0 |
AT9000 |
1 |
Azbil Corporation |
2 |
|
3 |
Analog/Digital Conversion Failure |
4 |
Analog/Digital Conversion Failure- An error occurred during analog-to-digital conversion. |
5 |
Sensor Characteristic Data Failure |
6 |
Sensor Characteristic Data Failure- An error was detected in the sensor characteristic data. |
7 |
Suspect Input |
8 |
Suspect Input- Input data error |
9 |
CPU Failure |
10 |
CPU Failure- CPU operation failure |
11 |
NVM Failure |
12 |
NVM Failure- Nonvolatile memory error |
13 |
RAM Failure |
14 |
RAM Failure- RAM failure |
15 |
ROM Failure |
16 |
ROM Failure- ROM failure |
17 |
Output Circuit Failure |
18 |
Output Circuit Failure- Output circuit failure |
19 |
Meter Body Over Temperature |
20 |
Meter Body Over Temperature- Meter body is too hot. |
21 |
Excess Zero Calibration Value |
22 |
Excess Zero Calibration Value- The zero Calibration Value exceeds the limit for accurate measurement. |
23 |
Excess Span Calibration Value |
24 |
Excess Span Calibration Value- The span Calibration Value exceeds the limit for accurate measurement. |
25 |
In Output Mode |
26 |
In Output Mode- The device is operating in Output Mode. |
27 |
Meter Body Overload or Failure |
28 |
Meter Body Overload or Fault-
-The input pressure is more than two times the upper range limit for the device. Or,
-Device error |
29 |
Fld dev stat 1-6- Unknown error |
30 |
Not Calibrated |
31 |
Not Calibrated- Clears the calibration data. |
32 |
Fld dev stat 1-8- Unknown error |
33 |
External Zero/Span Adjustment Failure |
34 |
External Zero/Span Adjustment Failure- An error occurred during external adjustment of the zero point or span. |
35 |
Fld dev stat 2-2- Unknown error |
36 |
Fld dev stat 2-3- Unknown error |
37 |
Fld dev stat 2-4- Unknown error |
38 |
Failure Alarm Simulation Mode |
39 |
Failure Alarm Simulation Mode- The device is operating in Failure Alarm Simulation Mode. |
40 |
Contact Output Simulation Mode |
41 |
Contact Output Simulation Mode- The device is operating in Contact Output Simulation Mode. |
42 |
Output Alarm Detected |
43 |
Output Alarm Detected- The output is going over upper/lower limit of output alarm. |
44 |
Sensor Temp. Alarm Detected |
45 |
Sensor Temp. Alarm Detected- The sensor temperature is going over upper/lower limit of sensor temp. alarm. |
46 |
Invalid Database |
47 |
Invalid Database- Configuration data and/or calibration data is invalid. |
48 |
Fld dev stat 3-2- Unknown error |
49 |
Fld dev stat 3-3- Unknown error |
50 |
Fld dev stat 3-4- Unknown error |
51 |
Fld dev stat 3-5- Unknown error |
52 |
Fld dev stat 3-6- Unknown error |
53 |
Fld dev stat 3-7- Unknown error |
54 |
Fld dev stat 3-8- Unknown error |
55 |
Low Output Alarm |
56 |
Low Output Alarm- Indicates whether the present output level is lower than the Output Alarm Low Limit setting. |
57 |
High Output Alarm |
58 |
High Output Alarm- Indicates whether the present output level is higher than the Output Alarm High Limit setting. |
59 |
Low Sensor Temp. Alarm |
60 |
Low Sensor Temp. Alarm- Indicates whether the present sensor temperature is lower than the Sensor Temp. Alarm Low Limit setting. |
61 |
High Sensor Temp. Alarm |
62 |
High Sensor Temp. Alarm- Indicates whether the present sensor temperature is higher than the Sensor Temp. Alarm High Limit setting. |
63 |
Fld dev stat 4-5- Unknown error |
64 |
Fld dev stat 4-6- Unknown error |
65 |
Fld dev stat 4-7- Unknown error |
66 |
Fld dev stat 4-8- Unknown error |
67 |
Pressure Frequency Index Alarm |
68 |
Pressure Frequency Index Alarm - Check Process and/or Impulse Line. |
69 |
Standard Deviation Alarm |
70 |
Standard Deviation Alarm - Check Process Conditions. |
71 |
Out-of-Range Count Alarm |
72 |
Out-of-Range Count Alarm - Check Operating Conditions. |
73 |
Fld dev stat 5-4- Unknown error |
74 |
Fld dev stat 5-5- Unknown error |
75 |
Fld dev stat 5-6- Unknown error |
76 |
Fld dev stat 5-7- Unknown error |
77 |
Fld dev stat 5-8- Unknown error |
78 |
.3f |
79 |
ExtSw mode |
80 |
OnePush |
81 |
RaiseLower |
82 |
SI Unit |
83 |
ALL Unit |
84 |
SI Only |
85 |
Alarm Status |
86 |
Alarm Status- Displays the current alarm output status. |
87 |
Output Alarm Low |
88 |
Output Alarm Low- Indicates whether or not an alarm is detected.
If an alarm is detected, the present output level is lower than the Lower Output Alarm setting. |
89 |
Output Alarm High |
90 |
Output Alarm High- Indicates whether or not an alarm is detected.
If an alarm is detected, the present output level is higher than the Upper Output Alarm setting. |
91 |
Sensor Temp. Alarm Low |
92 |
Sensor Temp. Alarm Low- Indicates whether or not an alarm is detected.
If an alarm is detected, the present sensor temperature is lower than the Lower Sensor Temp. Alarm setting. |
93 |
Sensor Temp. Alarm High |
94 |
Sensor Temp. Alarm High- Indicates whether or not an alarm is detected.
If an alarm is detected, the present sensor temperature is higher than the Upper Sensor Temp. Alarm setting. |
95 |
Contact Output Status |
96 |
Contact Output Status- Displays the status of the contact output.
--Open: The contact output circuit is open.
--Closed: The contact output circuit is closed.
--Sim Open: The contact output circuit is open in simulation mode.
--Sim Closed: The contact output circuit is closed in simulation mode. |
97 |
Open |
98 |
Open- The contact output circuit will open when an alarm occurs. |
99 |
Closed |
100 |
Closed- The contact output circuit will close when an alarm occurs. |
101 |
Sim Open |
102 |
Sim Open- Device will go into simulation mode and the contact output circuit will open. |
103 |
Sim Closed |
104 |
Sim Closed- Device will go into simulation mode and the contact output circuit will close. |
105 |
Output Alarm |
106 |
Output Alarm- Configures the output alarm.
--No Alarm: Output alarm is not used.
--Low Limit: If the output is below the low limit for the output alarm, the alarm is activated.
--High Limit: If the output is above the high limit for the output alarm, the alarm is activated.
--Low/High Limit: If the output is either below the low limit or above the high limit for the output alarm, the alarm is activated. |
107 |
No Alarm |
108 |
No Alarm- Output alarm is not used. |
109 |
Low Limit |
110 |
Low Limit- If the output is below the low limit for the output alarm, the alarm is activated. |
111 |
High Limit |
112 |
High Limit- If the output is above the high limit for the output alarm, the alarm is activated. |
113 |
Low/High Limit |
114 |
Low/High Limit- If the output is either below the low limit or above the high limit for the output alarm, the alarm is activated. |
115 |
Output Alarm Low Limit |
116 |
Output Alarm Low Limit- Sets the low limit for the Output Alarm. |
117 |
.1f |
118 |
6.1f |
119 |
Output Alarm High Limit |
120 |
Output Alarm High Limit- Sets the high limit for the Output Alarm. |
121 |
Sensor Temp. Alarm |
122 |
Sensor Temp. Alarm- Configures the temperature alarm.
--No Alarm: Sensor temperature alarm is not used.
--Low Limit: If the sensor temperature is below the low limit for the sensor temperature alarm, the alarm is activated.
--High Limit: If the sensor temperature is above the high limit for the sensor temperature alarm, the alarm is activated.
--Low/High Limit: If the sensor temperature is either below the low limit or above the high limit for the sensor temperature alarm, the alarm is activated. |
123 |
No Alarm- Sensor temperature alarm is not used. |
124 |
Low Limit- If the sensor temperature is below the low limit for the sensor temperature alarm, the alarm is activated. |
125 |
High Limit- If the sensor temperature is above the high limit for the sensor temperature alarm, the alarm is activated. |
126 |
Low/High Limit- If the sensor temperature is either below the low limit or above the high limit for the sensor temperature alarm, the alarm is activated. |
127 |
Sensor Temp. Alarm Low Limit |
128 |
Sensor Temp. Alarm Low Limit- Sets the low limit for the Sensor Temp. Alarm. |
129 |
Sensor Temp. Alarm High Limit |
130 |
Sensor Temp. Alarm High Limit- Sets the high limit for the Sensor Temp. Alarm. |
131 |
Contact Output Mode |
132 |
Contact Output Mode- Sets the contact output mode to either Normally Open or Normally Closed.
--Normally Open: Contacts are normally open, but close when an alarm is detected.
--Normally Closed: Contacts are normally closed, but open when an alarm is detected. |
133 |
Normally Open |
134 |
Normally Open- Sets the contact output circuit to normally open, and when the alarm is detected, becomes closed. |
135 |
Normally Closed |
136 |
Normally Closed- Sets the contact output circuit to normally closed, and when the alarm is detected, becomes open. |
137 |
Contact Output ON/OFF |
138 |
Contact Output ON/OFF- Selects the alarms to reflect the contact output. |
139 |
Output Low |
140 |
Output Low- This parameter determines whether or not detection of the output low-limit alarm is connected to the contact output circuit. |
141 |
Output High |
142 |
Output High- This parameter determines whether or not detection of the output high-limit alarm is connected to the contact output circuit. |
143 |
Sensor Temp. Low |
144 |
Sensor Temp. Low- This parameter determines whether or not detection of the sensor's low-limit temperature alarm is connected to the contact output circuit. |
145 |
Sensor Temp. High |
146 |
Sensor Temp. High- This parameter determines whether or not detection of the sensor's high-limit temperature alarm is connected to the contact output circuit. |
147 |
Off |
148 |
On |
149 |
alarm_select_dummy |
150 |
Display Mode |
151 |
Display Mode- Sets the device indicator's method of display.
--%: Displays the currently measured pressure as a percentage of the preset measurement range.
--Pressure: Displays the currently measured pressure in the selected pressure unit.
--Scale: Displays the measured value (flow rate, level etc.) in user-selected or user-defined units. |
152 |
% |
153 |
%- Displays the currently meauserd pressure as a percentage of the preset measuring range. |
154 |
Pressure |
155 |
Pressure- Displays the currently measured pressure in the selected engineering units. |
156 |
Scale |
157 |
Scale- Displays the measured value in user-selected or user-defined units (flow rate, level etc). |
158 |
Display Function |
159 |
Display Function- Sets the type of analog output from the device.
--Linear: Displays PV by linear scale.
--Square root: Displays flow by square root extraction. |
160 |
Linear |
161 |
Linear- Displays PV by linear scale. |
162 |
Square root |
163 |
Square root- Displays flow by square root extraction. |
164 |
Disp. Unit |
165 |
Disp. Unit- Selects the engineering unit displayed on the indicator of the device. Valid when Scale is selected for Display Mode. |
166 |
user define unit |
167 |
mmH2O |
168 |
mmAq |
169 |
mH2O |
170 |
inH2O |
171 |
ftH2O |
172 |
kPa |
173 |
MPa |
174 |
Pa |
175 |
hPa |
176 |
kPaG |
177 |
MPaG |
178 |
kPa abs |
179 |
MPa abs |
180 |
Pa abs |
181 |
hPa abs |
182 |
bar |
183 |
mbar |
184 |
barG |
185 |
mbarG |
186 |
mmHg |
187 |
inHg |
188 |
mmHg abs |
189 |
inHg abs |
190 |
gf/cm2 |
191 |
kgf/cm2 |
192 |
g/cm2 |
193 |
kg/cm2 |
194 |
kgf/cm2G |
195 |
kgf/cm2 abs |
196 |
atm |
197 |
Torr |
198 |
psi |
199 |
g/cm3 |
200 |
kg/m3 |
201 |
m3 |
202 |
l |
203 |
kl |
204 |
ml/h |
205 |
l/h |
206 |
kl/h |
207 |
Skl/h |
208 |
Sm3/h |
209 |
t/h |
210 |
m3/h |
211 |
km3/h |
212 |
Nm3/h |
213 |
kNm3/h |
214 |
l/min |
215 |
kl/min |
216 |
m3/min |
217 |
Nml/min |
218 |
Nl/min |
219 |
Nm3/min |
220 |
kl/d |
221 |
m3/d |
222 |
t/d |
223 |
Nm3/d |
224 |
kg/h |
225 |
gal/min |
226 |
gal/h |
227 |
mm |
228 |
m |
229 |
t |
230 |
kg |
231 |
none |
232 |
Exponent |
233 |
Exponent- Selects the exponent (X10, X100, etc.) displayed on the indicator of the device.
Valid when Scale is selected for Display Mode. |
234 |
X1 |
235 |
X10 |
236 |
X100 |
237 |
X1000 |
238 |
Cutoff Mode |
239 |
Cutoff Mode- Displays one of the following Cutoff Mode status indications.
--Def (7.1%Lin): The default Cutoff Mode is in effect. (Flow mode is the default value, dropout is linear, and cutoff point is 7.1%)
--Zero Flow & Dflt: Dropout is zero and flow mode is the default value.
--Lin. Flow & Dflt: Dropout is linear and flow mode is the default value
--Zero Flow & Bi-dir: Dropout is zero and flow mode is bi-directional.
--Lin. Flow & Bi-dir: Dropout is linear and flow mode is bi-directional.
--Dropout: Determines behavior below the cutoff value (either zero or linear).
--Flow mode: Determines the method of calculating the flow output, either default (square root of positive output) or bi-directional (square root of both positive and negative outputs). |
240 |
Zero Flow:Def. |
241 |
Zero Flow:Def.- Dropout is zero, and flow mode is default. |
242 |
Lin. Flow:Def. |
243 |
Lin. Flow:Def.- Dropout is linear, and flow mode is default. |
244 |
Zero Flow:Bi-dir. |
245 |
Zero Flow:Bi-dir.- Dropout is zero, and flow mode is bi-directional. |
246 |
Lin. Flow:Bi-dir. |
247 |
Lin. Flow:Bi-dir.- Dropout is linear, and flow mode is bi-directional. |
248 |
Def. (7.1% Lin.) |
249 |
Def. (7.1% Lin.)- Cutoff mode is default. (Flow mode is default, dropout is linear, and cutoff point is 7.1%). |
250 |
Cutoff Point |
251 |
Cutoff Point- Displays the cutoff point (%) that is valid when Square Root is selected as the Xfer fnctn. |
252 |
.2f |
253 |
5.2f |
254 |
EULO (0%) |
255 |
EULO (0%)- The value displayed on the indicator of the device when the output from the device is 0%. Valid when Scale is selected for Display Mode. |
256 |
7.3f |
257 |
EUHI (100%) |
258 |
EUHI (100%)- The value displayed on the indicator of the device when the output from the device is 100%. Valid when Scale is selected for Display Mode. |
259 |
Output Low Limit |
260 |
Output Low Limit- Sets the low limit for the Loop current. |
261 |
6.3f |
262 |
Output High Limit |
263 |
Output High Limit- Sets the high limit for the Loop current. |
264 |
Measurement Type |
265 |
Measurement Type- Displays the device's sensor type.
--DP: Differential pressure
--GP: Gauge pressure
--AP: Absolute pressure |
266 |
DP |
267 |
GP |
268 |
AP |
269 |
Reserved |
270 |
PROM No. |
271 |
PROM No.- Displays PROM No. used for the device. |
272 |
SFN Communicator |
273 |
Enable |
274 |
Disable |
275 |
User Unit |
276 |
User Unit- Sets the user-defined unit displayed on the indicator of the device. Valid when Scale is selected for Display Mode. |
277 |
Status Record 0 |
278 |
Status Record 1 |
279 |
Status Record 2 |
280 |
Status Record 3 |
281 |
1 (Latest) |
282 |
1 (Latest)- Displays the zero corrected value as a percentage of the Upper Range Limit. |
283 |
% of URL |
284 |
7.4f |
285 |
2 |
286 |
2- Displays the zero corrected value as a percentage of the Upper Range Limit. |
287 |
3 |
288 |
3- Displays the zero corrected value as a percentage of the Upper Range Limit. |
289 |
4 |
290 |
4- Displays the zero corrected value as a percentage of the Upper Range Limit. |
291 |
5 |
292 |
5- Displays the zero corrected value as a percentage of the Upper Range Limit. |
293 |
Zero Correct Current |
294 |
Count |
295 |
Software Write Protect |
296 |
Software Write Protect- Displays the status of software write protection. |
297 |
No |
298 |
Yes |
299 |
Hardware Write Protect |
300 |
Hardware Write Protect- Displays the status of hardware write protection. |
301 |
Analog Output Configuration |
302 |
Normal |
303 |
NAMUR |
304 |
Hardware Equipment |
305 |
Meter |
306 |
Contact Output |
307 |
Software Version |
308 |
Height |
309 |
Height- Compensates for the effects of temperature change on the fill fluid of the remote sealed transmitter. User must enter the distance between the flanges in order to use this compensation. |
310 |
Set Output Value |
311 |
6.2f |
312 |
Burnout Simulation |
313 |
Clear |
314 |
Burnout Simulaiton |
315 |
Production Number |
316 |
Production Number- Displays the pressure transmitter Production Number. |
317 |
Pressure Frequency Index |
318 |
Pressure Frequency Index- The value indicating Pressure Frequency index. |
319 |
Press Freq Index Min |
320 |
Press Freq Index Min- Displays the minimum value of Pressure Frequency Index during a certain period. |
321 |
Press Freq Index Max |
322 |
Press Freq Index Max- Displays the maximum value of Pressure Frequency Index during a certain period. |
323 |
Press Freq Index Sensor Selection |
324 |
Press Freq Index Sensor Selection- Displays the sensor (Differential Pressure Sensor or Static Pressure Sensor) used to calculate the Pressure Frequency Index and the sampling period of the sensor if the device is a Differential Pressure Transmitter.
Also, displays the sampling period of the Differential Pressure Sensor if the device is a Gauge Pressure Transmitter or Absolute Pressure Transmitter. |
325 |
DP, 120ms |
326 |
DP, 240ms |
327 |
DP, 360ms |
328 |
SP, 360ms |
329 |
Press Freq Calc PV Low Limit |
330 |
Press Freq Calc PV Low Limit- The Pressure Frequency Index is calculated only when the pressure is greater than or equal to this low limit. |
331 |
.4f |
332 |
Press Freq Calc PV High Limit |
333 |
Press Freq Calc PV High Limit- The Pressure Frequency Index is calculated only when the pressure is less than or equal to this high limit. |
334 |
Press Freq Filter Constant |
335 |
Press Freq Filter Constant- Sets the filter constant used to calculate the Pressure Frequency Index.
The larger the index value, the fewer the detection errors, because low-frequency components are more thoroughly removed. However, the amount of variation in the index also becomes smaller, requiring a more strict setting of the alarm. |
336 |
Press Freq Index Low Limit |
337 |
Press Freq Index Low Limit- Sets the low limit for the Pressure Frequency Index.
If Press Freq Index Alarm Use is set to "Enabled (Low)" or "Enabled (High and Low)," and the Pressure Frequency Index is less than this limit, the alarm will turn ON. |
338 |
Press Freq Index High Limit |
339 |
Press Freq Index High Limit- Sets the high limit for the Pressure Frequency Index.
If Press Freq Index Alarm Use is set to "Enabled (High)" or "Enabled (High and Low)" and the Pressure Frequency Index is greater than this limit, the alarm will turn ON. |
340 |
Press Freq Index Alarm Use |
341 |
Press Freq Index Alarm Use- Sets whether the Pressure Frequency Alarm is activated, for example when the Pressure Frequency Index is greater than Press Freq High Limit, or when the Pressure Frequency Index is less than Press Freq Low Limit. |
342 |
Disabled |
343 |
Enabled (High) |
344 |
Enabled (Low) |
345 |
Enabled (High and Low) |
346 |
OOR Pressure Count |
347 |
OOR Pressure Count- Displays the number of occurrences of out-of-range pressure. This value can be reset. |
348 |
Counts |
349 |
6u |
350 |
Normal Pressure Low Limit |
351 |
Normal Pressure Low Limit- This is the low limit for normal pressure. After calbration, if the pressure drops below this limit, OOR Pressure Count will increase by 1. |
352 |
Normal Pressure High Limit |
353 |
Normal Pressure High Limit- This is the high limit for normal pressure. After calbration, if the pressure exceeds this limit, OOR Pressure Count will increase by 1. |
354 |
OOR Count Alarm Threshold |
355 |
OOR Count Alarm Threshold- Sets the threshold for the Out-of-Range Count Alarm.
If OOR Count Alarm Use is set to "Enabled" and the OOR Pressure Count is greater than or equal to this threshold, the alarm will turn ON. |
356 |
OOR Count Alarm Use |
357 |
OOR Count Alarm Use- Sets whether the Out-of-Range Count Alarm is activated, when OOR Pressure Count is greater than or equal to OOR Count Alarm Threshold. |
358 |
Enabled |
359 |
Average Pressure |
360 |
Average Pressure- Displays the average value for the measured pressure. |
361 |
Standard Deviation |
362 |
Standard Deviation- Displays the standard deviation for the pressure. |
363 |
Standard Deviation Max |
364 |
Standard Deviation Max- Displays the maximum standard deviation for the pressure in certain period. |
365 |
Standard Deviation Min |
366 |
Standard Deviation Min- Displays the minimum standard deviation for the pressure in certain period. |
367 |
Standard Deviation Unit |
368 |
Standard Deviation Unit- Selects the unit of measurement for the Standard Deviation and the Average Pressure. |
369 |
mmH2O@68degF |
370 |
inH2O@68degF |
371 |
Standard Deviation Sample Count |
372 |
Standard Deviation Sample Count- Displays the number of samples used to calculate the Standard Deviation and the Average Pressure. Sampling is done once every 60 ms. |
373 |
Samples |
374 |
5u |
375 |
Standard Deviation Low Limit |
376 |
Standard Deviation Low Limit- Sets the low limit for the Standard Deviation.
If Standard Deviation Alarm Use is set to "Enabled (Low)" or "Enabled (High and Low)," and the Standard Deviation is less than this limit, the alarm will turn ON. |
377 |
Standard Deviation High Limit |
378 |
Standard Deviation High Limit- Sets the high limit for the Standard Deviation.
If Standard Deviation Alarm Use is set to "Enabled (High)" or "Enabled (High and Low)," and the Standard Deviation is greater than this limit, the alarm will turn ON. |
379 |
Standard Deviation Alarm Use |
380 |
Standard Deviation Alarm Use- Sets whether the Standard Deviation Alarm is activated, for example when the Standard Deviation is greater than Standard Deviation High Limit, or when the Standard Deviation is less than Standard Deviation Low Limit. |
381 |
Advance Diag Option |
382 |
OOR Pressure Count Ratio |
383 |
Status Record 4 |
384 |
Status Record 5 |
385 |
dss_1_7_dummy |
386 |
dss_2_dummy |
387 |
dss_3_dummy |
388 |
dss_4_dummy |
389 |
dss_rec_1_7_dummy |
390 |
dss_rec_2_dummy |
391 |
dss_rec_3_dummy |
392 |
dss_rec_4_dummy |
393 |
dss_rec_5_dummy |
394 |
ρ |
395 |
λ |
396 |
d |
397 |
h |
398 |
Secondary Variable - Temperature Compensation |
399 |
Pressure Unit |
400 |
Field Device Variable Unit- The engineering unit (unit of measurement) to be displayed with the Digital Value, Upper Range Value, Lower Range Value, and Upper Range Limit.
Pressure- The pressure of the process measured with respect to a reference pressure. |
401 |
Upper Range Limit |
402 |
Upper Range Limit- Defines the maximum usable value for the Upper Range Value.
Pressure- Pressure of the process measured with respect to a reference pressure. |
403 |
Lower Range Limit |
404 |
Damping |
405 |
4.2f |
406 |
Sensor Serial Number |
407 |
7d |
408 |
Sensor Temp. |
409 |
Temp. Unit |
410 |
Temp. Unit- The engineering unit (unit of measurement) to be displayed with the sensor temperature. |
411 |
degC |
412 |
degF |
413 |
degR |
414 |
Kelvin |
415 |
Process Variables |
416 |
Device |
417 |
Basic Setup |
418 |
Configuration |
419 |
Output Condition |
420 |
Device Information |
421 |
Review |
422 |
Maintenance |
423 |
Adjustment |
424 |
Calibration |
425 |
Simulation |
426 |
Diagnostics |
427 |
Diagnostic Status |
428 |
Diagnostic Status Records |
429 |
Zero-Error Int. Data |
430 |
Diagnostic Setup |
431 |
Pressure Frequency Diagnostic |
432 |
Out-of-Range Pressure Count |
433 |
Failure |
434 |
Function Check |
435 |
Out of Specification |
436 |
Maintenance Required |
437 |
upload_valiables |
438 |
Monitoring |
439 |
Trend |
440 |
Gauge |
441 |
Offline |
442 |
Configuration Data |
443 |
ID |
444 |
Device Tag |
445 |
Analog Output Form |
446 |
Output Limit |
447 |
Fail Safe |
448 |
Revision Numbers |
449 |
Write Protect |
450 |
Input Calibration |
451 |
Output Calibration |
452 |
Factory Setting Information |
453 |
Range |
454 |
Display |
455 |
Remote-seal |
456 |
Alarm |
457 |
Alarm Settings |
458 |
Advance Diagnostic is invalid |
459 |
Input LRV Calibration |
460 |
Input LRV Calibration- Calibrates the zero point (Lower Range Value) based on the current input value. |
461 |
Input URV Calibration |
462 |
Input URV Calibration- Calibrates the span (Upper Range Value) based on the current input value. |
463 |
Reset Calibration |
464 |
Reset Calibration- Discards calibration data. |
465 |
Set LRV (Actual) |
466 |
Set LRV (Actual)- Sets the Lower Range Value at the present output value. |
467 |
Change Cutoff Mode |
468 |
Change Cutoff Mode- Selects the cutoff mode and sets the cutoff point. Valid when Square Root is selected for Transfer Function. |
469 |
Clear Status Records |
470 |
Clear status records- Clears the device's Status Record. |
471 |
Change SW Write Protect |
472 |
Change Software Write Protect- Can be set to 'Yes' or 'No'. |
473 |
Apply Zero Value |
474 |
Apply zero value- Sets the Lower Range Value at the present input pressure. |
475 |
Apply Span Value |
476 |
Apply span value- Sets the Upper Range Value at the present input pressure. |
477 |
Restore Factory Setting |
478 |
Restore factory settings- Return to the factory default calibration data. |
479 |
Open- Simulates the contact output.
Select [Open] to put the device in simulation mode, with contact output circuit is open. |
480 |
Closed- Simulates the contact output.
Select [Closed] to put the device in simulation mode, with contact output circuit is closed. |
481 |
Clear- Simulates the contact output.
Selects [Clear] to cancel simulation mode. |
482 |
Software Version- Displays the software version of device. |
483 |
Scaled D/A Trim |
484 |
Failure Alarm Simulation Mode- Simulates the output when the device is in a critical state. |
485 |
Clear- Simulates the output when the device is in a critical state.
Select [Clear] to cancel simulation mode. |
486 |
Amb. Temp. Compensate |
487 |
Ambient Temperature Compensation- Compensates for the effects of temperature on the fill fluid. The user must enter the flange span and fill fluid type. |
488 |
Auto Range (Closed Tank) |
489 |
Auto Range (closed tank)- Automatically calculates the range (LRV and URV). The user must enter the high pressure flange location, specific gravity, fill fluid, span, flange span, and distance of the zero level above the lower flange. This function is valid when Display Function is linear. |
490 |
Contact Output Simulation Mode- Simulates the contact output.
--Open: Select [Open] to put the device in simulation mode with the contact output circuit open.
--Closed: Select [Closed] to put the device in simulation mode with the contact output circuit closed. |
491 |
Reset Press Freq Index |
492 |
Reset Press Freq Index- Resets the Pressure Frequency Index and the Press Freq Index Max and Min values. |
493 |
Reset Standard Deviation |
494 |
Reset Standard Deviation- Resets the Standard Deviation, the Standard Deviation Max and Min values, and the Average Pressure. |
495 |
Reset OOR Pressure Count |
496 |
Reset OOR Pressure Count- Resets the Out-of-Range Pressure Count. |
497 |
Status |
498 |
Status- Displays the result of self-diagnosis of the device. |
499 |
Change Press Freq Index Sensor Selection |
500 |
Change Press Freq Index Sensor Selection- This function changes the Press Freq Index Sensor Selection.
If the Press Freq Index Sensor Selection is changed, other parameters (Pressure Frequency Index, Press Freq Index Max, Press Freq Index Min) will be reset. |
501 |
Change Standard Deviation Sample Count |
502 |
Change Standard Deviation Sample Count- This function changes the Standard Deviation Sample Count.
If the Standard Deviation Sample Count is changed, other parameters (Standard Deviation, Standard Deviation Max, Standard Deviation Min, Average Pressure) will be reset. |
503 |
PV % rnge Gauge |
504 |
Pressure Gauge |
505 |
Sensor Temp. Gauge |
506 |
Pressure Frequency Index Trend |
507 |
Press Freq Index High/Low Limit |
508 |
Standard Deviation Trend |
509 |
OOR Pressure Count Ratio (%) |
510 |
OOR Pressure Count Ratio (%) = OOR Pressure Count / OOR Count Alarm Threshold |
511 |
Displays Device Status |
512 |
Device Image |
513 |
Common Tables |
514 |
Interoperable * |
515 |
Universal |
516 |
Common Practice |
517 |
Pressure Universal |
518 |
Pressure Common |
519 |
Menu |
520 |
Rosemount Magflow |
521 |
Rosemount Pressure |
522 |
Moore Products |
523 |
Siemens Pressure |
524 |
P2 Common |
525 |
PV |
526 |
Wireless |
527 |
Munge |
528 |
Analytical Oxygen/Combustion/Methane/CO |
529 |
Flow Totalizer - Rate Indicator |
530 |
Flow Variable Area |
531 |
Flow Coriolis Mass |
532 |
Flow Ultrasonic - Gas |
533 |
Flow Ultrasonic - Liquid |
534 |
Level Ultrasonic |
535 |
Level Radar - Microwave |
536 |
Pressure Absolute |
537 |
Level Capacitance |
538 |
Pressure Differential |
539 |
Level Hydrostatic |
540 |
Level |
541 |
Pressure Absolute/Relative |
542 |
Level Microwave |
543 |
Level Radar |
544 |
Level Density - Nuclear |
545 |
Flow Electromagnetic |
546 |
Flow Vortex |
547 |
Flow Magmeter - Rack Version |
548 |
Flow High Performance Magmeter |
549 |
Flow Mass |
550 |
Flow Ultrasonic |
551 |
Flow Magnetic |
552 |
Flow Magnetic - Custody Transfer |
553 |
Flow 2-wire Magnetic |
554 |
Flow Coriolis Mass - Custody Transfer |
555 |
Flow 2-wire Vortex |
556 |
Flow Thermal |
557 |
Temperature Head Mounted |
558 |
Temperature Compact |
559 |
Temperature DIN Rail Mounted |
560 |
Analytical pH |
561 |
Analytical Conductivity - Conductive |
562 |
Analytical Conductivity - Inductive |
563 |
Temperature Sensor |
564 |
Level Servo |
565 |
Temperature DIN Rail |
566 |
Temperature |
567 |
Analytical ORP |
568 |
Analytical pION |
569 |
Analytical IConc |
570 |
Analytical Electrode Conductivity |
571 |
Analytical Toroidal Conductivity |
572 |
Valve Positioner |
573 |
Valve Controller |
574 |
Pressure Multivariable |
575 |
Analytical pH/ORP |
576 |
Flow Coriolis |
577 |
Analytical Conductivity |
578 |
Valve Positioner/Pneumatic |
579 |
Valve Actuator - Electrical |
580 |
Temperature Dual Input |
581 |
Valve Actuator |
582 |
Analytical Corrosion |
583 |
Analytical Density - Nuclear |
584 |
Level Continuous Gamma |
585 |
Level Laser Ranging |
586 |
Pressure Safety |
587 |
Flow Vortex/Swirl |
588 |
Flow |
589 |
Analytical 4 Electrode Conductivity |
590 |
Analytical 2 Electrode Conductivity |
591 |
Analytical Troidal Conductivity |
592 |
Analytical pH/ORP/pIon/Ion Conc |
593 |
Analytical Oxygen |
594 |
Pressure P, dP + Multivariable |
595 |
Temperature 2-wire |
596 |
Pressure w/ PID |
597 |
Pressure Saturated Steam |
598 |
Analytical Density - Gamma Ray |
599 |
Level Gamma Ray |
600 |
Analytical Weight - Gamma Ray |
601 |
Level w/ Density - Gamma Ray |
602 |
Analytical Motorized Density |
603 |
Flow Multivariable Mass |
604 |
Temperature Multivariable |
605 |
Specialty HART-to-Analog Signal Conv |
606 |
Flow dP |
607 |
Pressure Low Power |
608 |
Pressure Certified SIL2 |
609 |
Interoperable |
610 |
Analytical Density |
611 |
Valve Positioner/Electropneumatic |
612 |
Flow Magnetic Inductive |
613 |
Analytical |
614 |
Analytical Opacity Monitor |
615 |
Analytical Oxygen/Combustion |
616 |
Analytical 2-wire Conductivity |
617 |
Analytical 2-wire Conductivity w/ pure H20 |
618 |
Analytical 2-wire Conductivity w/ % conc |
619 |
Analytical 2-wire Oxygen/Chlorine/Ozone |
620 |
Analytical 2-wire pH/ORP |
621 |
Analytical 2-wire Contacting/Toroidal |
622 |
Analytical Oxygen - High Temperature |
623 |
Analytical Conductivity w/ % conc |
624 |
Analytical pH/ORP/ISE |
625 |
Analytical Amperometric Analyzer |
626 |
Analytical Dual Input Analyzer |
627 |
Pressure Pressure/Level |
628 |
Analytical Consistency |
629 |
Analytical Microwave-Pulp Consistency |
630 |
Analytical Microwave Consistency |
631 |
Valve Digital Positioner |
632 |
Valve Smart Valve Positioner |
633 |
Temperature Differential |
634 |
Analytical Dissolved Oxygen |
635 |
Analytical pH Controller |
636 |
Analytical Conductivity/Resistivity Controller |
637 |
Analytical Conductivity/Resistivity |
638 |
Level Displacement |
639 |
Temperature w/ PID |
640 |
Valve Universal Positioner |
641 |
Level Radar - Reflex |
642 |
Pressure Gauge |
643 |
Pressure Sealed Gauge |
644 |
Valve Digital Positioner w/ PID |
645 |
Valve Intelligent Positioner |
646 |
Analytical Gas Detector |
647 |
Analytical Infrared Gas Detector |
648 |
Analytical Electrochemical Gas Detector |
649 |
Analytical Measuring Instruments |
650 |
Level Magnostrictive |
651 |
Valve On/Off Valve Controller |
652 |
Level Radar - Pulse |
653 |
Level Continuous Capactive |
654 |
Valve Damper Actuator |
655 |
Valve Electric Damper Actuator |
656 |
Flow Electronic |
657 |
Flow Vortex - Multiparameter Mass |
658 |
Analytical UV Flame Detector |
659 |
Analytical UVIR Flame Detector |
660 |
Analytical IR Flame Detector |
661 |
Analytical Fire Detector |
662 |
Analytical Gas Detector Display Interface |
663 |
Analytical Toxic Gas Detector |
664 |
Analytical Flame Detector |
665 |
Level Nuclear |
666 |
Analytical Microwave |
667 |
Analytical Density - Radiometric |
668 |
Level Radiometric |
669 |
Analytical Gas/Steam |
670 |
Level Nucleonic |
671 |
Analytical Blade Consistency |
672 |
Analytical Rotary Consistency |
673 |
Pressure Absolute/Differential |
674 |
Positioner |
675 |
Analytical Combustible Gas Detector |
676 |
Analytical Point Infrared Gas Detector |
677 |
Analytical Multi-Detector Monitor System |
678 |
Valve Positioner/Electro-hydraulic |
679 |
Valve Monitor |
680 |
Enable on token-passing data link layer only |
681 |
Enable on TDMA data link layer only |
682 |
Enable on both TDMA and token-passing data link layer |
683 |
WirelessHART Device |
684 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
685 |
Primary |
686 |
Primary Analog Channel |
687 |
Secondary |
688 |
Secondary Analog Channel |
689 |
Tertiary |
690 |
Tertiary Analog Channel |
691 |
Quaternary |
692 |
Quaternary Analog Channel |
693 |
Quinary |
694 |
Quinary Analog Channel |
695 |
Critical Power Failure |
696 |
Lock all |
697 |
Trim not supported |
698 |
Configuration cannot be changed |
699 |
Locked by gateway |
700 |
Analog Input Channel |
701 |
Analog Output Channel |
702 |
Secondary Analog Channel Saturated |
703 |
Tertiary Analog Channel Saturated |
704 |
Quaternary Analog Channel Saturated |
705 |
Quinary Analog Channel Saturated |
706 |
Secondary Analog Channel Fixed |
707 |
Tertiary Analog Channel Fixed |
708 |
Quaternary Analog Channel Fixed |
709 |
Quinary Analog Channel Fixed |
710 |
Simulation active |
711 |
The device is in simulation mode and the level output is not representative of the process. |
712 |
Non-Volatile memory failure |
713 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
714 |
Volatile memory error |
715 |
The RAM memory check is invalid or maybe corrupt |
716 |
Watchdog reset executed |
717 |
A watchdog reset has been performed |
718 |
Voltage conditions out of range |
719 |
A voltage condition is outside its allowable range |
720 |
Environmental conditions out of range |
721 |
An internal or environmental condition is beyond acceptable limits. |
722 |
Electronic failure |
723 |
A hardware problem not related to the sensor has been detected. |
724 |
Subdevice list changed |
725 |
Duplicate master detected |
726 |
Capacity Denied |
727 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
728 |
Bandwidth allocation pending |
729 |
Block transfer pending |
730 |
Continuous |
731 |
The Burst Message is published continuously at (worst case) the Minimum Update Period. |
732 |
Window |
733 |
The Burst Message is triggered when the source value deviates more than the specified trigger value |
734 |
Rising |
735 |
The Burst Message is triggered when source value Rises Above the specified trigger value. |
736 |
Falling |
737 |
The Burst Message is triggered when the source value Falls Below the specified trigger value. |
738 |
On-change |
739 |
The Burst Message is triggered when any value in the message changes. |
740 |
Battery life |
741 |
Percent range |
742 |
Loop current |
743 |
Primary variable |
744 |
Secondary variable |
745 |
Tertiary variable |
746 |
Quaternary variable |
747 |
Enable event notification on token-passing data link layer |
748 |
Enable event notification on TDMA data link layer |
749 |
Enable event notification on TDMA and token-passing data link layer |
750 |
Configuration changed event pending |
751 |
Device status event pending |
752 |
More status available event pending |
753 |
Enable single data point trending |
754 |
Enable filtered trending |
755 |
The sample is filtered using a time constant equal to one-third the trend sample period |
756 |
Enable average trending |
757 |
The value obtained by dividing the sum of all samples in the period by the total number of samples |
758 |
Read receive time |
759 |
Write date and time |
760 |
Non-Volatile Clock |
761 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
762 |
Clock Uninitialized |
763 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
764 |
Line Power |
765 |
Device is powered via a hardwired connection to a plant power source. |
766 |
Battery Power |
767 |
Device operates solely of its (internal) battery. |
768 |
Recharg.B./Power Sc. |
769 |
Rechargeable battery power or power scavenging. The device has a short-term supply of energy (must be at least one hour) that is continuously being replenished. The device's power is being replenished by harvesting and converting energy from the environment surrounding the device (e.g. solar, vibration, heat). |
770 |
Idle |
771 |
Active search |
772 |
Looking for neighbors. |
773 |
Negotiating |
774 |
Quarantined |
775 |
Operational |
776 |
Suspended |
777 |
Deep sleep |
778 |
Ultra-low power |
779 |
Wireless Signal Found |
780 |
Network packets have been recognized |
781 |
Wireless Signal Identified |
782 |
Wireless Time Synchronized |
783 |
WirelessHart Signal Identified |
784 |
Network Admission Requested |
785 |
Set on transmission of first join request |
786 |
Join Retrying |
787 |
Set after the first join request retry (i.e. when number of join requests > 2). Cleared when device is authenticated or when Active Search started. |
788 |
Join Failed |
789 |
Set on transition from Active Search mode to Deep Sleep mode. Cleared on transition from Deep Sleep to Active Search. |
790 |
Network Security Clearance Granted |
791 |
Network Manager Session Established |
792 |
Network Joined |
793 |
Network Joined. Normal superframe and links obtained. |
794 |
Network Bandwidth Requested |
795 |
Negotiating Network Properties. Gateway session obtained. |
796 |
Join Complete |
797 |
Normal operating mode. |
798 |
None |
799 |
No restrictions |
800 |
Restricted to SI Unit codes only |
801 |
Open - accept all devices |
802 |
Use Whitelist for access |
803 |
Use Blacklist |
804 |
Use Whitelist and Blacklist |
805 |
Lockdown - disable join links |
806 |
Process automation device |
807 |
Discrete device |
808 |
Hybrid:Process automation+discrete |
809 |
I/O System |
810 |
WirelessHART Process automation device |
811 |
WirelessHART Discrete device |
812 |
WirelessHART Process automation+discrete |
813 |
WirelessHART Gateway |
814 |
WirelessHART Access point |
815 |
WirelessHART Process adapter |
816 |
WirelessHART Discrete adapter |
817 |
WirelessHART enabled handheld/portable maintenance tool |
818 |
Don't attempt to join |
819 |
Join now |
820 |
Attempt to join immediately on powerup or reset |
821 |
US |
822 |
United States of America |
823 |
JP |
824 |
Japan |
825 |
DE |
826 |
Germany |
827 |
FR |
828 |
France |
829 |
ES |
830 |
Spain |
831 |
RU |
832 |
Russian Federation |
833 |
CN |
834 |
People's Republic of China |
835 |
Nominal |
836 |
Device Power Status is nominal |
837 |
Low |
838 |
Device Power Status is low |
839 |
Critically Low |
840 |
Device Power Status is critically low |
841 |
Recharging - Low |
842 |
Recharging - High |
843 |
Secondary Master |
844 |
Primary Master |
845 |
PV Analog Channel Saturated |
846 |
PV Analog Channel Fixed |
847 |
Device Profile |
848 |
Device Diagnostic Status 0 |
849 |
Device Diagnostic Status 1 |
850 |
I/O and Subdevice Status |
851 |
WirelessHART Status |
852 |
Time Set Option |
853 |
Real Time Clock Flags |
854 |
Country |
855 |
Country of intended device installation |
856 |
SI Unit Control |
857 |
Indicates if device should only display SI unit codes |
858 |
Burst Trigger Units |
859 |
Burst Trigger Class |
860 |
Wired HART Enabled |
861 |
Enabled on Wired HART device connection only |
862 |
Wireless HART Enabled |
863 |
Enabled on WirelessHART device connection only |
864 |
Both Wired and WirelessHart Enabled |
865 |
Enabled on both Wired and WirelessHART connections |
866 |
Burst Msg Trigger Mode |
867 |
Event Status |
868 |
Event Notification Control |
869 |
Operate mode Mask |
870 |
Not Defined |
871 |
Power Source |
872 |
Power source of the device |
873 |
Wireless Mode |
874 |
Join Status |
875 |
The current location in the joining sequence of operations. |
876 |
Network Access Mode |
877 |
Join Mode |
878 |
Power Status |
879 |
Power status of the device |
880 |
Master Mode |
881 |
This sets the I/O System's Data Link Layer operation on all channels as either primary or secondary master mode. |
882 |
Trend Units |
883 |
Trend Device Variable Unit Code |
884 |
Trend Classification |
885 |
Trend Device Variable Classification |
886 |
Trend Control |
887 |
uint16_var_blk_char |
888 |
uint8_var_blk_char |
889 |
Device Status Mask |
890 |
Ext dev status Mask |
891 |
Device Diagnostic Status 0 Mask |
892 |
Device Diagnostic Status 1 Mask |
893 |
Device Diagnostic Status 2 Mask |
894 |
Device Diagnostic Status 3 Mask |
895 |
AO saturated Mask |
896 |
AO fixed Mask |
897 |
Latched Device Status |
898 |
Latched Ext dev status |
899 |
Latched Operate Mode |
900 |
Latched Device Diagnostic Status 0 |
901 |
Latched Device Diagnostic Status 1 |
902 |
Latched Device Diagnostic Status 2 |
903 |
Latched Device Diagnostic Status 3 |
904 |
Latched AO saturated |
905 |
Latched AO fixed |
906 |
Wireless Module Device Type |
907 |
Device type of the Wireless Radio |
908 |
Wireless Module Manufacturer ID |
909 |
Manufacturer of the Wireless Radio contained in this device |
910 |
Subdevice Device Type |
911 |
Device type of the subdevice attached to this system. |
912 |
Subdevice Manufacturer Id |
913 |
Manufacturer of the subdevice attached to this system. |
914 |
8u |
915 |
Dynamic Variables Returned for Device Variables |
916 |
Command Response Truncated |
917 |
Configuration Change Counter Mismatch |
918 |
Status bytes mismatch |
919 |
Device Variable |
920 |
STX Count |
921 |
Count of STX messages received by this device |
922 |
ACK Count |
923 |
Count of ACK messages received by this device |
924 |
BACK Count |
925 |
Count of BACK messages received by this device |
926 |
8.3f |
927 |
Output Status |
928 |
Output Status Byte |
929 |
Uninitialized |
930 |
Normal - one sensor |
931 |
Normal - two sensors |
932 |
Normal - three sensors |
933 |
Shorted |
934 |
Shorted- shorted field wiring. |
935 |
Limited high |
936 |
Limited high- block can not respond to upward movement of value. |
937 |
Open- open field wiring. |
938 |
Limited low |
939 |
Limited low- block can not respond to upward movement of target. |
940 |
Initialization done |
941 |
Initialization done- block initialization is complete. |
942 |
Not selected |
943 |
Not selected- target value provided by block was not selected. |
944 |
No path |
945 |
No path- there is no path to the process. |
946 |
Fail |
947 |
Fail- hardware failure. |
948 |
Bad |
949 |
Bad- block is out of service or has bad input. |
950 |
Not from process |
951 |
Not from process- value is not directly affected by any process information. |
952 |
Local override |
953 |
Local override- block has LO mode. |
954 |
Not Cascade Mode |
955 |
Not RCas Mode |
956 |
Not ROut Mode |
957 |
No communication |
958 |
No communication- (NO_COM). |
959 |
uint32_var_blk_char |
960 |
int8_var_blk_char |
961 |
new itest redef1 label |