density:
Value %2 %3
Temperature:
Value %4 %5 |
2257 |
Dynamic Viscosity:
Value %0 %1
Kinematic Viscosity:
Value %2 %3
Density:
Value %4 %5 |
2258 |
Temperature:
Value %0 %1 |
2259 |
Kinematic Viscosity:
Value %0 %1
Temperature:
Value %2 %3 |
2260 |
Keep Existing Assignment;Temperature;Density;Dynamic Viscosity;Kinematic Viscosity; |
2261 |
Keep Existing Assignment;Temperature; |
2262 |
Keep Existing Assignment;Kinematic Viscosity; |
2263 |
Density:
Value %0 %1
Temperature:
Value %2 %3 |
2264 |
Specific Gravity (Gas):
Value %0
Molecular Weight:
Value %1 %2
Density:
Value %3 %4 |
2265 |
Specific Gravity (Gas):
Value %0
Molecular Weight:
Value %1 %2
Base Density (Gas):
Value %3 %4 |
2266 |
Relative Density:
Value %0
Base Density (Gas):
Value %1 %2
Density:
Value %3 %4 |
2267 |
Current referred viscosity method is ASTM D341 Single.
Select referred viscosity method: |
2268 |
ASTM D341 Single;Multi-curve ASTM;Matrix Referral |
2269 |
Current referred viscosity method is Multi-curve ASTM.
Select referred viscosity method: |
2270 |
Current referred viscosity method is Matrix Referral.
Select referred viscosity method: |
2271 |
Select the fluid: |
2272 |
Fluid 1;Fluid 2;Fluid 3;Fluid 4;Fluid 5;Fluid 6;Fluid 7;Fluid 8;End; |
2273 |
The ASTM Referral Matrix data was successfully written to the device. |
2274 |
Matrix Referral Results:
Fit Accuracy: %0
Fit Result: %1 |
2275 |
Click Next to begin viscosity scaling factor calibration. |
2276 |
A calibration is already in progress. Please complete the calibration before starting another one. |
2277 |
Before beginning calibration, the reference chamber must be purged. See the configuration manual for more information. |
2278 |
Select device measurement: |
2279 |
Specific Gravity;Molecular Weight;Relative Density; |
2280 |
Select calibration format (applies to all calibrations):
|
2281 |
2-point Calibration;3-point Calibration; |
2282 |
Select calibration number: |
2283 |
1;2;3;4; |
2284 |
Select the calibration option: |
2285 |
Calibration Gas Low;Calibration Gas High;Exit Gas Calibration; |
2286 |
Flow Calibration Gas Low through the meter, and click Next to start the calibration. |
2287 |
Flow Calibration Gas Low has been saved. Click Next to view the results. |
2288 |
The calibration is complete. Ensure that all required calibrations have been performed before beginning measurement:
• 2-point calibrations: Calibration Gas Low (K0) and Calibration Gas High (K2). |
2289 |
Flow Calibration Gas High through the meter and click Next to start the calibration. |
2290 |
Flow Calibration Gas High has been saved. Click Next to view the results. |
2291 |
Calibration Gas Low;Calibration Gas Medium;Calibration Gas High;Exit Gas Calibration; |
2292 |
The calibration is complete. Ensure that all required calibrations have been performed before beginning measurement:
• 3-point calibrations: Calibration Gas Low (K0), Calibration Gas Medium (K1), and Calibration Gas High (K2). |
2293 |
Flow Calibration Gas Medium through the meter and click Next to start the calibration. |
2294 |
Flow Calibration Gas Medium has been saved. Click Next to view the results. |
2295 |
Calibration Gas Low;Calibration Gas High; |
2296 |
Sensor Time Period: %0
Stability: %1
Calibration Point: %2 |
2297 |
Calibration Gas Low;Calibration Gas Medium;Calibration Gas High; |
2298 |
The calibration has been aborted. Ensure that all required calibrations have been performed before beginning measurement:
• 2-point calibrations: Calibration Gas Low (K0) and Calibration Gas High (K2). |
2299 |
Flow Calibration Gas Medium has been aborted. |
2300 |
Enter density of pure gas: |
2301 |
Enter density of second gas: |
2302 |
Resetting configuration changed flag. Please wait. |
2303 |
Configuration changed flag has been reset. |
2304 |
The calibration has been aborted. Ensure that all required calibrations have been performed before beginning measurement:
• 3-point calibrations: Calibration Gas Low (K0), Calibration Gas Medium (K1), and Calibration Gas High (K2). |
2305 |
The compressibility calculations is disabled. Do you want to enable it? |
2306 |
The compressibility calculations is enabled. |
2307 |
The compressibility calculations is enabled. Do you want to disable it? |
2308 |
The compressibility calculations is disabled. |
2309 |
Fit Accuracy:
%0
Fit Result:
%1 |
2310 |
Select Device Measurement: |
2311 |
Gas Specific Gravity;Relative Density;Molecular Weight |