Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
ao_1_fault_level | 408D | 故障レベル | This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. The range is 1.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale. | R/W | Float | 64 |
fo_fault_level | 4090 | 故障レベル | This parameter allows you to select the upper frequency output level that will indicate a fault condition. The maximum value is 15000 Hz. | R/W | Float | 64 |
density_for_cal | 4092 | R/W | Float | 4096 | ||
point_code | 4093 | R/W | Unsigned | 4096 | ||
densityfd_for_cal | 4094 | R/W | Float | 4096 | ||
temp_for_cal | 4095 | Ref temperature | User-entered temperature for the purpose of performing an on-line temperature calibration. | R/W | Float | 4096 |
cutoff_code | 4098 | カットオフ タイプ | R/W | Index | 4096 | |
D1 | 4099 | D1 | The density, in g/cc, of the air or other low-density fluid used during a 2-point density calibration. | R/W | Float | 8 |
D2 | 409A | D2 | The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. | R/W | Float | 8 |
D3 | 409B | D3 | The density, in g/cc, of the water or other high-density fluid that is used during a 4-point density calibration with Micro Motion T-series sensors. D3 can represent the D3 value for a Micro Motion T-series sensor. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |