0 |
TX200H |
1 |
UNITED_ELECTRIC |
2 |
PWM Saturated - Low |
3 |
PWM Saturated - High |
4 |
PWM Error - Measured value and commanded value do not match [+/-5%] |
5 |
Temperature Below -45C |
6 |
Temperature Above 90C |
7 |
A/D Va lue Saturated - Low |
8 |
A/D Va lue Saturated - High |
9 |
Vcore out of tolerance [+/- 5%] |
10 |
3.3V Supply out of tolerance [+/- 5%] |
11 |
5V Supply out of tolerance [+/- 5%] |
12 |
1.25V Reference Out of Tolerance [+/- 5%] |
13 |
Supply voltage fault VSUPPLY |
14 |
A/D Converter error - wrong # of bytes returned/Invalid data |
15 |
Cold Start Bit Set |
16 |
Pressure |
17 |
Temperature |
18 |
Simulate |
19 |
|
20 |
Update Period |
21 |
.3f |
22 |
Pres Dev fmly |
23 |
Pres Class |
24 |
Pres PDQ |
25 |
Pres LS |
26 |
Pres Fmly stat |
27 |
8.3f |
28 |
Pressure Update Period |
29 |
s |
30 |
7d |
31 |
Secondary Variable - Temperature Compensation |
32 |
Einh Snsr Temp |
33 |
Einheit Sensortemperatur- Temperatureinheit mit der die Sensortemperatur angezeigt wird. |
34 |
Temp Class |
35 |
Temp PDQ |
36 |
Temp LS |
37 |
Temp Dev Stat |
38 |
Temp USL |
39 |
Temp LSL |
40 |
Temp min span |
41 |
Temp Update Period |
42 |
Temp snsr s/n |
43 |
Pressure Value |
44 |
Temperature Value |
45 |
PV |
46 |
Rate |
47 |
Start |
48 |
Device Variables |
49 |
Primary Variables |
50 |
Identification |
51 |
PV Measurement |
52 |
Range Values |
53 |
Measured Values |
54 |
HART Tags |
55 |
Device Info. |
56 |
Revisions |
57 |
Diagnostics |
58 |
Status |
59 |
Byte 0 |
60 |
Byte 1 |
61 |
Byte 8 |
62 |
Byte 10 |
63 |
Byte 13 |
64 |
HART Status |
65 |
Ranging |
66 |
Temperature sensor |
67 |
Output Setup Menu |
68 |
HART Setup |
69 |
PV Analog Channel Saturated |
70 |
PV Analog Channel Fixed |
71 |
Wireless |
72 |
WirelessHART Device |
73 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
74 |
Primary |
75 |
Primary Analog Channel |
76 |
Secondary |
77 |
Secondary Analog Channel |
78 |
Tertiary |
79 |
Tertiary Analog Channel |
80 |
Quaternary |
81 |
Quaternary Analog Channel |
82 |
Quinary |
83 |
Quinary Analog Channel |
84 |
Critical Power Failure |
85 |
Lock all |
86 |
Trim not supported |
87 |
Configuration cannot be changed |
88 |
Locked by gateway |
89 |
Analog Input Channel |
90 |
Analog Output Channel |
91 |
Secondary Analog Channel Saturated |
92 |
Tertiary Analog Channel Saturated |
93 |
Quaternary Analog Channel Saturated |
94 |
Quinary Analog Channel Saturated |
95 |
Secondary Analog Channel Fixed |
96 |
Tertiary Analog Channel Fixed |
97 |
Quaternary Analog Channel Fixed |
98 |
Quinary Analog Channel Fixed |
99 |
Simulation active |
100 |
The device is in simulation mode and the level output is not representative of the process. |
101 |
Non-Volatile memory failure |
102 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
103 |
Volatile memory error |
104 |
The RAM memory check is invalid or maybe corrupt |
105 |
Watchdog reset executed |
106 |
A watchdog reset has been performed |
107 |
Voltage conditions out of range |
108 |
A voltage condition is outside its allowable range |
109 |
Environmental conditions out of range |
110 |
An internal or environmental condition is beyond acceptable limits. |
111 |
Electronic failure |
112 |
A hardware problem not related to the sensor has been detected. |
113 |
Subdevice list changed |
114 |
Duplicate master detected |
115 |
Capacity Denied |
116 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
117 |
Bandwidth allocation pending |
118 |
Block transfer pending |
119 |
Battery life |
120 |
Percent range |
121 |
Loop current |
122 |
Primary variable |
123 |
Secondary variable |
124 |
Tertiary variable |
125 |
Quaternary variable |
126 |
Read receive time |
127 |
Write date and time |
128 |
Non-Volatile Clock |
129 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
130 |
Clock Uninitialized |
131 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
132 |
None |
133 |
No restrictions |
134 |
SI Only |
135 |
Restricted to SI Unit codes only |
136 |
Process automation device |
137 |
Discrete device |
138 |
Hybrid:Process automation+discrete |
139 |
I/O System |
140 |
WirelessHART Process automation device |
141 |
WirelessHART Discrete device |
142 |
WirelessHART Process automation+discrete |
143 |
WirelessHART Gateway |
144 |
WirelessHART Access point |
145 |
WirelessHART Process adapter |
146 |
WirelessHART Discrete adapter |
147 |
WirelessHART enabled handheld/portable maintenance tool |
148 |
Device Profile |
149 |
US |
150 |
United States of America |
151 |
JP |
152 |
Japan |
153 |
DE |
154 |
Germany |
155 |
FR |
156 |
France |
157 |
ES |
158 |
Spain |
159 |
RU |
160 |
Russian Federation |
161 |
CN |
162 |
People's Republic of China |
163 |
Secondary Master |
164 |
Primary Master |
165 |
Device Diagnostic Status 0 |
166 |
Device Diagnostic Status 1 |
167 |
I/O and Subdevice Status |
168 |
WirelessHART Status |
169 |
Time Set Option |
170 |
Real Time Clock Flags |
171 |
Country |
172 |
Country of intended device installation |
173 |
SI Unit Control |
174 |
Indicates if device should only display SI unit codes |
175 |
8u |
176 |
Dynamic Variables Returned for Device Variables |
177 |
Command Response Truncated |
178 |
Configuration Change Counter Mismatch |
179 |
Status bytes mismatch |
180 |
Device Variable |
181 |
Output Status |
182 |
Output Status Byte |
183 |
Uninitialized |
184 |
Normal - one sensor |
185 |
Normal - two sensors |
186 |
Normal - three sensors |
187 |
Shorted |
188 |
Shorted- shorted field wiring. |
189 |
Limited high |
190 |
Limited high- block can not respond to upward movement of value. |
191 |
Open |
192 |
Open- open field wiring. |
193 |
Limited low |
194 |
Limited low- block can not respond to upward movement of target. |
195 |
Initialization done |
196 |
Initialization done- block initialization is complete. |
197 |
Not selected |
198 |
Not selected- target value provided by block was not selected. |
199 |
No path |
200 |
No path- there is no path to the process. |
201 |
Fail |
202 |
Fail- hardware failure. |
203 |
Bad |
204 |
Bad- block is out of service or has bad input. |
205 |
Not from process |
206 |
Not from process- value is not directly affected by any process information. |
207 |
Local override |
208 |
Local override- block has LO mode. |
209 |
Not Cascade Mode |
210 |
Not RCas Mode |
211 |
Not ROut Mode |
212 |
No communication |
213 |
No communication- (NO_COM). |
214 |
uint16_var_blk_char |
215 |
uint32_var_blk_char |
216 |
uint8_var_blk_char |
217 |
int8_var_blk_char |
218 |
new itest redef1 label |