Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
flow_set20mA | 401D | Flow 20 mA Set Point | Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. | R/W | Float | 64 |
sensor_serial_number | 401E | Snsr s/n | R | Unsigned | 2048 | |
Range | 401F | Range | The distance from the face of the transducer to the bottom of the vessel or flow element. | R/W | Float | 2048 |
vessel_type | 4020 | Vessel Type | Shape of vessel. | R/W | Enum | 2048 |
radius | 4021 | Radius | Radius of the cylindrical portion of the vessel. | R/W | Float | 2048 |
ellipse_height | 4022 | Ellipse Height | Height of the elliptical portion of the vessel. | R/W | Float | 2048 |
conical_height | 4023 | Conical Height | Height of the conical portion of the vessel. | R/W | Float | 2048 |
width | 4024 | Width | Horizontal width of the rectangular vessel. | R/W | Float | 2048 |
length | 4025 | Length | Horizontal length of the rectangular vessel or of the cylindrical portion of a vessel with elliptical or spherical ends. | R/W | Float | 2048 |
ellipse_end | 4026 | Ellipse End | Horizontal length of the elliptical end of the vessel. | R/W | Float | 2048 |
Name | ID | Label | Description | Type | Access | Class |