Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
D3 | 40DD | D3 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. | R/W | Float | 8 |
D4 | 40DE | D4 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D4 can represent the D4 value for a Micro Motion T-series sensor. | R/W | Float | 8 |
K1 | 40DF | K1 | The tube period, in microseconds, corresponding to a density equal to D1. | R/W | Float | 8 |
K2 | 40E0 | K2 | The tube period, in microseconds, corresponding to a density equal to D2. | R/W | Float | 8 |
K3 | 40E1 | K3 | The tube period, in microseconds, corresponding to a density equal to D3. | R/W | Float | 8 |
K4 | 40E2 | K4 | The tube period, in microseconds, corresponding to a density equal to D4. | R/W | Float | 8 |
DCoef | 40E3 | TC | The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. | R/W | Float | 8 |
DT | 40E4 | DT | The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. | R/W | Float | 8 |
alarm_number | 40F6 | Event num | R/W | Index | 4096 | |
alarm1_type | 40EA | Event 1 Type | Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. | R/W | Enum | 1024 |
Name | ID | Label | Description | Type | Access | Class |