0 |
6u |
1 |
FMR2xx |
2 |
|
3 |
electronics defect A111 |
4 |
electronics defect A113 |
5 |
electronics defect A114 |
6 |
electronics defect A164 |
7 |
checksum error general reset & new calibr.required A102 |
8 |
electronics defect A121 |
9 |
electronics defect A171 |
10 |
checksum error general reset & new calibr.required A110 |
11 |
checksum error general reset & new calibr.required A160 |
12 |
electronics defect A155 |
13 |
sensor 1 defect check connection A231 |
14 |
electronics defect A115 |
15 |
electronics defect amplifier A272 |
16 |
downloading please wait A106 |
17 |
download error repeat download A116 |
18 |
no usable echo channel 1 check calibr. E641 |
19 |
linearisation ch1 not complete, not usable A671 |
20 |
level in safety distance - risk of overspill E651 |
21 |
custody switch undef check position A270 |
22 |
reduced accuracy check installation W271 |
23 |
reduced accuracy W273 |
24 |
reduced accuracy check pipe diameter W274 |
25 |
electronics defect factory settings W275 |
26 |
current ch1 out of range W681 |
27 |
recording of mapping please wait W512 |
28 |
no factory cali- bration ch1 W511 |
29 |
less than 2 linea- risation points for channel 1 W611 |
30 |
linearisation ch1 curve not monotone W601 |
31 |
simulation ch. 1 on W621 |
32 |
initialising - please wait W103 |
33 |
initialising - please wait W153 |
34 |
PERCENT RANGE |
35 |
% |
36 |
6.1f |
37 |
horizonzal zyl. |
38 |
manual |
39 |
semi automatic |
40 |
activate table |
41 |
clear table |
42 |
MIN. SPAN |
43 |
LOW SENSOR LIMIT |
44 |
.4f |
45 |
HIGH SENSOR LIMIT |
46 |
SERIAL-NO. SENS. |
47 |
7u |
48 |
FORMAT |
49 |
01u |
50 |
NUL_UNIT |
51 |
none |
52 |
DUMMY_FLOAT |
53 |
measured value |
54 |
9.0f |
55 |
9.1f |
56 |
9.2f |
57 |
9.3f |
58 |
no.of decimals |
59 |
x |
60 |
x.x |
61 |
x.xx |
62 |
x.xxx |
63 |
unlock parameter |
64 |
5.1u |
65 |
format display |
66 |
decimal |
67 |
ft-in-1/16'' |
68 |
tank shape |
69 |
dome ceiling |
70 |
horizontal cyl |
71 |
bypass |
72 |
stilling well |
73 |
flat ceiling |
74 |
sphere |
75 |
overspill prot. |
76 |
standard |
77 |
german WHG |
78 |
medium property |
79 |
unknown |
80 |
DC: < 1.9 |
81 |
DC: 1.9 ... 4 |
82 |
DC: 4 ... 10 |
83 |
DC: > 10 |
84 |
process cond. |
85 |
calm surface |
86 |
turb. surface |
87 |
add. agitator |
88 |
fast change |
89 |
test:no filter |
90 |
distance process
conn. to min. level |
91 |
empty calibr. |
92 |
distance unit |
93 |
m |
94 |
ft |
95 |
mm |
96 |
inch |
97 |
max. meas. dist. |
98 |
span |
99 |
full calibr. |
100 |
inner diameter of
bypass/stilling well |
101 |
pipe diameter |
102 |
output on alarm |
103 |
MIN (<=3.6mA) |
104 |
MAX (22mA) |
105 |
hold |
106 |
user specific |
107 |
output on alarm |
108 |
mA |
109 |
4.2f |
110 |
wire type |
111 |
2 wire |
112 |
4 wire |
113 |
outp. echo loss |
114 |
alarm |
115 |
ramp %/min |
116 |
ramp %span/min |
117 |
%/min |
118 |
10.3f |
119 |
in case of echo loss
max. 4000 sec. |
120 |
delay time |
121 |
s |
122 |
4.1u |
123 |
level/ullage |
124 |
level CU |
125 |
level DU |
126 |
ullage CU |
127 |
ullage DU |
128 |
linearisation |
129 |
linear |
130 |
manual |
131 |
semi-automatic |
132 |
table on |
133 |
clear table |
134 |
from antenna tip/
lower edge of horn |
135 |
safety distance |
136 |
in safety dist. |
137 |
warning |
138 |
self holding |
139 |
ackn. alarm |
140 |
no |
141 |
yes |
142 |
customer unit |
143 |
l |
144 |
hl |
145 |
m3 |
146 |
dm3 |
147 |
cm3 |
148 |
ft3 |
149 |
usgal |
150 |
i gal |
151 |
kg |
152 |
t |
153 |
lb |
154 |
ton |
155 |
 j |
156 |
table no. |
157 |
2.1u |
158 |
input level |
159 |
input volume |
160 |
max. scale |
161 |
diameter vessel |
162 |
8.3f |
163 |
check distance |
164 |
distance = ok |
165 |
dist.too small |
166 |
dist. too big |
167 |
dist. unknown |
168 |
input of
mapping range |
169 |
range of mapping |
170 |
start mapping |
171 |
pres. map dist. |
172 |
cust. tank map |
173 |
inactive |
174 |
active |
175 |
reset |
176 |
echo quality |
177 |
dB |
178 |
3.1u |
179 |
will be added to the
measured level |
180 |
offset |
181 |
output damping |
182 |
5.1f |
183 |
blocking dist. |
184 |
commun. address |
185 |
2.1i |
186 |
no. of preambels |
187 |
low output limit |
188 |
off |
189 |
on |
190 |
curr.output mode |
191 |
curr.turn down |
192 |
fixed current |
193 |
fixed cur. value |
194 |
5.2f |
195 |
simulation |
196 |
sim. off |
197 |
sim. level |
198 |
sim. volume |
199 |
sim. current |
200 |
simulation value |
201 |
output current |
202 |
language |
203 |
English |
204 |
Deutsch |
205 |
Français |
206 |
Español |
207 |
Italiano |
208 |
Nederlands |
209 |
???? |
210 |
back to home |
211 |
4.1i |
212 |
sep. character |
213 |
. |
214 |
, |
215 |
pres. amplitude |
216 |
3.1i |
217 |
ampl. over map. |
218 |
ampl. over FAC |
219 |
unfilt. distance |
220 |
8.2f |
221 |
pres. edge param |
222 |
present FEF |
223 |
device name |
224 |
order code |
225 |
MAM filt. length |
226 |
MAM filt. border |
227 |
low pass filter |
228 |
hysterese width |
229 |
max. fill. speed |
230 |
inch/s |
231 |
mm/s |
232 |
max. drain speed |
233 |
unfilt. dist.raw |
234 |
env. statistics |
235 |
env. smoothing |
236 |
envel. energy |
237 |
mapping adder |
238 |
mapping scanrate |
239 |
mapping average |
240 |
fact. map. valid |
241 |
FAC adder |
242 |
FAC scan rate |
243 |
range fact. map |
244 |
not active |
245 |
active |
246 |
cust. map. valid |
247 |
edge detec. mode |
248 |
front of echo |
249 |
behind echo |
250 |
middle of echo |
251 |
edge parameter |
252 |
first echo fact. |
253 |
FEF threshold |
254 |
FEF at near dist |
255 |
FEF dist. near |
256 |
FEF dist. far |
257 |
FEF edge |
258 |
max.ampl.o.FAC |
259 |
Tank Bottom Det. |
260 |
min. ampl. TBD |
261 |
max. level TBD |
262 |
fill/drain speed |
263 |
HF module |
264 |
µP II.1 |
265 |
µP II.2 |
266 |
µP III.1 |
267 |
µP III.2 |
268 |
zero distance |
269 |
antenna type |
270 |
rod aerial |
271 |
horn aerial |
272 |
dish antenna |
273 |
planar antenna |
274 |
inactive length |
275 |
total length |
276 |
pos. ref. pulse |
277 |
ampl. ref. pulse |
278 |
Plausib.Function |
279 |
PF maximum delta |
280 |
PF upper area |
281 |
display version |
282 |
calc. cycle time |
283 |
ms |
284 |
sw. vers. appl. |
285 |
para. vers. appl |
286 |
D/A adjust 4mA |
287 |
D/A adjust 20mA |
288 |
debug index 1 |
289 |
debug index 2 |
290 |
debug value |
291 |
10.1f |
292 |
err. state alg. |
293 |
state algorithm |
294 |
present error |
295 |
|
296 |
A101
checksum error
general reset & new
calibr.required |
297 |
A102
checksum error
general reset & new
calibr.required |
298 |
W103
initialising -
please wait
|
299 |
A106
downloading
please wait
|
300 |
A110
checksum error
general reset & new
calibr.required |
301 |
A111
electronics defect
|
302 |
A113
electronics defect
|
303 |
A114
electronics defect
|
304 |
A115
fault power supply
|
305 |
A116
download error
repeat download
|
306 |
A121
electronics defect
|
307 |
A152
checksum error
general reset & new
calibr.required |
308 |
W153
initialising -
please wait
|
309 |
A155
electronics defect
|
310 |
A160
checksum error
general reset & new
calibr.required |
311 |
A164
electronics defect
|
312 |
A171
electronics defect
|
313 |
A231
sensor 1 defect
check connection
|
314 |
A270
custody switch undef
check position
|
315 |
W271
reduced accuracy
check installation
|
316 |
A272
electronics defect
amplifier
|
317 |
W273
reduced accuracy
|
318 |
W274
reduced accuracy
check pipe diameter
|
319 |
W275
electronics defect
factory settings
|
320 |
W511
no factory cali-
bration ch1
|
321 |
W512
recording of mapping
please wait
|
322 |
W601
linearisation ch1
curve not monotone
|
323 |
W611
less than 2 linea-
risation points
for channel 1 |
324 |
W621
simulation ch. 1
on
|
325 |
E641
no usable echo
channel 1
check calibr. |
326 |
E643
no usable echo
channel 1
check calibr. |
327 |
E651
level in safety
distance - risk
of overspill |
328 |
A671
linearisation ch1
not complete,
not usable |
329 |
W681
current ch1 out
of range
|
330 |
previous error |
331 |
clear last error |
332 |
keep |
333 |
erase |
334 |
for reset code
see manual |
335 |
reset |
336 |
measured dist. |
337 |
6.3f |
338 |
measured level |
339 |
application par. |
340 |
not modified |
341 |
modified |
342 |
tag no. |
343 |
protocol+sw-no. |
344 |
serial no. |
345 |
download mode |
346 |
parameter only |
347 |
param+cust.map |
348 |
mapping only |
349 |
4mA value |
350 |
20mA value |
351 |
max sample dist. |
352 |
length FAR10 -
for FMR230 only |
353 |
antenna extens. |
354 |
MEAS_VAL_UNIT |
355 |
VARIABLE_ARRAY_INDEX |
356 |
VARIABLE_ARRAY_VALUE |
357 |
VARIABLE_ARRAY_UNIT |
358 |
FORMAT_TABLE_INDEX |
359 |
FORMAT_TABLE |
360 |
%-RANGE |
361 |
Set to nearest possible value |
362 |
Invalid value selection |
363 |
basic setup |
364 |
safety settings |
365 |
linearisation |
366 |
extended calibr. |
367 |
output |
368 |
display |
369 |
service |
370 |
info |
371 |
distance |
372 |
envelope |
373 |
mapping |
374 |
edge |
375 |
first echo |
376 |
tank bottom |
377 |
module |
378 |
antenna table |
379 |
system 1 |
380 |
system 2 |
381 |
debug |
382 |
diagnostics |
383 |
system parameters |
384 |
MATRIX GROUP SEL. |
385 |
DEVICE DATA |
386 |
HART OUTPUT |
387 |
UPLOAD |
388 |
Reset Configuration Changed Flag |
389 |
Restart Device |
390 |
FMU860 |
391 |
DT |
392 |
DVC |
393 |
FCX |
394 |
AS800 |
395 |
340T |
396 |
340C |
397 |
344 |
398 |
Siemens Micro K |
399 |
SITRANS L |
400 |
SIPAN PH |
401 |
SITRANS P |
402 |
75SMT |
403 |
UNIDELTA |
404 |
8320 |
405 |
SX5100 |
406 |
TSV175 |
407 |
DMU |
408 |
3780 |
409 |
BM70 |
410 |
8u |
411 |
Do you want to reset the 'Config Changed Flag' ? |
412 |
Attention ! Device will be restarted !