Loading…
Name ID Label Description Type Access Class
xmtr_chk_abort_flag 4164 R/W Unsigned 4096
static_pressure_value 4018 SP Static Pressure- Pressure of the process measured with respect to the reference pressure equal to 0. R Float 10
static_pres_percent 4019 SP % Static pressure percent of range- Static Pressure Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34
snsr_temp_value 401A Snsr temp Sensor temperature- Temperature measured at sensor capsule. R Float 10
easy_engr_unit 4088 Engineer Unit Select Code for Easy Setting R/W Enum 256
engr_disp_value 401B Engr Disp Engineering display- Process variable displayed in user scale and unit. R Float 258
engr_exp 401C Engr exp Engineering display exponents- Exponents for user scale display. R/W Enum 256
engr_disp_unit 401D Engr Unit Engineering unit- Engineering unit for user scale display. R/W ASCII 256
p_lower_trim_dev 400F P LTD Lower pressure trim deviation- Deviation from the original value. R Float 8
p_upper_trim_dev 4010 P UTD Upper pressure trim deviation- Deviation from the original value. R Float 8
Name ID Label Description Type Access Class