Name ID Label Description Type Access Class
tube_amplitude_L 40D1 Left Pickoff Amplitude R Float 10
tube_amplitude_R 40D2 Right Pickoff Amplitude R Float 10
drive_gain 40D3 Drive The drive gain of the transmitter. R Float 10
live_zero 40D4 Live Zero Flow Rate This flow value is not effected by the flow cutoff and is damped at 8 seconds. It is to be used for diagnostic purposes only and is not used in ANY flow calculations or totalization. R Float 10
board_temp_700 40D5 Board Temperature The board temperature of the core processor. R Float 10
case_temp 40D6 Meter Temperature (T-Series) The meter temperature of the T-Series sensor. R Float 10
FD 40D7 FD A calibration constant generated during a flowing density calibration. R/W Float 8
tot_reset 40D8 Totalizer Reset When this parameter is enabled, you can use the display to reset the totalizer. R/W Enum 256
auto_scroll 40D9 Auto Scroll When this parameter is enabled, you do not need to press the scroll button to scroll through the configured process variables. R/W Enum 256
auto_scroll_rate 40DA Scroll Time (1-10) This parameter allows you to set the rate at which the display will scroll from one process variable to the next. Display Auto Scroll must be enabled. R/W Unsigned 256
Name ID Label Description Type Access Class