Loading…
Name ID Label Description Type Access Class
module_fill_fluid 139E Fill fluid Fill Fluid- Fluid material in the module. R Enum 2048
number_of_remote_seals 139F Num remote seal Number of Remote Seals- Physical quantity of installed Remote Seals. R/W Enum 2048
analog_output_value 1779 AO Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes R/W Float 66
analog_output_alarm_select 177A AO alrm typ Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. R/W Enum 64
pressure_sensor_serial_number 1770 Pres snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. 0 indicates, that there is currently no Sensor Serial Number determined, because the sensor is uniquely defined by the Device ID. R Unsigned 2048
module_type_code 177B Meas typ Measurement Type- Type of measurement application this Field Device is intended to be used with (Diff, Gauge, Absolute, Liquid-lvl). R Enum 2048
sensor_a_lower_sensor_trim_point 1774 Lower sensor trim point Lower Sensor Trim Point- Applied process pressureat which the last lower sensor trim was performed. R/W Float 8
sensor_a_upper_sensor_trim_point 1776 Upper sensor trim point Upper Sensor Trim Point- Applied process pressureat which the last upper sensor trim was performed. R/W Float 8
flange_material 177C Flnge matl Flange Material- Construction material of flange type. R/W Enum 2048
remote_seal_isolator_material 177E RS isoltr matl Remote Seal Isolator Material- Construction material of a part of the remote seal type that comes in contact with the process material. This is only unambiguous if 1 remote seal is installed. R/W Enum 2048
Name ID Label Description Type Access Class