Loading…
Name ID Label Description Type Access Class
oxygen_units 4013 Unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R Enum 8
oxygen_value 4014 O2 value Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R Float 10
oxygen_damping_value 4015 Damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R Float 8
oxygen_upper_sensor_limit 4016 USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
oxygen_lower_sensor_limit 4017 LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
oxygen_minimum_span 4018 Min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
oxygen_pv_upper_range_value 4019 URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
oxygen_pv_lower_range_value 401A LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
ao1_transfer_function 401B Xfer fnctn Transfer Function- Defines the transformation function that will be applied from the Field Device Variable to the Analog Output, and Percent Range. The Transfer Function does not affect the Digital Value representation. R/W Enum 32
oxygen_pv_percent_range 401C % rnge Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34
Name ID Label Description Type Access Class