Loading…
Loading…
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
sp_volume_total_text | 40C3 | 总量标签 | A 1 to 8 character description of a special unit of volume, which is based on the special unit of flow to represent total. | 读/写 | ASCII | 2048 |
totctl | 40C4 | 累加器控制 | Allows the user to Reset, Stop, or Start all totalizers. NOTE - Inventory totals are NOT reset. | 读/写 | 枚举 | 4096 |
dens_tgco | 40C5 | DTG | This parameter represents the Density Temperature Gradient Coefficient (DTG) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
dens_fqco1 | 40C6 | DFQ1 | This parameter represents the Density Frequency Coefficient #1 (DFQ1) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
dens_fqco2 | 40C7 | DFQ2 | This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
flow_tgco | 40C8 | FTG | This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
flow_fqco | 40C9 | FFQ | This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. | 读/写 | 浮点 | 8 |
max_zero | 40CA | Zero Time | The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. | 读/写 | 无符号 | 12 |
tempcal | 40CB | Temperature Calibration Factor | A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. | 读/写 | ASCII | 32 |
rt_zero | 40CC | Zero Value | A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. | 读/写 | 浮点 | 10 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |