名称 ID 标签 描述 类型 访问 分类
sp_volume_total_text 40C3 总量标签 A 1 to 8 character description of a special unit of volume, which is based on the special unit of flow to represent total. 读/写 ASCII 2048
totctl 40C4 累加器控制 Allows the user to Reset, Stop, or Start all totalizers. NOTE - Inventory totals are NOT reset. 读/写 枚举 4096
dens_tgco 40C5 DTG This parameter represents the Density Temperature Gradient Coefficient (DTG) that is on the T-series sensor tag. 读/写 浮点 8
dens_fqco1 40C6 DFQ1 This parameter represents the Density Frequency Coefficient #1 (DFQ1) that is on the T-series sensor tag. 读/写 浮点 8
dens_fqco2 40C7 DFQ2 This parameter represents the Density Frequency Coefficient #2 (DFQ2) that is on the T-series sensor tag. 读/写 浮点 8
flow_tgco 40C8 FTG This parameter represents the Flow Temperature Gradient Coefficient (FTG) that is on the T-series sensor tag. 读/写 浮点 8
flow_fqco 40C9 FFQ This parameter represents the Flow Frequency Coefficient (FFQ) that is on the T-series sensor tag. 读/写 浮点 8
max_zero 40CA Zero Time The amount of time the transmitter takes to determine its zero-flow reference point. The default zero time is 20 seconds. 读/写 无符号 12
tempcal 40CB Temperature Calibration Factor A 14-character factor that describes the slope and offset of the linear output representing flow tube temperature. 读/写 ASCII 32
rt_zero 40CC Zero Value A value, in microseconds, representing the phase shift between velocity signals from the sensor when measuring zero flow. 读/写 浮点 10
名称 ID 标签 描述 类型 访问 分类