Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
D2 | 40BD | Dens B | The density, in g/cc, of the water or other high-density fluid used during a 2-point density calibration. Dens B can represent: The Dens B (D2) value on an ELITE sensor, a value of 1.0000 g/cc for a D, DL, or DT sensor with a 14-character dens cal factor. | R/W | Float | 8 |
K1 | 40BE | K1 | The tube period, in microseconds, corresponding to a density equal to Dens A. | R/W | Float | 8 |
K2 | 40BF | K2 | The tube period, in microseconds, corresponding to a density equal to Dens B. | R/W | Float | 8 |
DCoef | 40C0 | Temp coeff | The percent change in the elasticity of the flow tubes around the bending axis per 100 degrees Celsius. | R/W | Float | 8 |
alarm_number | 40D2 | Event num | R/W | Index | 4096 | |
alarm1_type | 40C6 | Event1 type | Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. | R/W | Enum | 1024 |
alarm2_type | 40CB | Event2 type | Determines whether the Event is activated when it goes above (High) or below (Low) the setpoint. | R/W | Enum | 1024 |
alarm1_setpoint | 40C2 | Event1 setpoint | For the Model RFT9739, a programmable flow rate, total, density, temperature, differential pressure, or viscosity at which an output set to represent event 1 or event 2 switches states. | R/W | Float | 16 |
alarm2_setpoint | 40C9 | Event2 setpoint | For the Model RFT9739, a programmable flow rate, total, density, temperature, differential pressure, or viscosity at which an output set to represent event 1 or event 2 switches states. | R/W | Float | 16 |
alarm1_varcode | 40C4 | Event1 var | Assigns flow rate, total, density, temperature, differential pressure, or viscosity to event 1 or event 2. | R/W | Index | 16 |
Name | ID | Label | Description | Type | Access | Class |