Loading…
Name ID Label Description Type Access Class
diff_mass_flow_digital_value 4308 Differential Mass Flow Differential Mass Flow R Float 10
analog_damping 4006 Added Damping Analog Output Added Damping Value- Damping time constant applied the Analog Output. The units of this variable are always in seconds. R/W Float 32
mass_flow_digital_damping 4021 Flow Damping Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 32
serial_number 4309 Sensor Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
mass_flow_upper_sensor_limit 4307 Mass Flow Upper Sensor Limit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
mass_flow_lower_sensor_limit 430A Mass Flow Lower Sensor Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
mass_flow_minimum_span 430B Mass Flow Minimum Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
massflowClassificationCode 430E Mass Flow Classification Classification: Indicates the unit code expansion table used for a particular dynamic or device variable. R Enum 0
massflowFamily 430F Mass Flow Family Device Family: Indicates the family (i.e., pressure, temperature, flow, etc.) to which the device belongs. R Enum 2048
massflowPDQ 4310 Status Process Data Quality: Indicates the overall process data quality of the device variable value. R Enum 10
Name ID Label Description Type Access Class