pressureUpdatePeriod 4379 Pressure Update Period R Time 0
drive_digital_units 437A Drive unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit. R Enum 32
drive_digital_value 414C Drive Gain Drive gain is the amount of current required to drive the vibrating element, expressed as a percentage of the maximum allowable drive gain. R Float 10
drive_upper_sensor_limit 437B Upper Sensor Limit Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
drive_lower_sensor_limit 437C Lower Sensor Limit Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
drive_minimum_span 437D Minimum Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
driveClassificationCode 437F Drive Classification Classification: Indicates the unit code expansion table used for a particular dynamic or device variable. R Enum 0
driveFamily 4380 Drive Family Device Family: Indicates the family (i.e., pressure, temperature, flow, etc.) to which the device belongs. R Enum 2048
drivePDQ 4381 Drive PDQ Process Data Quality: Indicates the overall process data quality of the device variable value. R Enum 10
driveLS 4382 Drive LS Limit Status: Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 10
Name ID Label Description Type Access Class