Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
pressure_lower_range_limit | 1394 | Pres LSL | Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. | R | Float | 8 |
pressure_minimum_span | 1390 | Pres min span | Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. | R | Float | 8 |
pressure_output_transfer_function | 1391 | Pres xfer fnctn | Transfer Function- Defines the transformation function that will be applied from the Field Device Variable to the Analog Output, and Percent Range. The Transfer Function does not affect the Digital Value representation. | R/W | Enum | 32 |
pressure_upper_range_value | 1396 | Pres URV | Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. | R/W | Float | 32 |
pressure_lower_range_value | 1397 | Pres LRV | Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. | R/W | Float | 32 |
pressure_damping_value | 1398 | Pres damp | Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. | R/W | Float | 8 |
flange_type | 139A | Flange type | Flange Type- Hardware, adjacent to the sensor, that physically connects the process to the sensor. | R/W | Enum | 2048 |
o_ring_material | 139B | O ring matl | O Ring Material- Construction material of a seal that exists between the module and flange. | R/W | Enum | 2048 |
meter_installation | 139C | Local meter | Local meter- Defines whether there is a meter installed or not and possibly what information the meter will display. | R/W | Enum | 256 |
remote_seal_type | 139D | RS type | Remote Seal Type- Type of device that is capable of detecting and passing on the process pressure on to the module. This is only unambiguous if 1 remote seal is installed. | R/W | Enum | 2048 |
Name | ID | Label | Description | Type | Access | Class |